Waters 2432 Conductivity Detector for HPLC Systems
Waters ·Made in USA · In Active Production

Waters 2432 Conductivity Detector for HPLC Systems

Waters 2432 Conductivity Detector for HPLC Systems — Chromatography detectors.

The Waters 2432 Conductivity Detector is a HPLC detector using ring-shaped stainless-steel electrodes in a 0.8 µL flow cell, with chemically inert PCTFE and stainless-steel wetted parts. It covers 0 to 10,000 µS/cm without range switching and provides active temperature control from 20 °C to 50 °C in 5 °C steps. It is intended for continuous conductivity monitoring in HPLC systems, delivering low noise (≤0.005 µS/cm) and drift below 0.5%/hr.

Full specifications

Core Performance & Functional Specifications
Detector type Conductivity Detector
Measuring range mfr 0 to 10,000 μS/cm without range switching
Drift mfr <0.5%/hr
Sampling rate mfr 10 points/s
Electrodes mfr Ring-shaped electrodes, made from stainless stee
Cell volume mfr 0.8 μL
Pressure limit mfr 4999 kPa (49,989 mbar, 725 psi)
Wetted materials mfr Chemically inert polychlorotrifluoroethylene (PC
Moisture protection mfr Normal (IPX0)
Line voltages, nominal mfr Grounded ac
Measurement range mfr 0 to 10,000 μS/cm (without range switching)
Drift (alt var) mfr <0.5 %/hr
Maximum backpressure mfr 5000 kPa (50 mbar, 725 psi)
Dimensions, Weight & Mechanical Design
Weight 12.3 kg
Width mfr 34.3 cm
Height mfr 20.8 cm
Depth mfr 61.0 cm
Minimum pulse width mfr 100 ms
Electrical, Power & Drive Specifications
Power consumption 160 VA (nominal)
Line frequency mfr 50 to 60 Hz
Line voltages, nominal mfr 100 to 240 Vac (grounded ac)
Power requirements mfr 100 to 240 Vac
Fuse ratings mfr 3.15 A, 250 V (fast-blow), 5 x 20 mm
Inject-start trigger V max mfr +5 Vdc
Inject-start trigger V min mfr +0 Vdc
Inject-start trigger VIH mfr +2.0 Vdc
Inject-start trigger VIL mfr +0.8 Vdc
Operating Conditions, Environment & Protection
Noise ≤0.005 μS/cm
Audible noise (instrument generated) (alt var) mfr <58 dBA
Operating humidity range mfr 20 to 80 % non-condensing
Shipping and storage humidity range mfr 20 to 85 % non-condensing
Protection class mfr Class I
Overvoltage category mfr II
Pollution degree mfr 2
Noise (alt var) mfr ≤0.005 μS/cm

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Do not exceed 725 psi system pressure

The flow cell is rated to a maximum pressure of 725 psi. Operating above this limit can cause cell rupture or leakage, leading to inaccurate readings and potential safety hazards. Install appropriate pressure relief or monitoring upstream.

Factory Specification Sheet
Process Limits
Temperature control range 20–50 °C in 5 °C steps

The detector can actively control temperature only within 20 °C to 50 °C, selectable in 5 °C increments, or turned off. Ensure your method's required temperature falls within this range and set it accordingly; otherwise, rely on ambient stability.

Factory Specification Sheet
Maintenance
Periodic cleaning of the flow cell is required

The flow cell may accumulate deposits from eluents, affecting noise and drift. Flush with appropriate solvents and follow the manual's cleaning procedure at recommended intervals to maintain performance.

Instruction Manual, Section 5 Maintenance
Electrical
Verify power and signal connections before operation

The detector must be connected to a suitable power source and integrated with the HPLC system's data acquisition unit. Refer to the manual for specific voltage and interface requirements to avoid damage.

Instruction Manual, Section 2 Installation

Selection & Integration Q&A

When should I choose the 2432 Conductivity Detector over other detectors in the Waters HPLC series?

Choose the 2432 when your analytes are ionic and require conductivity detection with a wide dynamic range (0–10,000 µS/cm) and active temperature control. For UV-absorbing compounds, a UV/Vis detector is more appropriate; for trace-level ion analysis, the 2432's low noise and drift are advantageous.

What are the requirements for connecting the 2432 to my HPLC system?

Ensure the system pressure does not exceed 725 psi and that wetted materials (PCTFE, stainless steel) are compatible with your mobile phase. Use low-volume tubing to preserve the 0.8 µL cell volume and minimize band broadening.

What are the environmental and duty-cycle limits for the 2432?

Operate within 20–50 °C if using active temperature control; otherwise, ambient fluctuations may affect stability. The detector is designed for continuous operation but should not exceed 725 psi. Avoid corrosive atmospheres and ensure adequate ventilation.

What maintenance is required and what accessories are needed but not included?

Regularly flush the flow cell to prevent fouling. Replacement parts such as ferrules, tubing, and cleaning solutions are not included and must be ordered separately. Refer to the manual for the recommended maintenance schedule and spare parts list.