Waters Alliance iS HPLC System - Analysis for Quality Control Labs
Waters Alliance iS HPLC System - Analysis for Quality Control Labs — Chromatographs with Flow rate 10 milliliter/minute, Max pressure 82.74 megapascal.
The Alliance iS HPLC System is an integrated quaternary HPLC platform with flow-through needle injection and stainless-steel flow path, supporting up to 10.0 mL/min and 12,000 psi. It distinguishes itself within the Alliance series by its integrated TUV or PDA detector options and RFID-enabled column tracking, targeting quality control laboratories that require high-throughput, reliable analysis with minimal carryover (≤0.002% caffeine).
Full specifications
| Flow rate | 10.0 mL/min |
| Max pressure | 12000 psi |
| Column temperature | 40 °C |
| Injection type mfr | Flow-through needle |
| Sample flow path mfr | Stainless steel |
| Sample carryover mfr | ≤0.002% [Caffeine] under UV conditions |
| Sample compartment mfr | 4 to 40 °C |
| Pump options mfr | Quaternary |
| Solvent conditioning mfr | Passive pre-heating |
| Column tracking mfr | Yes, with RFID enabled eConnect Waters Columns |
| Column capacity mfr | 1 column up to 300 mm length, up to 8 mm interna |
| Maximum sample capacity mfr | (any three of the following) 48-position, 2.00-m |
| Integrated Detector Options mfr | TUV Detector, PDA Detector |
| External Detectors Supported mfr | CAD, ACQUITY QDa II, ELS, RI, FLR, MALS |
| Informatics compatibility mfr | Empower Chromatography Data System (CDS) (Featur |
| Height | 742 mm |
| Weight | 72.5 kg |
| Width | 499 mm |
| Depth mfr | 63.5 cm (24.99 in.) |
| pH range mfr | 1 to 13 |
Description
Construction and Positioning
- Integrated HPLC system with quaternary pump, flow-through needle injector, and column compartment (4–40 °C) in a single chassis (742 mm H × 499 mm W × 635 mm D, 72.5 kg). - Supports one column up to 300 mm length and 8 mm I.D., with RFID-enabled eConnect column tracking for method compliance. - Sample compartment temperature control (4–40 °C) and multiple sample formats: any three of 48/54-position vial holders, 96/384-well plates, or micro-centrifuge plates. - Integrated detector options: TUV or PDA; external detectors supported include CAD, ACQUITY QDa II, ELS, RI, FLR, MALS. ### Behavior Under Load
- Quaternary pump delivers up to 10.0 mL/min at pressures up to 12,000 psi (≈827 bar), suitable for sub-2 µm particle columns and high-resolution separations. - Passive pre-heating of solvents and column temperature control (up to 40 °C) ensure retention time stability. - Flow-through needle design with stainless-steel flow path achieves sample carryover ≤0.002% (caffeine, UV), critical for trace analysis. - pH range 1–13 accommodates a wide variety of mobile phases. ### Installation, Control, and Operation
- Controlled via Empower CDS (Feature Release 4, Windows 10 or later); no standalone control panel described. - Requires connection to external detectors and informatics network; integrated detector options simplify setup. - Column tracking via RFID requires eConnect columns; sample capacity is configurable at order time (three plates/holders).
Integration & Operating Notes
The system is controlled exclusively via Empower Chromatography Data System (Feature Release 4) running on Windows 10 or later. Ensure the PC meets Empower specifications and that network infrastructure supports instrument control and data transfer.
The column compartment accepts only one column up to 300 mm length and 8 mm internal diameter. Using larger columns may cause fit issues or performance degradation. Verify column dimensions before installation.
Selection & Integration Q&A
Choose integrated TUV for routine QC assays with fixed wavelengths and simplicity. Choose integrated PDA for method development or peak purity. For specialized detection (e.g., mass spec, fluorescence), select a configuration without integrated detector and connect an external detector via the supported interfaces.
The system supports external detectors including CAD, ACQUITY QDa II, ELS, RI, FLR, and MALS. Connection requires appropriate interface cables and configuration in Empower CDS. Ensure the detector's output range and communication protocol are compatible with the system's I/O.
The base delivery includes the system and standard accessories. Consumables such as pump seals, injector needles, and detector lamps are not included. Waters offers preventive maintenance kits; refer to the instruction manual for part numbers and replacement intervals.