Shimadzu Nexera Prep
Shimadzu ·Made in Japan · In Active Production

Shimadzu Nexera Prep

Shimadzu Nexera Prep — Chromatographs.

The Nexera Prep is a parallel-type double-plunger solvent delivery pump for preparative chromatography, with a maximum flow rate of 150.0 mL/min and a maximum discharge pressure of 44 MPa (10 µL plunger) or 49 MPa (47 µL plunger). It is distinguished within the series by its high-flow capability and optional plunger rinsing mechanism, making it suitable for scaling up purification processes from analytical to preparative scale.

Direct drop-in replacements

0%
match
⚖️ Compare
0%
match
⚖️ Compare

Full specifications

Core Performance & Functional Specifications
Detector types UV-VIS SPD-40, PDA SPD-M40, RID-20A, ELSD-LT III, LCMS-2050
Plunger capacity (LC-40D) mfr 10 μL (LC-40D)
Plunger capacity (LC-20AR) mfr 47 μL (LC-20AR)
Plunger capacity (LC-20AP) mfr 250 μL (LC-20AP)
Maximum discharge pressure (LC-40D) mfr 44 MPa (LC-40D)
Maximum discharge pressure (LC-20AR) mfr 49 MPa (LC-20AR)
Maximum discharge pressure (LC-20AP) mfr 42 MPa (LC-20AP)
Flow rate setting range (LC-40D) mfr 0.01 to 20.00 mL/min (LC-40D)
Flow rate setting range (LC-20AR / LC-20AP) mfr 0.01 to 150.00 mL/min (LC-20AR / LC-20AP)
Flow rate (SIL-40C autosampler) mfr Up to 10 mL/min (SIL-40C autosampler)
Size (Column Holder) mfr W 250 × H 465 × D 400 mm (Column Holder)
Size (Column Holder SLIM) mfr W 110 × H 625 × D 500 mm (Column Holder SLIM)
Installable columns (Column Holder) mfr I.D. 20 to 50 mm columns, 2 pc. (Column Holder)
Pore size (C18 / C8 / PFPP columns) mfr 12 nm (C18 / C8 / PFPP columns)
Pore size (HD-C18 column) mfr 8 nm (HD-C18 column)
Column I.D. (conventional scale) (Conventional) mfr 3 to 8 mm (Conventional)
Flowrate (conventional scale) (Conventional) mfr 0.5 to 3 mL/min (Conventional)
Column I.D. (semi-preparative scale) (Semi-preparative) mfr 10 to 20 mm (Semi-preparative)
Flowrate (semi-preparative scale) (Semi-preparative) mfr 3 to 10 mL/min (Semi-preparative)
Column I.D. (preparative scale) (Preparative) mfr 30 to 50 mm (Preparative)
Flowrate (preparative scale) (Preparative) mfr 10 to 100 mL/min (Preparative)
Maximum flowrate capacity (LC-20AR) mfr 20 mL/min (LC-20AR)
Maximum flowrate capacity (LC-40D) mfr 10 mL/min (LC-40D)
Maximum sample injection volume (SIL-40 setup) mfr 2 mL (SIL-40 setup)
Maximum number of LH-40 and/or FRC-40 units (system expansion) mfr six units (system expansion)
Detector signal input channels mfr up to four channels
Dimensions, Weight & Mechanical Design
Size and weight (LC-40D) mfr W260 × H140 × D500 mm, 10 kg (LC-40D)
Size and weight (LC-20AR) mfr W260 × H140 × D500 mm, 16 kg (LC-20AR)
Size and weight (LC-20AP) mfr W260 × H210 × D500 mm, 19 kg (LC-20AP)
Materials, Media & Process Compatibility
Flow rate (LH-40 liquid handler) mfr Up to 150 mL/min (LH-40 liquid handler)
pH range (C18 / C8 columns) mfr 1 – 12 (C18 / C8 columns)
pH range (HD-C18 column) mfr 1 – 10 (HD-C18 column)
pH range (Phenyl / PFPP columns) mfr 1 – 8 (Phenyl / PFPP columns)

Description

Construction and Positioning

The Nexera Prep is a solvent delivery module built around a parallel-type double-plunger pump. It is available with three plunger capacities—10 µL, 47 µL, and 250 µL—each corresponding to a different maximum discharge pressure (44 MPa, 49 MPa, and 42 MPa, respectively). The 47 µL variant offers the highest pressure tolerance, while the 250 µL variant is optimized for maximum flow rate. The pump supports constant-pressure solvent delivery, which is essential for maintaining stable flow under varying backpressures typical in preparative separations. The optional plunger rinsing mechanism (syringe or rinsing pump, part number 228-39625-41) helps extend seal life when using buffers or aggressive solvents. ### Performance Characteristics

Flow rate accuracy is specified as no more than ±1% or ±2 µL/min (whichever is greater) under defined conditions, and flow rate precision is ≤0.06% RSD or 0.02 min SD. These specifications ensure reliable delivery for both analytical and preparative workflows. The pump can deliver up to 150.0 mL/min, making it suitable for high-throughput purification. The parallel plunger design minimizes pulsation, and the constant-pressure mode is particularly useful when backpressure fluctuates, such as during gradient elution or column switching. The pump is compact (W260 × H140 × D500 mm for 10 µL and 47 µL variants; W260 × H210 × D500 mm for 250 µL) and weighs between 10 kg and 19 kg depending on configuration. It is intended for integration into a chromatography system, requiring connection to solvent reservoirs, a sample injector, and a column. The plunger rinsing mechanism, if used, requires an additional rinsing pump or syringe.

Integration & Operating Notes

Process Limits
Maximum pressure depends on plunger size

The 10 µL plunger supports up to 44 MPa, the 47 µL up to 49 MPa, and the 250 µL up to 42 MPa. Exceeding these limits can damage the plunger seals and affect flow accuracy. Ensure system backpressure remains below the applicable limit for the installed plunger.

Factory Specification Sheet
Maintenance
Plunger rinsing extends seal life with buffers

When using buffer solutions, the optional plunger rinsing mechanism (syringe or rinsing pump, part number 228-39625-41) should be employed to prevent salt crystallization and prolong seal life. Without rinsing, maintenance intervals may be shortened.

Factory Specification Sheet
Process Limits
Flow accuracy varies with flow rate range

Flow rate accuracy is specified as ±1% or ±2 µL/min (whichever is greater) under defined conditions, but for flow rates between 0.1 and 5.0 mL/min, the tolerance is ±1% or ±10 µL/min. This must be considered when precise low-flow delivery is required.

Factory Specification Sheet
Info
Constant pressure mode available

The pump supports constant-pressure solvent delivery, which can be advantageous when system backpressure varies, such as during gradient methods or when using columns with changing permeability. This mode helps maintain stable flow.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the Nexera Prep with 47 µL plunger over the 10 µL or 250 µL versions?

Choose the 47 µL plunger when you need the highest pressure tolerance (49 MPa) and moderate flow rates. The 10 µL is for lower flow rates with slightly lower pressure (44 MPa), while the 250 µL is for maximum flow (150 mL/min) but lower pressure (42 MPa). Select based on required flow and backpressure.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
FRC-40 Precautions for Use - Fraction Collector
Manual ·1,1 МБ ·25 pp.
View PDF
PDF
c193-e034
Datasheet ·3,9 МБ ·24 pp.
View PDF
PDF
LH-40 Instruction Manual - Liquid Handler
Manual ·6,1 МБ ·175 pp.
View PDF
PDF
Nexera Prep System Guide
Datasheet ·8,3 МБ ·234 pp.
View PDF