Shimadzu LC 2030C NT
Shimadzu ·Made in Japan · In Active Production

Shimadzu LC 2030C NT

Shimadzu LC 2030C NT — Chromatographs.

The Shimadzu LC 2030C NT is a parallel-type double-plunger HPLC pump configured for four-solvent low-pressure gradient operation, with a five-line degassing unit (four mobile phases plus one rinse solution). It delivers flow rates from 0.0001 to 10 mL/min against packed-column backpressures up to 34 MPa (5 mL/min and below) or 22 MPa (above 5 mL/min), and supports slide-in columns up to 7 MPa at low flow. This modification is intended for analytical and semi-preparative separations requiring precise, low-pulsation solvent delivery with automated gradient control.

Direct drop-in replacements

0%
match
⚖️ Compare
0%
match
⚖️ Compare

Full specifications

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Do not exceed 22 MPa above 5 mL/min on packed columns

The pump's maximum pressure drops from 34 MPa to 22 MPa when flow exceeds 5 mL/min. Methods that ramp flow above 5 mL/min while a high-impedance column is installed can trip overpressure protection or damage the column; verify the pressure-flow envelope before writing the method.

Factory Specification Sheet, Maximum pressure
Process Limits
Slide-in column path is limited to 7 MPa and 0.3 mL/min

The slide-in column side has a much lower pressure and flow ceiling than the packed-column side. It is suitable only for short guard or slide-in columns; using it with an analytical column will exceed the 7 MPa limit and stop the run.

Factory Specification Sheet, Maximum pressure
Control & I/O
Gradient accuracy is valid only within 0.1–2 mL/min and 1–20 MPa

The ±0.5% gradient concentration accuracy is specified for 0.1–2 mL/min and 1–20 MPa. Outside this window, gradient composition may deviate; methods requiring tight compositional control should keep flow and pressure inside these bounds or be validated separately.

Factory Specification Sheet, Gradient / concentration accuracy
Mounting
Allow clearance for solvent lines and waste tubing

Four mobile-phase inlet lines plus one rinse line and the waste outlet must be routed without kinks or upward loops that trap air. Plan bench space so the degasser inlets and the pump outlet to the injection valve are accessible for priming and leak checks.

Factory Specification Sheet, Degassing unit

Selection & Integration Q&A

When should I choose the LC 2030C NT instead of a sibling variant in the series?

Choose this model when you need four-solvent low-pressure gradient operation with a five-line degasser and flow up to 10 mL/min at 22–34 MPa. A sibling with a different pump head or gradient type would be preferred if your method requires higher pressure at high flow or binary high-pressure mixing.

What are the requirements for the column and plumbing connected to this pump?

The packed-column side accepts columns rated for at least 34 MPa at flows up to 5 mL/min; above 5 mL/min the column and fittings must withstand 22 MPa. The slide-in side requires a column rated for at least 7 MPa and flows no higher than 0.3 mL/min. Use fittings compatible with the system delay volume of about 615 µL.