Waters Alliance iS Bio HPLC System
Waters Alliance iS Bio HPLC System — Chromatographs with Flow rate 10 milliliter/minute, Max pressure 82.74 megapascal.
The Alliance iS Bio HPLC System is a quaternary, inert flow-path HPLC system with a flow-through needle and MaxPeak HPS technology, supporting up to 10.0 mL/min and 12,000 psi. It accommodates one column up to 300 mm length and 8 mm I.D., with sample capacity for any three of six vial/plate formats. It is intended for bio-separation methods requiring pH 1–13 compatibility and low carryover (≤0.002% caffeine).
Full specifications
| Flow rate | 10.0 mL/min |
| Max pressure | 12000 psi |
| Injection type mfr | Flow-through needle |
| Sample flow path mfr | Inert (MaxPeak HPS Technology) |
| Sample carryover mfr | ≤0.002% [Caffeine] under UV conditions |
| Sample compartment mfr | 4 to 40 °C |
| Pump options mfr | Quaternary |
| Solvent conditioning mfr | Passive pre-heating |
| Column tracking mfr | Yes, with RFID enabled eConnect Waters Columns |
| Column capacity mfr | 1 column up to 300 mm length, up to 8 mm interna |
| Maximum sample capacity mfr | (any three of the following) 48-position, 2.00-m |
| Integrated Detector Options mfr | TUV Detector, PDA Detector |
| External Detectors Supported mfr | CAD, ACQUITY QDa II, ELS, RI, FLR, MALS |
| Informatics compatibility mfr | Empower Chromatography Data System (CDS) (Featur |
| Height | 742 mm |
| Weight | 72.5 kg |
| Width | 499 mm |
| Depth mfr | 63.5 cm (24.99 in.) |
| pH range mfr | 1 to 13 |
Description
Construction and positioning
- Integrated quaternary pump with passive pre-heating and flow-through needle injector; all wetted sample-path components are inert (MaxPeak HPS). - Single-column capacity (up to 300 mm × 8 mm I.D.) with RFID-based eConnect column tracking. ### Behavior under load and media
- Maximum operating pressure 12,000 psi (≈82.7 MPa) and flow rate 10.0 mL/min define the hydraulic envelope. - pH range 1–13 supports acidic to mildly alkaline mobile phases; sample carryover ≤0.002% (caffeine, UV) reduces ghost-peak risk in trace analysis. - Sample compartment holds 4–40 °C; no column oven or gradient-curve data are provided. ### Installation, control and operation
- Integrated detector options: TUV or PDA; external detectors supported include CAD, ACQUITY QDa II, ELS, RI, FLR, and MALS. - Informatics compatibility: Empower CDS Feature Release 4 with Windows 10 and later. - Physical envelope: 742 mm height, 499 mm width, 635 mm depth, 72.5 kg. No electrical, connection, or manual-derived operating details are available in the provided facts.
Integration & Operating Notes
Operating above the 12,000 psi (≈82.7 MPa) limit risks seal, fitting, or column hardware failure and invalidates results. Verify backpressure at the maximum method flow (10.0 mL/min) with the installed column I.D. and length before routine use.
The inert MaxPeak HPS flow path is rated for pH 1–13. Methods outside this window can degrade wetted materials and produce metal-ion interference or leaks; confirm buffer pH before priming the quaternary pump.
The system is controlled through Empower Chromatography Data System Feature Release 4 on Windows 10 or later. Older CDS releases or unsupported operating systems will not provide instrument control or data acquisition; plan the informatics upgrade before installation.
Selection & Integration Q&A
Integrated TUV or PDA detectors are available; external CAD, ACQUITY QDa II, ELS, RI, FLR, and MALS detectors are supported. Control and data acquisition require Empower CDS Feature Release 4 on Windows 10 or later.
Column tracking requires RFID-enabled eConnect Waters Columns; non-RFID columns need a manual log. No maintenance schedule, service intervals, cables, or consumables are specified in the provided facts; confirm the base-delivery scope with the vendor.