Waters ACQUITY RDa Detector

Waters ACQUITY RDa Detector — Mass spectrometers with Resolution 10000.

The ACQUITY RDa Detector is a high-resolution time-of-flight mass spectrometer with an electrospray ionization source, covering 50–2000 m/z in both positive and negative modes (extended ranges to 9000 and 5000 m/z respectively). It delivers 10,000 FWHM resolution and <2.5 ppm mass accuracy, distinguishing it as a dedicated intact-mass and data-independent acquisition detector within the Waters ACQUITY series. It is intended for integration into UPLC/MS systems for accurate mass confirmation and profiling.

Resolution
10000

Full specifications

Core Performance & Functional Specifications
Resolution 10000 FWHM
Ion source Electrospray Ionization (ESI)
Scan speed 1 to 20 Hz
Vacuum system mfr fully automated and differentially pumped vacuum
Acquisition modes mfr Full scan (MS); Full scan with fragmentation (da
Positive ion MS sensitivity mfr Intensity ≥22,000 ions/counts per second, measur
Negative ion MS sensitivity mfr Intensity ≥13,750 ions/counts per second, measur
Chromatographic dynamic range mfr ≥4 orders of magnitude
Regulatory approvals/marks mfr CE, CB, NRTL (CAN/US), RCM
Software mfr Systems supported in waters_connect Software ver
Dimensions, Weight & Mechanical Design
Mass range Positive ion: 50–2000 or 400–9000 m/z; Negative ion: 50–2000 or 400–5000 m/z
Mass analyzer mfr high resolution, high stability Tof mass analyze
Mass measurement accuracy mfr RMS error <2.5 ppm
Mass scale calibration accuracy mfr <1 ppm with internal Lock Mass correction over r

Description

Construction and architecture

Integration & Operating Notes

Cooling & Media
Air-cooled turbo pump needs unobstructed ventilation

The split-flow turbo pump is air-cooled; installation must provide sufficient clearance and ambient airflow to prevent overheating. The rotary backing pump is oil-free only if that option is specified.

Factory Specification Sheet
Process Limits
Mass range selection must match ion mode and application

Positive ion mode offers 50–2000 or 400–9000 m/z; negative ion mode offers 50–2000 or 400–5000 m/z. Exceeding the selected range or using a range not supported in the chosen polarity will yield no valid data.

Factory Specification Sheet
Maintenance
Lock Mass correction required for <1 ppm calibration accuracy

Mass scale calibration accuracy of <1 ppm is achieved only with internal Lock Mass correction over 50–2000 m/z. Without Lock Mass, accuracy degrades to the RMS <2.5 ppm level; routine verification of Lock Mass reference is necessary.

Factory Specification Sheet

Selection & Integration Q&A

What are the requirements for the LC and software to integrate this detector?

The detector must be controlled by waters_connect Software version 4.0.0. It uses an ESI source; no specific LC coupling requirements are stated, but the system must support the selected acquisition mode (full scan or full scan with fragmentation).

What maintenance and consumables are required, and what is not included in the base delivery?

The vacuum system is fully automated; the rotary backing pump is oil-free only if that option is ordered. No maintenance schedule, cables, or consumables are listed in the supplied facts; Lock Mass reference materials are needed for <1 ppm calibration.