Waters Charged Aerosol Detector
Waters ·Made in USA · In Active Production

Waters Charged Aerosol Detector

Waters Charged Aerosol Detector — Chromatography detectors.

The Waters Charged Aerosol Detector is a universal HPLC detector that nebulizes the column effluent, evaporates the mobile phase, and measures the charge imparted to analyte particles. It supports flow rates down to 0.01 mL/min and evaporation temperatures from ambient to 100 °C, with four independent channels for parallel detection. It is used for sensitive, gradient-compatible detection of non-volatile and semi-volatile compounds lacking chromophores.

Full specifications

Core Performance & Functional Specifications
Detector type Charged Aerosol Detector
Flow rate 0.01 mL/min
Evaporation temperature mfr Ambient (off) or settable controlled range: 10 °
Sampling rate mfr Up to 200 Hz
Biocompatible mfr No
Inert gas supply mfr Nitrogen (recommended) or compressed air
Inlet pressure mfr 4.83–5.52 bar (483–552 kPa, 70.0–80.0 psi)
Gas consumption mfr Typically ≤ 5 L/min
Diagnostic channels mfr Nebulizer: Factory-optimized gas flow gas pressu
Dynamic range mfr Four individual power value
Power requirements mfr 100 to 240 VAC
Dimensions mfr Width: 34.3 cm (13.5 in); Height: 20.8 cm (8.2 i
Dimensions, Weight & Mechanical Design
Number of channels mfr Four (4) supported by four (4) independent backg

Description

Construction and positioning

Integration & Operating Notes

Process Limits
Evaporation temperature must be at least 5 °C above ambient

Setting the evaporator below ambient+5 °C risks incomplete evaporation, leading to baseline noise and contamination. Always verify ambient temperature and set the evaporator accordingly, or use the ambient (off) mode if the mobile phase is sufficiently volatile.

Factory Specification Sheet
Process Media
Gas supply pressure must be regulated to 70–80 psi

The nebulizer requires a stable inlet pressure of 4.83–5.52 bar (70–80 psi). Exceeding this range can damage the nebulizer; falling below it causes poor aerosol formation and unreliable results. Use a two-stage regulator and ensure the gas is dry and oil-free.

Factory Specification Sheet
Maintenance
Regular nebulizer and evaporator cleaning is essential

The nebulizer and evaporation chamber are prone to buildup from non-volatile residues. Periodic cleaning (frequency depends on usage) prevents clogging and baseline drift. Refer to the operator's manual for cleaning procedures and recommended intervals.

Instruction Manual (general maintenance section

Selection & Integration Q&A

When should I choose this detector over a UV or fluorescence detector?

Choose the charged aerosol detector for analytes lacking chromophores or fluorophores, or for universal detection across a wide polarity range. UV/fluorescence detectors offer higher sensitivity for specific compound classes but require appropriate functional groups.

What maintenance items and accessories are not included in the base delivery?

The base unit does not include gas regulators, tubing, or filters; these must be sourced separately. A preventive maintenance kit (nebulizer, filters, seals) is recommended but not included. Refer to the manual for the full list of consumables.

Dimensional Drawings & Technical Documentation

Technical documentation