Shimadzu LC 2030C NT
Shimadzu LC 2030C NT — Chromatographs.
The Shimadzu LC 2030C NT is a parallel-type double-plunger HPLC pump configured for four-solvent low-pressure gradient operation, with a five-line degassing unit (four mobile phases plus one rinse solution). It delivers flow rates from 0.0001 to 10 mL/min against packed-column backpressures up to 34 MPa (5 mL/min and below) or 22 MPa (above 5 mL/min), and supports slide-in columns up to 7 MPa at low flow. This modification is intended for analytical and semi-preparative separations requiring precise, low-pulsation solvent delivery with automated gradient control.
Direct drop-in replacements
Full specifications
Description
Construction and positioning
- Parallel-type double-plunger pump head with independent plunger drives, reducing pulsation to ≤0.1 MPa under standard test conditions (1.0 mL/min, 10 MPa, water). - Integrated five-line degasser: four mobile-phase channels plus one rinse-solution channel (400 µL per line), eliminating separate vacuum-degassing modules for standard four-solvent methods. - Low-pressure gradient proportioning valve mixes up to four solvents in 0.1% steps from 0 to 100%; no high-pressure mixing hardware is required. - Two flow-path configurations: packed-column side (system delay volume ≈615 µL) and slide-in column side (≈620 µL), the latter limited to 7 MPa and 0.3 mL/min. ### Behavior under load and in different modes
- Maximum pressure is flow-dependent: 34 MPa for 0.0001–5 mL/min and 22 MPa for 5.0001–10 mL/min on the packed-column side. Exceeding these limits triggers overpressure protection. - Flow accuracy is ≤±1% or ±2 µL/min (whichever is greater) from 0.01 to 2 mL/min, and ≤±2% from 2 to 5 mL/min under specified conditions; precision is ≤0.06 %RSD or ≤0.02 minSD. - Gradient concentration accuracy is ±0.5% at 0.1–2 mL/min and 1–20 MPa; precision is ±0.1% at 1 mL/min and 10 MPa. These tolerances apply only within the stated pressure and flow windows. - The slide-in column side is a low-pressure path (≤7 MPa, ≤0.3 mL/min) intended for guard or short columns, not for high-efficiency analytical separations. - Solvent lines connect to the four degasser inlets and the rinse port; the gradient proportioning valve is plumbed to the pump inlet. Tubing and fittings are not included in the base delivery unless specified. - Control is via the chromatograph's integrated system controller or external software; no separate pump controller is required. The pump responds to flow, gradient and pressure setpoints from the method. - No PDF manual digests were supplied with this record, so detailed wiring, I/O and startup sequencing must be taken from the official instruction manual.
Integration & Operating Notes
The pump's maximum pressure drops from 34 MPa to 22 MPa when flow exceeds 5 mL/min. Methods that ramp flow above 5 mL/min while a high-impedance column is installed can trip overpressure protection or damage the column; verify the pressure-flow envelope before writing the method.
The slide-in column side has a much lower pressure and flow ceiling than the packed-column side. It is suitable only for short guard or slide-in columns; using it with an analytical column will exceed the 7 MPa limit and stop the run.
The ±0.5% gradient concentration accuracy is specified for 0.1–2 mL/min and 1–20 MPa. Outside this window, gradient composition may deviate; methods requiring tight compositional control should keep flow and pressure inside these bounds or be validated separately.
Four mobile-phase inlet lines plus one rinse line and the waste outlet must be routed without kinks or upward loops that trap air. Plan bench space so the degasser inlets and the pump outlet to the injection valve are accessible for priming and leak checks.
Selection & Integration Q&A
Choose this model when you need four-solvent low-pressure gradient operation with a five-line degasser and flow up to 10 mL/min at 22–34 MPa. A sibling with a different pump head or gradient type would be preferred if your method requires higher pressure at high flow or binary high-pressure mixing.
The packed-column side accepts columns rated for at least 34 MPa at flows up to 5 mL/min; above 5 mL/min the column and fittings must withstand 22 MPa. The slide-in side requires a column rated for at least 7 MPa and flows no higher than 0.3 mL/min. Use fittings compatible with the system delay volume of about 615 µL.