Waters ACQUITY UPLC H-Class PLUS System
Waters ACQUITY UPLC H-Class PLUS System — Chromatographs with Flow rate 2.2 milliliter/minute, Max pressure 103.4 megapascal.
The ACQUITY UPLC H-Class PLUS System is a stainless-steel flow-path UPLC chromatograph with binary or quaternary pumping, active pre-heating, and up to 15,000 psi operation at 2.2 mL/min. It accommodates up to six 150 mm columns and 20 sample plates with the Sample Organizer, positioning it for high-throughput method development and routine analysis. Its open detector architecture supports PDA, TUV, FLR, charged aerosol, RI, ELS, and mass spectrometric detection under Empower, MassLynx, waters_connect, or UNIFI control.
Full specifications
| Flow rate | 2.2 mL/min |
| Max pressure | 15000 psi |
| Sample flow path mfr | Stainless-steel |
| Sample carryover mfr | ≤0.002% (Caffeine) under UV conditions |
| Sample compartment mfr | 4–40 ̊C |
| Pump options mfr | Binary, Quaternary |
| Solvent conditioning mfr | Active Pre-heating |
| Column capacity mfr | Up to 6 columns of 150 mm length and up to 4.6 m |
| Maximum sample capacity mfr | Up to 20 plates (with Sample Organizer) |
| Detector options mfr | PDA Detector, TUV Detector, FLR Detector, Charge |
| External control mfr | Empower Software, MassLynx Software, waters_conn |
| Height | 711 cm |
| Width | 343 cm |
| Depth mfr | 71.2 cm (28.0 in.) |
| pH range mfr | 1–12.5 |
Description
Construction and positioning
- Stainless-steel sample flow path with pH range 1–12.5; sample carryover ≤0.002% (caffeine, UV conditions). - Pump options: binary or quaternary, with active solvent pre-heating for thermal equilibration before the column. - Column capacity: up to 6 columns of 150 mm length and up to 4.6 mm I.D.; sample capacity up to 20 plates with the Sample Organizer. - Sample compartment temperature control: 4–40 °C. - Detector options: PDA, TUV, FLR, charged aerosol, RI, ELS, and mass spectrometers; external control via Empower, MassLynx, waters_connect, or UNIFI. ### Behavior under load and in different modes
- Maximum pressure 15,000 psi (≈103.4 MPa) at a flow rate of 2.2 mL/min; the stainless-steel flow path supports acidic to mildly alkaline mobile phases (pH 1–12.5). - Binary pumping suits fixed-ratio high-pressure gradients; quaternary pumping enables low-pressure mixing of up to four solvents for method scouting. - Active pre-heating reduces thermal mismatch at the column inlet, supporting reproducible retention at elevated flow rates. - Carryover ≤0.002% (caffeine) minimizes ghost peaks in high-sensitivity UV or MS detection. ### Installation, connection, and operation
- The system is controlled externally via Empower, MassLynx, waters_connect, or UNIFI; no standalone control panel is specified. - Detector selection determines the required bench space and data system configuration; mass spectrometers require additional interface and vacuum provisions. - Column capacity and sample organizer capacity affect bench footprint and plate-handling workflow; plan for access to the sample compartment for plate loading.
Integration & Operating Notes
Operating above the 15,000 psi maximum risks seal, tubing, and column hardware failure; method gradients and column dimensions must be selected so that backpressure stays below this limit at the intended 2.2 mL/min flow.
The stainless-steel flow path tolerates pH 1–12.5; outside this range, corrosion and metal leaching can occur, compromising both hardware life and chromatographic results, especially with MS detection.
The system is controlled via Empower, MassLynx, waters_connect, or UNIFI; no standalone control is specified. Integration planning must include the data-system workstation, network, and license before the instrument can run methods.
Sample stability and injection reproducibility depend on holding the sample compartment within its 4–40 °C range; lab ambient conditions and plate residence time should be checked for thermally labile samples.
With up to six 150 mm columns and up to 20 plates with the Sample Organizer, routine column changes and plate loading require unobstructed front and side access; layout the bench so these tasks do not disturb detector or MS connections.
Selection & Integration Q&A
Choose this model when you need up to 15,000 psi at 2.2 mL/min, pH 1–12.5, and flexibility across binary or quaternary pumping with active pre-heating. A binary-only variant suits fixed high-pressure gradients; this system adds quaternary mixing and broad detector options for method development.
The system requires an external control platform (Empower, MassLynx, waters_connect, or UNIFI) plus the selected detector (PDA, TUV, FLR, charged aerosol, RI, ELS, or MS). Confirm bench space for up to six 150 mm columns and, if used, the Sample Organizer for up to 20 plates.
Operate within 15,000 psi maximum pressure and pH 1–12.5; keep the sample compartment at 4–40 °C. The stainless-steel flow path and active pre-heating require stable lab ambient conditions. No PDF manual was supplied, so confirm duty-cycle and ventilation requirements with Waters.
Dimensional Drawings & Technical Documentation
Dimensional drawings
Click drawing to inspect vector dimensions