Waters Arc Premier System for HPLC/UHPLC Applications
Waters Arc Premier System for HPLC/UHPLC Applications — Chromatographs with Flow rate 5 milliliter/minute, Max pressure 65.5 megapascal.
The Arc Premier System is an HPLC/UHPLC chromatograph with an inert sample flow path (MaxPeak HPS Technology) and a binary or quaternary pump, supporting up to 9500 psi and 5 mL/min. It distinguishes itself within the Arc series by its active solvent pre-heating and eCord column tracking, enabling precise, reproducible separations. The system is designed for method development and routine analysis in pharmaceutical, environmental, and food testing laboratories.
Full specifications
| Flow rate | 5 mL/min |
| Max pressure | 9500 psi |
| Column temperature | 90 °C |
| Injection volume | 1 mL |
| Sample flow path mfr | Inert (MaxPeak HPS Technology) |
| Sample carryover mfr | <0.002% (Caffeine) under UV conditions |
| Sample compartment mfr | 4 to 40 °C |
| Pump options mfr | Binary, Quaternary |
| Solvent conditioning mfr | Active pre-heating; Passive pre-heating |
| Column tracking mfr | eCord Technology column information tracks and a |
| Column capacity mfr | Up to 6 columns of 150 mm length and up to 4.6 m |
| Maximum sample capacity mfr | Up to 20 plates (with Sample Organizer) |
| Detector options mfr | TUV Detector; PDA Detector; FLR Detector; Charge |
| External control mfr | Empower Software; MassLynx Software |
| Injection volume range mfr | 0.1 μL up to 1 mL |
| Column temperature range mfr | 4 °C to 90 °C |
| Height | 71.1 cm |
| Width | 34.3 cm |
| Depth mfr | 71.2 cm (28.0 in.) |
| Column length range mfr | 50 mm to 300 mm |
| pH range mfr | 1 to 12.5 |
Description
Construction and Positioning
The Arc Premier System is a modular HPLC/UHPLC instrument featuring an inert sample flow path constructed with MaxPeak HPS Technology, which minimizes analyte adsorption and carryover (measured at <0.002% for caffeine under UV conditions). The system is available with either a binary or quaternary pump, allowing flexibility in gradient formation. It accommodates up to six columns of 150 mm length and 4.6 mm internal diameter, and the column compartment supports a temperature range from 4 °C to 90 °C. The system dimensions are 71.1 cm (height) × 34.3 cm (width) × 71.2 cm (depth). ### Behavior Under Load and in Different Modes
The pump delivers flow rates up to 5 mL/min and can generate pressures up to 9500 psi (approximately 65.5 MPa), suitable for UHPLC separations with small-particle columns. The injection volume range spans from 0.1 µL to 1 mL, covering both analytical and semi-preparative injections. The sample compartment maintains a temperature between 4 °C and 40 °C, ensuring sample stability. The system supports active or passive solvent pre-heating, which improves retention time reproducibility by pre-conditioning the mobile phase before it reaches the column. The inert flow path is compatible with mobile phases from pH 1 to 12.5, reducing the risk of corrosion and interaction with sensitive analytes. ### Installation, Connection, and Control
The Arc Premier System is controlled via Empower or MassLynx software, providing seamless integration into laboratory data systems. The system includes eCord Technology, which tracks and archives column usage information, aiding in method transfer and column lifetime management. Detector options include TUV, PDA, FLR, Charged Aerosol, RI, and ELS detectors, allowing configuration for a wide range of detection needs. The system requires a stable laboratory environment; the sample compartment temperature control (4–40 °C) and column oven (4–90 °C) demand adequate ventilation and power supply. Installation should account for the system's footprint and access to the sample organizer, which can hold up to 20 plates.
Integration & Operating Notes
The Arc Premier System, when configured with multiple detectors and a column oven, can draw significant current. Ensure the laboratory supply provides a dedicated 15 A circuit to avoid voltage drops that may affect pump precision and detector baseline stability. Refer to the site preparation guide for exact power requirements.
Operating outside the pH range of 1 to 12.5 can damage the inert flow path components and the column. Use appropriate buffers and rinse the system thoroughly when switching between extreme pH values to prevent precipitation and corrosion.
Opening the column oven door during a run triggers an interlock that pauses the pump and stops data acquisition to protect the column from temperature fluctuations. Ensure the door is closed and the oven has reached setpoint before starting a sequence.
Under typical UHPLC conditions (pressures above 8000 psi), pump seals should be replaced every six months to prevent leaks and flow inaccuracies. The eCord column tracking can also log column usage, but pump maintenance is scheduled separately based on operating hours.
The system, when fully configured with detectors and sample organizer, can weigh over 100 kg. The bench must be level and capable of supporting this load without vibration, as vibrations can affect detector performance and retention time reproducibility.
Selection & Integration Q&A
Choose the binary pump for high-pressure UHPLC applications requiring precise gradients and maximum pressure (9500 psi). The quaternary pump is better for low-pressure gradient methods and higher flow rates up to 5 mL/min, but may not reach the same pressure limits. Select based on method requirements and solvent compatibility.
The system can be controlled by MassLynx software and configured with an appropriate detector interface. Ensure the flow path is compatible with the MS ionization source, and use the inert flow path to minimize metal adducts. Consult the MS manufacturer's guidelines for flow rate and pressure limits.
The base system does not include consumables such as pump seals, check valves, column filters, sample vials, and solvent lines. Additionally, the eCord column tracking requires eCord-enabled columns, which are purchased separately. A preventive maintenance kit is recommended for annual service.