Waters LC Prep AutoPurification System for Liquid Chromatography
Waters ·Made in USA · In Active Production

Waters LC Prep AutoPurification System for Liquid Chromatography

Waters LC Prep AutoPurification System for Liquid Chromatography — Chromatographs with Flow rate 150 milliliter/minute, Max pressure 41.37 megapascal.

The Waters LC Prep AutoPurification System is a preparative-scale liquid chromatography instrument configured with a binary high-pressure gradient pump, delivering up to 150 mL/min at pressures to 6000 psi. Its 0.35 mL delay volume and 20 µL analytical injection loop (1.0 mL preparative loop standard) distinguish it as a dedicated purification platform within the Waters LC Prep series. It is intended for automated fraction collection and compound isolation in pharmaceutical, biotech, and chemical research workflows.

Flow rate
150 milliliter/minute
Max pressure
41.37 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 150.0 mL/min
Max pressure 6000 psi
Injection volume 20 µL
Delay volume mfr 0.35 mL
Number of solvents mfr Up to 4: A1 or A2 and B1 or B2
Size (W x H x D, cm) mfr 28 x 43.2 x 66
External Control mfr MassLynx Software with FractionLynx
Detectors mfr Ultraviolet/visible (UV/Vis) absorbance; Evapora
Injector volume (fixed sample loop) mfr Analytical valve: 20 μL loop standard; Preparati
Syringe sizes mfr 1.0 mL (standard), 2.5 mL, 5 mL, and 12.5 mL
Dimensions, Weight & Mechanical Design
Height 43.2 cm
Weight 39.5 kg
Width 28 cm

Description

Construction and Architecture

Integration & Operating Notes

Process Limits
Do not exceed 6000 psi system pressure

Operating above the maximum pressure rating of 6000 psi (≈41.4 MPa) can damage pump seals, tubing, and fittings, and may cause leaks or unsafe operation. Always verify that the column and plumbing are rated for the intended pressure and set the pump's upper pressure limit accordingly in the software.

Factory Specification Sheet
Control & I/O
FractionLynx controls valve switching and collection

The preparative and analytical valves are switched automatically by FractionLynx based on the run method. Ensure that the method specifies the correct valve position and that the injector is configured for the installed loop size (20 µL analytical or 1.0 mL preparative) to avoid sample loss or overloading.

MassLynx/FractionLynx Software User Guide
Maintenance
Flush the system after buffers and before storage

After using buffer-containing mobile phases, flush the entire flow path (pump, injector, valves, and tubing) with HPLC-grade water followed by an appropriate organic solvent (e.g., methanol or acetonitrile) to prevent salt precipitation and microbial growth. This is especially critical for the preparative valve and loop, which have small internal passages.

Instruction Manual, Section on Maintenance and Storage

Selection & Integration Q&A

When should I choose this LC Prep AutoPurification System instead of a sibling model in the Waters LC Prep series?

Choose this model when you need automated preparative purification with flow rates up to 150 mL/min and pressures to 6000 psi, plus flexibility for up to four solvents and multiple detectors (UV/Vis, ELS, PDA, MS). Sibling models may offer different flow or pressure ranges; select based on your required scale and detection needs.

What are the requirements for connecting a detector to this system?

The system supports UV/Vis absorbance, evaporative light scattering (ELS), photodiode array (PDA), and mass spectrometry (MS) detectors. Ensure the detector's flow cell is rated for the system's maximum pressure (6000 psi) and that its communication interface (e.g., Ethernet or analog) is compatible with MassLynx/FractionLynx control.

What are the environmental and duty-cycle limits for operating this system?

Operate in a temperature-controlled laboratory (typically 15–30 °C) with <80% relative humidity and away from direct sunlight and vibration. The system is designed for continuous duty but requires periodic maintenance (e.g., seal replacement) as specified in the manual; duty cycle limits are not explicitly stated in the provided facts.

What maintenance items and accessories are not included in the base delivery?

The base system includes the pump, autosampler, fraction collector, and standard 20 µL analytical loop and 1.0 mL preparative loop. Detectors, columns, solvent reservoirs, waste containers, and additional syringes (2.5 mL, 5 mL, 12.5 mL) are not included and must be ordered separately. Routine maintenance parts such as pump seals and check valves are also consumables.