Shimadzu Tescan Dynatom — Spectrometers.

The Tescan Dynatom is a laboratory micro-CT scanner built around a rotating gantry with a fixed sample, enabling 4D dynamic micro-CT with sub-10-second temporal resolution. Unlike conventional systems that rotate the sample, this configuration keeps the specimen stationary, allowing in-situ experiments under heating, compression, flow, or electrochemical conditions. It is used for real-time, non-destructive 3D imaging of evolving microstructures in materials and biological samples.

Full specifications

Description

Construction and positioning

The Dynatom uses a rotating gantry architecture: the X-ray source and detector rotate around a stationary sample. This differs from typical micro-CT designs where the sample rotates, and it is the key enabler for dynamic studies because the sample and any in-situ loading rig remain fixed in space. The system is a laboratory instrument, not a portable or industrial inline scanner. ### Behavior and modes

The defining capability is 4D dynamic micro-CT: volumetric datasets are acquired with a temporal resolution of under 10 seconds per full rotation. This allows time-resolved imaging of processes such as heating, compression, fluid flow, or electrochemical reactions. Real-time 4D reconstruction is performed, so the evolving 3D structure can be visualized during or immediately after acquisition. The fixed-sample geometry avoids motion artifacts from sample rotation and simplifies the integration of in-situ cells. ### Control and operation

Acquisition and reconstruction are controlled by the Acquila and Panthera software suites. These handle scan planning, real-time reconstruction, and visualization of 4D data.

Integration & Operating Notes

Mounting
Fixed sample stage simplifies in-situ rig integration

Because the gantry rotates and the sample remains stationary, in-situ loading devices (heating stages, compression cells, flow cells) can be mounted directly on the sample stage without rotating unions or slip rings. This reduces mechanical complexity and potential leak paths for fluid or electrical connections to the sample environment.

Factory Specification Sheet
Control & I/O
Acquisition and reconstruction via Acquila and Panthera

The system relies on these software packages for scan control, real-time 4D reconstruction, and visualization.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the Dynatom over a conventional sample-rotating micro-CT?

Choose the Dynatom when the sample must remain stationary, especially for in-situ experiments with heating, compression, flow, or electrochemical cells. For static samples where in-situ capability is not needed, a sample-rotating system may offer higher resolution or lower cost.

What are the requirements for integrating an in-situ compression rig with the Dynatom?

The rig must fit on the fixed sample stage and remain within the field of view. It should not obstruct the rotating gantry path. Electrical and fluid connections to the rig can be static, simplifying integration. Synchronization with acquisition may be needed for time-resolved studies.

Dimensional Drawings & Technical Documentation