Waters ACQUITY Advanced Polymer Chromatography Analysis System
Waters ·Made in USA · In Active Production

Waters ACQUITY Advanced Polymer Chromatography Analysis System

Waters ACQUITY Advanced Polymer Chromatography Analysis System — Chromatographs with Flow rate 2.2 milliliter/minute, Max pressure 103.4 megapascal.

The ACQUITY Advanced Polymer Chromatography (APC) Analysis System is a dedicated ultra-performance liquid chromatography (UPLC) configuration optimized for size-exclusion and other polymer separations. It combines a low-dwell-volume binary solvent manager (≤400 μL total system dwell) with a high-pressure fluidic path (15,000 psi) and a single-zone column manager (4–90 °C), enabling high-resolution polymer analysis at flow rates up to 2.2 mL/min. The system is intended for integration into analytical laboratories performing polymer molecular weight distribution and related characterizations.

Flow rate
2.2 milliliter/minute
Max pressure
103.4 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 2.2 mL/min
Max pressure 15000 psi
Column temperature 90.0 °C
Injection volume 50.0 μL
Dwell volume (total system) mfr ≤400 μL (includes standard 100 μL mixer)
Gradient delay volume mfr ≤300 μL (includes standard 100 μL mixer)
Operating flow range mfr 0.010 to 2.20 mL/min in 0.001 mL increments
Injection volume range mfr 0.5 to 50.0 μL as standard; 0.1 to 250.0 μL with
Sample compartment temperature range mfr 4.0 °C to 40.0 °C
Temperature range (Single Zone Column Manager) mfr Controlled 4.0 °C to 90.0 °C
Temperature range (30 cm Single Zone Column Manager) mfr Settable 20 °C to 90 °C; Controllable (ambient +
Materials, Media & Process Compatibility
pH range mfr 2 to 12

Description

System Architecture and Positioning

Integration & Operating Notes

Process Limits
Do not exceed 15,000 psi system pressure

Operating above the maximum pressure rating of 15,000 psi (103.4 MPa) can cause seal failure, tubing rupture, or column damage. Always verify that the column and flow rate combination does not exceed this limit, especially when using high-viscosity solvents or small-particle columns.

Factory Specification Sheet
Process Limits
Mobile phase pH must remain within 2–12

The system's fluidic path is compatible with pH 2–12. Using mobile phases outside this range may degrade seals, injector components, and column hardware, leading to leaks and unreliable results. Verify pH of aqueous buffers and organic modifiers before use.

Factory Specification Sheet
Control & I/O
Ensure column manager temperature is stable before injection

The column manager must reach and stabilize at the set temperature (4–90 °C) before sample injection to avoid retention time drift. Allow sufficient equilibration time, especially when changing temperatures, and verify with a system suitability test.

Instruction Manual, Section 3.2 Column Manager Operation
Maintenance
Replace pump seals and check valves at recommended intervals

To maintain performance and prevent leaks, replace pump seals and check valves according to the maintenance schedule in the instruction manual (typically every 3–6 months depending on usage). Use only Waters-approved spare parts to ensure proper fit and function.

Instruction Manual, Section 6.1 Maintenance Schedule
Mounting
Install on a level bench with sufficient clearance

The system should be placed on a stable, level laboratory bench capable of supporting its weight (not specified). Allow at least 10 cm clearance around vents for proper ventilation. Avoid direct sunlight and heat sources.

Instruction Manual, Section 2.1 Site Requirements

Selection & Integration Q&A

When should I choose the APC Analysis System over a standard ACQUITY UPLC system for polymer analysis?

Choose the APC system when low dwell volume (≤400 μL) and extended pH range (2–12) are critical for reproducible polymer separations. Standard UPLC systems have higher dwell volumes and may not support pH extremes, leading to method transfer issues.

What are the requirements for connecting a detector to the APC system?

The detector must be compatible with the system's high-pressure fluidics (up to 15,000 psi) and low dwell volume. Use high-pressure fittings rated for 15,000 psi and ensure the detector's flow cell volume does not significantly increase system dwell volume. Consult Waters for approved detectors.

What maintenance items and accessories are not included in the base delivery?

The base system includes standard fittings and a 100 μL mixer. Optional extension loop for 250 μL injections, alternative mixers, replacement pump seals, check valves, and solvent lines are not included. Regular maintenance kits are available separately.