Waters 3767 Sample Manager (Injector/Collector)
Waters 3767 Sample Manager (Injector/Collector) — Laboratory automation & equipment.
The Waters 3767 Sample Manager is a dual-mode autosampler for analytical and preparative LC, supporting injection volumes from 20 µL to 30 mL via interchangeable fixed loops. It handles up to 15 microplates or 544 test tubes, with <0.75% RSD injection reproducibility and <0.05% carryover. Its 6000 psi high-pressure side and 100 psi collection side enable high-throughput sample introduction and fraction collection in a single unit.
Full specifications
| Sample capacity (Injection) mfr | Up to 15 microplates (up to 384-wells); Up to 72 |
| Syringe delivery accuracy mfr | >99.0% at full stroke with 1000-μL syringe |
| Injection reproducibility mfr | <0.75 RSD; 20-μL injection volume, 20-μL fixed l |
| Carryover mfr | <0.05% |
| Injection volume (Analytical) mfr | 20 μL standard (fixed loop), (0.150-in valve pas |
| Injection volume (Preparative) mfr | 1 mL standard (fixed loop), (0.024-in valve pass |
| System pressure ratings mfr | 6000 psi high pressure side, including injector; |
| Wash pump mfr | >30 mL/min flow through standard tubing and need |
| Available sample loops mfr | 20 uL (standard analytical); 1.0 mL (standard pr |
| Supported syringes mfr | 1.0 mL (standard); 2.5 mL, 5.0 mL, 12.5 mL |
| Injection cycle time mfr | <30 s based on standard wash cycle |
| Custom rack calibration mfr | Control software |
| Communications mfr | MassLynx v4.2 or higher; Ethernet to PC/workstat |
| Dimensions without hood mfr | Width: 959 mm (37.75 in); Depth: 629 mm (24.75 i |
| Dimensions with hood mfr | Width: 959 mm (37.75 in); Depth: 629 mm (24.75 i |
| Operating temperature mfr | 15 to 40 °C (59 to 104 °F) |
| Fume hood (optional) mfr | 4-in. duct with >1.25 CFM, minimum face velocity |
| Weight | 86.2 kg |
Description
Construction and positioning
- The 3767 is a single autosampler chassis (86.2 kg, 959 × 629 × 1020 mm without hood) that combines analytical injection and preparative fraction collection. Unlike fixed-loop analytical-only samplers, it uses user-replaceable sample loops (20 µL to 30 mL) and syringes (1.0 to 12.5 mL) to cover both high-resolution and milligram-scale workflows. - The high-pressure side (injector valve and needle) is rated to 6000 psi, while the collection side and wash system are limited to 100 psi. This split architecture allows standard analytical columns and preparative columns to be used without re-plumbing. ### Behavior under load and in different modes
- Injection reproducibility is <0.75% RSD (20 µL injection, 20 µL loop, 4× overfill), and carryover is <0.05% with the standard wash. The wash pump delivers >30 mL/min through the needle and standard tubing, using two wash solvents to minimize cross-contamination between samples. - The standard analytical configuration uses a 20 µL fixed loop and 0.150-in. valve passages; the preparative configuration uses a 1 mL loop and 0.024-in. passages. Larger loops (2, 5, 10, 30 mL) are available for scale-up, but the 100 psi collection-side limit must be respected. ### Installation, connection, and control
- The 3767 requires a benchtop footprint of at least 959 × 629 mm and a height clearance of 1020 mm (1060 mm with the optional fume hood). It communicates via Ethernet to a PC running MassLynx v4.2 or higher, and provides TTL (open collector) outputs on the rear panel for peripheral devices such as fraction collectors or valves. - The optional fume hood requires a 4-in. duct with >1.25 CFM and a minimum face velocity of 10 ft/min. Operating temperature must be 15–40 °C. Custom rack calibration is performed through the control software.
Integration & Operating Notes
The high-pressure side (injector) is rated to 6000 psi, but the collection side and syringe/needle wash are limited to 100 psi. Exceeding 100 psi on the collection side can damage the valve or tubing. Use appropriate back-pressure regulation when collecting fractions.
The rear-panel TTL (open collector) outputs are used to trigger fraction collectors or other peripherals. Open-collector outputs require pull-up resistors or compatible input circuitry on the peripheral. Verify voltage and current ratings to avoid missed triggers or damage.
If the optional fume hood is used, connect a 4-in. duct capable of >1.25 CFM and maintain a minimum face velocity of 10 ft/min. Inadequate exhaust may allow solvent vapors to escape into the laboratory, posing health and safety risks.
Selection & Integration Q&A
Choose the 3767 when you need both analytical injection (20 µL loop) and preparative fraction collection (up to 30 mL loops) in one instrument. A dedicated analytical autosampler lacks the 100 psi collection side and large-loop capability.
The 3767 connects to a PC via Ethernet running MassLynx v4.2 or higher. TTL outputs on the rear panel trigger peripherals like fraction collectors. Ensure the MS can accept the 6000 psi high-pressure side and that the collector's input is compatible with open-collector signals.
The base unit includes a 20 µL analytical loop and 1.0 mL syringe. Additional loops (1, 2, 5, 10, 30 mL), syringes (2.5, 5, 12.5 mL), the fume hood, and custom racks are not included. Wash solvents and replacement tubing are consumables. Regular maintenance includes checking wash solvent levels and replacing syringes/loops per the user manual.