Shimadzu USF 2000A
Shimadzu ·Made in Japan · In Active Production

Shimadzu USF 2000A

Shimadzu USF 2000A — Testing machines.

The Shimadzu USF-2000A is an ultrasonic fatigue testing system that uses a piezoelectric element with horn amplification to drive specimens at 20 kHz ±500 Hz, with a recommended test range of 20 kHz ±30 Hz. It applies fully reversed (stress ratio -1) loading via ±10 to ±50 μm horn-tip displacement, with actual stress depending on specimen geometry and material properties. This model is suited for high-cycle fatigue testing of metals such as high-strength steel, duralumin, titanium alloy, and aluminum that resonate at 20 kHz with minimal heat generation.

Full specifications

Description

Construction and positioning

Integration & Operating Notes

Electrical
Verify power availability for three separate voltage feeds

The system requires 3-phase 200 V for the air compressor, 1-phase 200 V for the ultrasonic fatigue testing system, and 1-phase 100 V for the computer and peripherals. Ensure all three supplies are available and properly rated; missing any one will prevent full system operation.

Factory Specification Sheet
Process Limits
Specimen must resonate at 20 kHz with minimal self-heating

The test frequency is derived from the specimen's resonance, so specimens must be designed to resonate within 20 kHz ±500 Hz (recommended ±30 Hz). Materials that generate significant heat during resonance, such as resins, cannot be tested because excessive heating would invalidate results or damage the specimen.

Factory Specification Sheet
Mounting
Specimen attachment is critical for reliable fatigue data

Samples must be securely attached to the ultrasonic resonance system. Materials for which samples are difficult to attach are listed as not testable. Proper specimen geometry and clamping are essential to achieve the required resonance and avoid energy loss or premature failure at the grip.

Factory Specification Sheet
Cooling & Media
Air dryer and compressor are required for cooling and operation

The standard system includes a cooling unit (air dryer), but an air compressor (3-phase 200 V, 2 kVA) is a required optional product that must be supplied separately. The air dryer depends on a continuous compressed air supply; without it, the system cannot maintain proper cooling during extended high-frequency testing.

Factory Specification Sheet
Control & I/O
Software controls resonance tracking and displacement amplitude

The included control computer and ultrasonic testing control and measurement software manage the excitation frequency and horn-tip displacement. The software must maintain resonance as the specimen heats or stiffens; operators should monitor frequency drift and displacement to ensure the test remains within the recommended 20 kHz ±30 Hz range.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the USF-2000A instead of other ultrasonic fatigue systems or conventional servo-hydraulic machines?

Choose the USF-2000A for very high-cycle fatigue testing (10^9 cycles and beyond) of metals that resonate at 20 kHz with minimal heat, such as high-strength steel, duralumin, titanium, and aluminum. Conventional machines are better for low-cycle fatigue or materials that heat significantly at 20 kHz.

What are the facility requirements for installing the USF-2000A, particularly regarding power and air supply?

The system needs 3-phase 200 V (2 kVA) for the air compressor, 1-phase 200 V (3.5 kVA) for the ultrasonic system, and 1-phase 100 V (1 kVA) for the computer and peripherals. An air compressor is a required optional product and must be ordered separately if not available.

What specimen design and material constraints must be considered to avoid invalid tests or equipment damage?

Specimens must be designed to resonate at 20 kHz ±500 Hz (recommended ±30 Hz) and must not generate significant heat during resonance. Materials like resins are not testable. Attachment must be secure; samples difficult to attach are excluded. Stress depends on specimen shape and properties.

What maintenance and consumables are required, and what is not included in the standard delivery?

The standard delivery includes the main unit, ultrasonic resonance system, control computer, software, and cooling unit (air dryer). An air compressor (3-phase 200 V, 2 kVA) is a required optional product and is not included. Regular maintenance includes air dryer servicing and specimen fixture inspection.