Shimadzu Quantasep Adept 300
Shimadzu Quantasep Adept 300 — Laboratory balances.
The Quantasep Adept 300 is a single-use, automated chromatography system with a 20 mL hold-up volume and two feed/delivery pumps, each supporting four lines (expandable to six). It operates at flow rates of 10–100 mL/min or 30–300 mL/min (±3%) and system pressure up to 50 psi, with integrated UV, conductivity, and pH monitoring. This modification is distinguished by its single-use flow path and absence of inline buffer dilution/mixing, positioning it for process development and small-scale production where cross-contamination risk must be minimized.
Full specifications
Description
Construction and Positioning
- Single-use flow path with wetted materials including quartz/polypropylene flow cell, silicone tubing and gaskets, and glass pH electrode. - Two feed/delivery pumps, each with four lines (expandable to six); six fractionation lines (expandable to ten). - Supports one connected column with forward, reverse, or bypass flow direction. - Compared to higher-end models in the series, this unit omits inline buffer dilution and mixing, making it suitable for applications where buffers are pre-prepared. ### Hydraulic and Sensing Behavior
- Flow rate ranges: 10–100 mL/min and 30–300 mL/min, both ±3% accuracy. - System pressure limit: 50 psi; pressure sensor range 0–100 psi (±1 psi). - UV detection: 0–2 AU (±0.1 AU) at 280 nm standard (254 nm optional). - Conductivity: 0–200 mS/cm (±5% F.S); pH: 1–14 (±0.2 pH). - Optional flow meter, pre-column pH and conductivity sensors available. - Hold-up volume: 20 mL, minimizing delay between pump and column. ### Installation and Operation
- Operating temperature range: 4–37 °C; ensure ambient conditions within this range. - Utilities: 110/220 V, 6 A; provide stable power supply. - Dimensions: 762 × 673 × 610 mm (W × H × D); allow clearance for tubing and service access. - No specific installation or control details provided in attached documents; refer to manufacturer's manual for connection and operation specifics.
Integration & Operating Notes
Exceeding the 50 psi limit may damage single-use flow path components or cause leaks. The pressure sensor range is 0–100 psi, but the system is rated for 50 psi; ensure pressure alarms are set accordingly.
The flow path is intended for single use; replace tubing, gaskets, and other wetted parts according to the manufacturer's recommendations or after each batch to prevent cross-contamination and ensure consistent performance.
Selection & Integration Q&A
Choose the Adept 300 when your process uses pre-prepared buffers and does not require inline dilution or mixing. Higher-end models include those features but may have a larger footprint or higher cost. The Adept 300 is ideal for process development and small-scale production with single-use requirements.
The system supports one column with forward, reverse, or bypass flow. Ensure the column's pressure rating exceeds the system's 50 psi limit. Use appropriate fittings compatible with the single-use flow path; the hold-up volume is 20 mL, so consider delay volume in method development.
The base system includes the instrument and standard flow path components. Optional items such as the flow meter, pre-column pH and conductivity sensors, and additional pump lines (up to six per pump) or fractionation lines (up to ten) are not included. Replacement single-use tubing and sensors are consumables and must be ordered separately.