Waters Alliance HPLC System
Waters ·Made in USA · In Active Production

Waters Alliance HPLC System

Waters Alliance HPLC System — Chromatographs with Flow rate 10 milliliter/minute, Max pressure 34.47 megapascal.

The Alliance HPLC System is a quaternary, low-pressure-gradient HPLC platform with vacuum degassing and Auto-Blend technology for up to four mobile phases. It delivers flow rates to 10 mL/min against 5000 psi (34.5 MPa) and supports injection volumes from 0.1 to 100 µL standard, with an optional 2 mL loop. It serves as the separation and delivery core in analytical HPLC workflows, compatible with a broad range of detectors and controlled via Empower or MassLynx software.

Flow rate
10 milliliter/minute
Max pressure
34.47 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 10 mL/min
Max pressure 5000 psi
Column temperature 4–40 °C
Injection volume 100 μL
Mobile phases mfr Up to 4 with Auto-Blend Technology
Sample carryover mfr ≤0.0025% (Caffeine) under specified conditions
Sample compartment mfr 4–40 °C
Pump options mfr Quaternary
Solvent conditioning mfr Vacuum Degas
Column capacity mfr Standard up to 3 columns
Maximum sample capacity mfr 120
Detector options mfr Tunable UV, Photo Diode Aray, Fluorescence, Char
External control mfr Empower Software, MassLynx Software
Injection volume range mfr 0.1 to 100 μL standard, up to 2 mL optional
Dimensions, Weight & Mechanical Design
Height 571 cm
Width 457 cm
Depth mfr 57.1 cm (22.5 in.) or 64.8 cm (25.5 in.) with sa
Width with column heater mfr 58.4 cm (23.0 in.)
Interfaces, Connections & Control
Front panel display mfr Easy to use control panel

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Do not exceed 5000 psi system pressure

Operating above 5000 psi (34.5 MPa) risks seal damage, column failure, and invalid results. The pump and plumbing are rated for this maximum; ensure method pressures remain below this limit, especially with small-particle columns or high flow rates.

Factory Specification Sheet
Process Limits
Column temperature must stay within 4–40 °C

The column heater/cooler maintains 4–40 °C. Exceeding this range may cause inaccurate temperature control, affecting retention times and separation. Verify that the setpoint and ambient conditions allow the module to hold the required temperature.

Factory Specification Sheet
Control & I/O
Software control required for automated sequences

Empower or MassLynx software is required for automated sample injection, gradient programming, and data acquisition. The front panel supports manual control but not full automation. Ensure the control software is installed and licensed before attempting automated runs.

Factory Specification Sheet
Mounting
Allow extra bench depth with sample heater/cooler

The system depth is 57.1 cm (22.5 in.) without the sample heater/cooler and 64.8 cm (25.5 in.) with it. Ensure the bench or stack accommodates the deeper configuration and allows adequate ventilation and access for service.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the Alliance HPLC System instead of a binary or isocratic variant?

Choose the Alliance when method development or multi-solvent applications require quaternary low-pressure gradients with up to four mobile phases and Auto-Blend. For routine isocratic assays, a simpler binary or isocratic pump may suffice and reduce complexity.

What detector and software compatibility should I verify before integrating the Alliance into my lab?

The Alliance supports a wide range of detectors including UV, PDA, fluorescence, CAD, RI, ELSD, electrochemical, conductivity, and MS. Ensure your chosen detector is compatible with the system's flow path and that Empower or MassLynx software is available for control and data handling.

What maintenance items and accessories are not included in the base delivery?

The base system typically includes the pump, autosampler, column heater, and necessary tubing, but detectors, columns, and software licenses are usually ordered separately. Consumables such as pump seals, degasser membranes, and sample vials are not included and should be stocked per the maintenance schedule.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation