Shimadzu TOC 1000E S
Shimadzu TOC 1000E S — TOC analyzers.
The TOC 1000E S is a UV oxidation-conductivity TOC analyzer with a 0–500 µg/L range and 0.02 µg/L detection limit, packaged in a 270×180×140 mm, 3.0 kg enclosure. It is distinguished within the series by its 5-minute startup and selectable 5/10/15/30-minute measurement cycles, making it suited to continuous low-level organic carbon monitoring in pure-water systems.
Full specifications
| Detection limit | 0.02 μg/L |
| Measurement range | 0 to 500 μg/L |
| Repeatability (when concentration is no greater than 1 (unit not printed)) mfr | ±0.05 (when concentration is no greater than 1 (unit not printed)) |
| Accuracy (500 sucrose) mfr | ±5% (500 sucrose) |
| Linearity (when measuring 250 in the 0-500 μg/L range) mfr | R2 = 0.98 or greater (when measuring 250 in the 0-500 μg/L range) |
| TOC measurement principle mfr | UV oxidation-conductivity measurement method |
| Repeatability *1 mfr | 0.05 μg/L |
| Accuracy *1 mfr | 5 % |
| Linearity *1 mfr | 2 |
| Startup time mfr | 5 minutes |
| Measurement cycle mfr | 5 minutes |
| External dimensions mfr | 270 |
| Protection from dust and water spray mfr | 33 |
| Weight | 3.0 kg TOC-1000e S: approximately 3 kg |
| Wiring length mfr | 10 m |
| Required power supply mfr | 100 to 240 V |
Description
Construction and positioning
- Single-channel online TOC analyzer using UV oxidation followed by conductivity detection; no combustion or reagent-based oxidation path is described in the supplied facts. - Compact enclosure: 270 mm W × 180 mm H × 140 mm D at 3.0 kg, rated IP33 for dust and water-spray protection. ### Measurement behavior
- Working range 0–500 µg/L TOC with a 0.02 µg/L detection limit and 0.05 µg repeatability; accuracy stated as 5.0 %. - Linearity R² ≥ 0.98 when measuring 250 µg/L across the 0–500 µg/L range. - Selectable measurement cycles of 5, 10, 15 or 30 minutes; startup time is 5 minutes. - No digitized curve points were supplied, so behavior under varying load or media cannot be characterized beyond the stated linearity and repeatability figures. ### Installation and operation
- Powered from 100–240 V AC, 50/60 Hz at 100 VA. - Signal and control wiring (LAN, analog, contact) is limited to 10 m or less.
Integration & Operating Notes
IP33 protects against solid objects above 2.5 mm and water spray at up to 60° from vertical, but not against immersion or directed jets. Install where the enclosure is not submerged or hosed directly, consistent with the stated protection rating.
The specified measurement range is 0–500 µg/L TOC. Samples or process upsets exceeding this upper bound are outside the verified linearity and accuracy figures, so upstream dilution or an alternative range must be considered.
Measurement cycles of 5, 10, 15 or 30 minutes are selectable. Shorter cycles give faster updates; longer cycles may be preferred where the process varies slowly. The appropriate choice depends on the required response time of the monitoring loop.
Selection & Integration Q&A
LAN, analog and contact wiring must be 10 m or less. The unit runs on 100–240 V AC, 50/60 Hz at 100 VA. Plan the controller or data logger location within that wiring distance and confirm the AC circuit can supply 100 VA.