Data Sheet Model FD S 56
Technical Reference Summary
Covered Models & Document Scope
Model FD-S 56 drying and heating chamber with forced convection, interior volume 55 L. Two standard variants: FDS056-230V (Art.-No. 9090-0018, 230 V 1~ 50/60 Hz, CEE 7/7 plug) and FDS056UL-120V (Art.-No. 9090-0019, 120 V 1~ 60 Hz, NEMA 5-15 plug).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | +10 °C above ambient temperature to 250 °C p.2 ↗ | FDS056-230V |
| Temperature uniformity at 150°C | 2.6 ± K p.2 ↗ | FDS056-230V |
| Temperature fluctuation at 150°C | 0.4 ± K p.2 ↗ | FDS056-230V |
| Heating up time to 150°C | 15 min p.2 ↗ | FDS056-230V |
| Recovery time after door was opened for 30 s at 150°C | 5 min p.2 ↗ | FDS056-230V |
| Air change rate at 100°C | 4 x/h p.2 ↗ | FDS056-230V |
| Rated voltage | 230 V p.2 ↗ | FDS056-230V |
| Power frequency | 50/60 Hz p.2 ↗ | FDS056-230V |
| Nominal power | 1,1 kW p.2 ↗ | FDS056-230V |
| Unit fuse | 6,3 A p.2 ↗ | FDS056-230V |
| Phase (nominal voltage) | 1~ p.2 ↗ | FDS056-230V |
| Interior volume | 55 L p.2 ↗ | FDS056-230V |
| Net weight of the unit (empty) | 35 kg p.2 ↗ | FDS056-230V |
| Load per rack | 15 kg p.2 ↗ | FDS056-230V |
| Permitted load | 45 kg p.2 ↗ | FDS056-230V |
| Wall clearance back | 160 mm p.2 ↗ | FDS056-230V |
| Wall clearance sidewise | 100 mm p.2 ↗ | FDS056-230V |
| Width net | 614 mm p.2 ↗ | FDS056-230V |
| Height net | 625 mm p.2 ↗ | FDS056-230V |
| Depth net | 577 mm p.2 ↗ | FDS056-230V |
| Interior width | 400 mm p.3 ↗ | FDS056-230V |
| Interior height | 440 mm p.3 ↗ | FDS056-230V |
| Interior depth | 310 mm p.3 ↗ | FDS056-230V |
| Sound-pressure level | 43 dB(A) p.3 ↗ | FDS056-230V |
| Energy consumption at 150°C | 290 Wh/h p.3 ↗ | FDS056-230V |
Operating Limits
- Temperature range: +10 °C above ambient temperature to 250 °C p.2 ↗
- Wall clearance back: 160 mm p.2 ↗
- Wall clearance sidewise: 100 mm p.2 ↗
- Load per rack: 15 kg p.2 ↗
- Permitted load: 45 kg p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-3 p.1 ↗ | FDS | Product family: drying and heating chamber, Solid.Line |
| 4-6 p.1 ↗ | 056 | Model size / interior volume class (55 L) |
| 7-9 p.1 ↗ | UL | UL variant for 120 V / 60 Hz mains (absent for 230 V variant) |
| 10-13 p.1 ↗ | 230V / 120V | Rated voltage variant |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| FDS056-230V p.1 ↗ | 9090-0018 | 230 V 1~ 50/60 Hz, CEE 7/7 plug, standard version | Standard model | |
| FDS056UL-120V p.1 ↗ | 9090-0019 | 120 V 1~ 60 Hz, NEMA 5-15 plug, standard version | UL standard model |