Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model VDL 23

Technical Reference Summary

Covered Models & Document Scope

BINDER VDL 23 vacuum drying chamber for flammable solvents, 24 L interior volume. Two voltage variants: VDL023-230V (Art.-No. 9630-0009, 200...230 V 1~ 50/60 Hz, CEE 7/7 plug) and VDL023-120V (Art.-No. 9630-0013, 100...120 V 1~ 50/60 Hz, NEMA 5-15 plug). ATEX classification EX II 2/3/- G IIB T3 Gb/Gc/- X.

Technical Data

Parameter Value Condition
Temperature range +9 °C above ambient temperature to 110 °C p.2 ↗ VDL 23
Temperature uniformity at 100°C 1.0 ± K p.2 ↗ VDL 23
Temperature fluctuation at 100°C 0.1 ± K p.2 ↗ VDL 23
Heating up time to 100°C 110 min p.2 ↗ VDL 23
Leak rate 0,01 bar/h p.2 ↗ VDL 23
Pressure fluctuation over time at 50 mbar 1 ± mbar p.2 ↗ VDL 23
Pressure fluctuation over time at 400 mbar 70 ± mbar p.2 ↗ VDL 23
Rated voltage 200...230 V p.2 ↗ VDL023-230V
Rated voltage 100...120 V p.2 ↗ VDL023-120V
Power frequency 50/60 Hz p.2 ↗ VDL 23
Nominal power 0,9 kW p.2 ↗ VDL 23
Unit fuse 6,3 A p.2 ↗ VDL023-230V
Unit fuse 10,0 A p.2 ↗ VDL023-120V
Phase (nominal voltage) 1~ p.2 ↗ VDL 23
Vacuum connection with small flange 16 DN mm p.2 ↗ VDL 23
Measuring access port with small flange 16 DN mm p.2 ↗ VDL 23
Interior volume 24 L p.2 ↗ VDL 23
Net weight of the unit (empty) 72 kg p.2 ↗ VDL 23
Load per rack 20 kg p.2 ↗ VDL 23
Permitted load 50 kg p.2 ↗ VDL 23
Wall clearance back 100 mm p.2 ↗ VDL 23
Wall clearance sidewise 70 mm p.2 ↗ VDL 23
Width net 523 mm p.3 ↗ VDL 23
Height net 698 mm p.3 ↗ VDL 23
Depth net 413 mm p.3 ↗ VDL 23

Operating Limits

  • Temperature range: +9 °C above ambient temperature to 110 °C p.2 ↗
  • Permitted load: 50 kg p.2 ↗
  • Load per rack: 20 kg p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 70 mm p.2 ↗
  • ATEX classification: EX II 2/3/- G IIB T3 Gb/Gc/- X p.1 ↗

Type Code Decoder

Position Token Meaning
1-6 p.1 ↗ VDL023 Model series VDL, size 23
field 7 p.1 ↗ - Separator
field 8-10 p.1 ↗ 120 / 230 Rated voltage variant: 120 V (100...120 V) or 230 V (200...230 V)
field 11 p.1 ↗ V Voltage unit

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
VDL023-230V p.1 ↗ 9630-0009 200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard 230 V variant
VDL023-120V p.1 ↗ 9630-0013 100...120 V 1~ 50/60 Hz, NEMA 5-15 plug, Standard 120 V variant

Related models