Shimadzu Tescan Solaris 2
Shimadzu Tescan Solaris 2 — Microscopes.
The Tescan Solaris 2 is a field-emission scanning electron microscope (FE-SEM) from Shimadzu, configured as a high-resolution imaging platform. It is distinguished within its class by a fully integrated electron optics column and detection system, optimized for nanoscale surface characterization. Its primary purpose is to provide high-resolution secondary and backscattered electron images for materials science and quality control.
Full specifications
Description
Construction and Architecture
- The Solaris 2 is a standalone FE-SEM with a Schottky field-emission gun, electromagnetic lenses, and a vacuum chamber. The column and chamber are integrated into a single unit, with vibration isolation and thermal stabilization for stable imaging. ### Behavior Under Load and Media
- The instrument operates under high vacuum (typically <10^-3 Pa) in the specimen chamber. - No load-dependent or media-specific behavior data is available. ### Installation, Connection, and Control
- Installation requires a stable power supply, cooling water, and compressed air (if applicable). The control system is integrated, with a PC workstation running proprietary software.
Integration & Operating Notes
The FE-SEM column is sensitive to mechanical vibration and electromagnetic interference; site survey must confirm floor vibration below 1 µm/s and AC magnetic field below 0.1 µT to avoid resolution loss.
The instrument requires a dedicated power line with voltage stability within ±5% and total harmonic distortion below 5% to prevent beam instability and electronics faults.
The cooling water circuit must provide flow of at least 2 L/min at 20°C with filtration to 50 µm; failure to maintain this can cause overheating and shutdown.
The Schottky emitter requires periodic conditioning and the apertures need cleaning every 6 months or after 1000 hours of operation to maintain beam current stability.
Selection & Integration Q&A
The system requires a dedicated 208-240 V AC single-phase supply with stable voltage and low harmonic distortion, plus cooling water at 20°C with a flow of at least 2 L/min and filtration to 50 µm.
Install in a room with temperature stability of ±1°C per hour, relative humidity below 60%, and low vibration and magnetic fields. Avoid locations near heavy machinery or AC power lines.
The base delivery typically excludes consumables such as apertures, filaments, and sample holders. A service contract and spare parts kit are recommended. Refer to the maintenance schedule for periodic replacement intervals.
Dimensional Drawings & Technical Documentation
Dimensional drawings
Click drawing to inspect vector dimensions