Waters LC Prep 150 System for Sample Purification

Waters LC Prep 150 System for Sample Purification — Chromatographs with Flow rate 150 milliliter/minute, Max pressure 34.5 megapascal.

The Waters LC Prep 150 System is a preparative HPLC platform configured for sample purification, featuring a 150 mL/min pump, 345 bar (5000 psi) maximum operating pressure, and a 10 mL sample loop. It supports columns from 10 to 50 mm I.D. and offers up to four solvents with 11 gradient curves, distinguishing it as a dedicated purification instrument within the Waters preparative series. Its primary purpose is to deliver reproducible, scalable separations for milligram-to-gram scale purification in pharmaceutical and biochemical workflows.

Flow rate
150 milliliter/minute
Max pressure
34.5 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 150 mL/min
Max pressure 345 bar
Injection volume 10000 μL
Operating pressure range mfr 5000 psi (345 bar)
Supported column dimensions mfr 10, 19, 30, and 50 mm I.D.
Software mfr ChromScope
Dwell volume mfr ≤350 μL; <6.5 mL
Flow accuracy mfr ±1.0% of required flow or 10 μL/min, whichever i
Composition accuracy mfr ±1.0% of required flow or 5% to 95%, and from 1
Composition precision mfr Retention time variation <0.3% RSD or 0.1 min SD
Number of solvents mfr Up to four: A1 or A2, B1 or B2
Gradient composition curves mfr Eleven (11) gradient curves including linear (1)
Compressibility compensation mfr Automatic and continuous
Pump seal wash mfr Integral, active, programmable
Priming mfr Self-prime from solvent container on floor with
Collection capacity mfr Standard test tube 12.0-18.0 O.D. rack Max. 162;
Injector mfr 3725i continuous flow design (MBB)
Injectors mfr 7725i analytical continuous flow design (MMB); 3
Sample loop volume mfr 10 mL sample loop
Sample capacity mfr Two 12-vial trays for 10 mL vials
Syringe size mfr 2.5 mL
Dimensions, Weight & Mechanical Design
Height 64 cm
Weight 60 kg
Width 71 cm
Electrical, Power & Drive Specifications
Voltage mfr 100 to 240 VAC
Materials, Media & Process Compatibility
Wetted material mfr 316 stainless steel, UHMWPE, sapphire, ruby, Tef

Description

Construction and Positioning

Integration & Operating Notes

Process Limits
Do not exceed 345 bar (5000 psi) system pressure

Operating above the maximum pressure rating can damage pump seals, column hardware, and tubing. The pressure limit applies to the entire flow path, including the injector and column. Ensure the column and connections are rated for at least 345 bar, and monitor pressure during method development to avoid overpressure trips.

Factory Specification Sheet
Maintenance
Regularly prime and flush the pump seal wash system

The integral pump seal wash is programmable and active, but it requires periodic replenishment of wash solvent and verification of flow. Neglecting the seal wash can lead to premature seal wear, especially when using buffers or aggressive solvents. Follow the recommended maintenance schedule in the user manual.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the LC Prep 150 over other preparative HPLC systems in the Waters series?

Choose the LC Prep 150 when you need flow rates up to 150 mL/min and pressures up to 345 bar for milligram-to-gram purification. Its 10–50 mm column compatibility and 10 mL injection volume suit scale-up from analytical methods. For lower flow rates or higher resolution, consider analytical siblings.

What are the requirements for connecting the LC Prep 150 to a fraction collector and detector?

The system integrates with fraction collectors via ChromScope software, which controls collection based on time or peak detection. Ensure the collector supports the flow rates and volumes used. Detector connections require analog or digital I/O compatible with the system's communication ports.

What maintenance items and accessories are not included in the base delivery?

The base system includes the pump, injector, and software, but consumables such as columns, solvent filters, and sample loops are not included. Fraction collection racks for test tubes, scintillation vials, and microtiter plates are optional. Regular maintenance kits for pump seals and check valves are available separately.