Waters 2475 Fluorescence (FLR) Detector for HPLC Systems
Waters 2475 Fluorescence (FLR) Detector for HPLC Systems — Chromatography detectors.
The Waters 2475 Fluorescence (FLR) Detector is a xenon-lamp-based HPLC detector with axially illuminated 13 µL flow cell, covering 200–890 nm excitation and 210–900 nm emission. It provides high sensitivity (Raman S/N ≥1000) and up to four 2D channels or one 3D channel, serving as a selective fluorescence detector in chromatographic systems.
Full specifications
| Detector type | Fluorescence (FLR) |
| Excitation wavelength range mfr | 200 to 890 nm |
| Emission wavelength range mfr | 210 to 900 nm |
| Bandwidth mfr | 20 nm |
| Wavelength accuracy mfr | ±3 nm |
| Wavelength repeatability mfr | ±0.25 nm |
| Sensitivity mfr | S/N, Raman peak of water ≥1000 |
| Measurement range mfr | 0.001 to 100,000.000 emission units |
| Filter time constant, single λ mode mfr | 0.1 to 5.0 s Hamming (default); 0.1 to 99.0 s (o |
| Filter time constant, dual λ mode mfr | 1.0 to 50.0 s Hamming (default); 1.0 to 99.0 s ( |
| Data channels mfr | Up to four 2D channels or one 3D channel |
| Sampling rate mfr | Up to 20 points/s in single mode; 1 point/s in 3 |
| Light source mfr | Xenon lamp, 150 W |
| Flow cell design mfr | Axially illuminated |
| Flow cell volume mfr | 13 μL |
| Pressure limit mfr | 145 psi (10 bar) |
| Wetted materials mfr | 316 stainless-steel, fused silica, Teflon |
| Power requirements mfr | 100 to 240 VAC |
| Inputs mfr | Four event inputs |
| Outputs mfr | Four outputs (2 analog, 2 event) |
| Weight | 18.1 kg |
| Power consumption | 280 VA |
| Line frequency mfr | 50 to 60 Hz |
Description
Construction and positioning
- Axially illuminated flow cell (13 µL) with 316 stainless-steel, fused silica, and Teflon wetted parts. - 150 W xenon lamp with 20 nm bandwidth; wavelength accuracy ±3 nm, repeatability ±0.25 nm. - Standalone benchtop unit (18.1 kg) with 100–240 VAC power, 280 VA consumption. ### Behavior under load and in different modes
- Measurement range 0.001–100,000.000 emission units. - Single-λ mode: filter time constant 0.1–5.0 s Hamming (default) or 0.1–99.0 s optional RC; sampling up to 20 points/s. - Dual-λ mode: filter time constant 1.0–50.0 s Hamming (default) or 1.0–99.0 s optional RC. - 3D mode: sampling limited to 1 point/s; supports one 3D channel. - Flow cell pressure limit 145 psi (10 bar). ### Installation, connection, and control
- Four event inputs and four outputs (2 analog, 2 event) for integration with HPLC systems. - Power input 100–240 VAC, 50–60 Hz; requires 280 VA.
Integration & Operating Notes
The detector draws 280 VA; verify the branch circuit and any UPS can handle this continuous load plus inrush. Operation outside 100–240 VAC or 50–60 Hz may cause malfunction or damage.
The 13 µL flow cell is rated to 145 psi (10 bar). Exceeding this pressure can rupture the cell or cause leaks, invalidating results and risking operator safety. Install a backpressure regulator or pressure relief upstream if system pressures may spike.
The 150 W xenon lamp has finite life; when Raman S/N of water falls below 1000, lamp replacement is indicated. Follow manufacturer's procedure for lamp replacement and recalibration of wavelength accuracy to ±3 nm.
Selection & Integration Q&A
The detector requires 100–240 VAC, 50–60 Hz, 280 VA. It provides four event inputs and four outputs (2 analog, 2 event) for synchronization with pumps, autosamplers, and data systems. Ensure the controller or software supports event-based triggering and analog signal acquisition.
The base delivery does not include cables for event/analog connections, replacement xenon lamps, or flow cell spare parts. A maintenance schedule should include lamp replacement when S/N drops below 1000 and periodic cleaning of the flow cell. Refer to the manual for specific intervals.