Shimadzu MCT Series
Shimadzu ·Made in Japan · In Active Production

Shimadzu MCT Series

Shimadzu MCT Series — Hardness testers.

The MCT Series is a compression testing system using electromagnetic force loading with a differential-transformer displacement sensor, designed for micro-scale compression tests on particles and fibers. It offers two loading ranges (9.8–4903 mN and 9.8–1961 mN) and supports compression, load-unload, and cycle test modes. Its optical monitor and motorized stage enable precise sample positioning and observation during testing.

Full specifications

Core Performance & Functional Specifications
Digital camera 1/2.9-inch color camera (210, 211)
Eyepiece 10× (210, 211)
Field of view Vertical: Approx. 90 μm; Horizontal: App (210, 211)
Horizontal positioning motor Stepping motor (210, 211)
Illumination method Epi-illumination (210, 211)
Light source lamp LED 3W / 3V (210, 211)
Loading method Electromagnetic force
Lower compression plate SKS flat plate, 25 mm dia. (210, 211)
Max objective lens attachments 10×, 50× (20×, 40×, 100× with option) (510, 511)
210, 211: 10×, 40×
Measurement method 0 to 100 μm (510, 210)
511, 211: 0 to 10 μm · 210, 211: Particle area / area Equivalent Diameter
Minimum increment 0.001 μm (510, 210)
511, 211: 0.0001 μm
Continuous test count Max. 100 points
Test force accuracy ±1 %
Test mode Compression test, load-unload test (210, 211)
Upper pressure indenter Flat indenter (50 μm dia.)
Vertical positioning range Approx. 20 mm (210, 211)
Loading range (max test force, manual model) mfr 1961 mN (max test force, manual model)
Min. increment (with a 40× objective lens) mfr Approx. 0.2 μm (with a 40× objective lens)
Area mfr Approx. W125 × D125 mm (210, 211)
Horizontal positioning range (both X and Y directions) mfr 50 mm (both X and Y directions)

Description

Construction and Positioning

Integration & Operating Notes

Electrical
Grounding resistance must be ≤100 Ω to avoid measurement noise

The 3-pin power connector's grounding pin must be connected to a ground with resistance of 100 Ω or less. Poor grounding can introduce electrical noise into the sensitive displacement and load measurements, compromising test accuracy and potentially damaging the instrument.

Factory Specification Sheet, Utility section
Process Limits
Humidity must stay between 20 and 60% RH without condensation

Operating the MCT Series outside the 20–60% relative humidity range or with condensation may cause drift in the differential-transformer displacement sensor and affect the optical components. Ensure the laboratory environment is controlled accordingly.

Factory Specification Sheet, Humidity section

Selection & Integration Q&A

When should I choose the MCT Series over other hardness testers for small particles?

Choose the MCT Series when you need compression testing of individual particles or fibers with force up to 4903 mN and displacement resolution in the micrometer range. It is not a conventional indentation hardness tester; it is specialized for micro-compression and load-unload/cycle tests on small samples.

What are the power and grounding requirements for installing the MCT Series?

The system requires single-phase 100–115 V AC or 230 V AC (±10%), 50/60 Hz, with power consumption of 100 W. The 3-pin connector's ground pin must be connected to a ground resistance of 100 Ω or less to ensure measurement stability.

What environmental conditions are necessary for reliable operation?

Operate the MCT Series at 20–60% relative humidity without condensation and within the specified temperature range (1.0 °C tolerance). Avoid direct sunlight, drafts, and vibration sources that could affect the sensitive displacement measurement.

What accessories or consumables are not included in the base delivery?

The base unit includes a flat diamond indenter (50 µm) and standard 10× and 50× objective lenses. Optional items include a 500 µm flat indenter, diamond pressure plate, and additional objective lenses (20×, 40×, 100×) for higher magnification. No sample preparation tools or calibration standards are mentioned as standard.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
347-05375
Manual ·3,1 МБ ·182 pp.
View PDF
PDF
c227-e030
Datasheet ·2,5 МБ ·16 pp.
View PDF