Window and source replacement maintenance procedures
Orderingpp. 61-62
Spare parts and consumables list
Specp. 63
Instrument technical specifications
Warningpp. 64-66
Safety information and guide conventions
p. 18table, 5 rows
p. 11table, 10 rows
p. 61table, 16 rows
p. 16table, 5 rows
Table 2 Diagnostic values
Value
Optimal (green)
Marginal (yellow)
Critical (red)
Comment
Energy (center-burst)
28,000-15,000
>28,000 or <10000
> 30,000 or < 3000
Indicates the overall alignment and proper gain adjustment for the system.
Source
2.1 A
>2.4 A or <1.6 A
>2.7 A or <1.4 A
Indicates a problem with the infrared source control voltage, or that the source has burnt out. Both voltage and amperage are displayed; however, amperage provides a sufficient diagnostic.
Laser signal
15,000-3,000
>15,000 or <3,000
>17,000 or < 2,000
Can verify gross alignment errors even if the reflectance cap is not in place.
Detector temperature
38–42
<38 or >42
<36 or >44
Indicates a problem with the cooling circuitry or an ambient temperature over the specified range.
CPU temperature
10–75
>75
>80
Indicates the ambient temperature over the specified range.
When using the Diffuse Reflectance sampling accessory, the appropriate reflectance disk must be in place to make these measurements. The diffuse gold reference disk should be used with the Diffuse Reflectance sampling accessory. This disk is permanently mounted into Position 1 of the sampling tray.
Table
Раздел
Страница
Unpacking
11
Installation
12
Components
13
Connecting power
13
Power cord selection
14
Environment
15
MicroLab software
16
Turning On the Instrument
16
Turning Off
17
Performance Verification
17
Оглавление раздела Getting Started; номера страниц приведены как напечатано.
Table 3 Spare parts
Part number
Description
G8043-67483
Replacement flat tip for ATR
G8043-67484
Replacement pellet tip for ATR
G8043-67405
Polystyrene film for ATR/Transmission sampling accessory
G8043-67406
Polystyrene film for TumblIR/DialPath sampling accessory
PIKE-162-5450
NIST traceable polystyrene from PIKE approximately 38 µm (for IQOQ)
925-0128
NIST traceable polystyrene from ICL approximately 35 µm (for IQOQ)
G8043-67400
Infrared source with purge fitting
G8043-67401
Desiccant with purge fitting
G8043-67402
KBr window with removal tool
G8043-67403
ZnSe window with removal tool
G8043-67003
Cary FTIR Swivel Diamond ATR Press
G8043-67004
Cary FTIR Swivel Ge-ZnSe Press
G8043-67012
Purge kit for engine and Transmission sampling accessory
G8043-67524
Cary FTIR Base Plate Rubber Bumper+Screw
G8043-67407
Cary 635 Power Supply
G8043-67408
Cary 635 USB Cable
For service contracts, repairs and refurbishment services, contact Agilent.
Table 1 Power switch LED status
Status
LED-color
Action (% duty cycle)
System OFF
No LED illumination
N/A
System OFF
Red
100%
System startup
Red / green
Green 0.5 s on / red 0.5 s on
System ON
Green
100%
Firmware update
Red / green
Green blinks rapidly twice then red blinks rapidly twice
Engineering Facts
14 facts · every claim linked to a page
Installation
The instrument must be allowed to reach room temperature before unpacking to prevent condensation.
Prevents condensation damage during initial setup and ensures proper operation.
Source
“place the shipping case in a room-temperature environment and allow several hours for the contents of the container to reach the room ambient temperature.”
Installation
Shipping in non-original packaging voids the warranty.
Mandates use of original packaging for any transport to preserve warranty coverage.
Source
“Shipping the Cary 635 FTIR in packaging other than the original shipping container will void the warranty.”
Installation
The instrument must be kept away from hot surfaces and electromagnetic interference sources.
Ensures environmental conditions do not degrade performance or cause malfunction.
Source
“Your Cary 635 FTIR must be kept away from hot surfaces and sources of electromagnetic interference.”
Operation
A 5-minute warmup is required before reliable analysis.
Guarantees stable measurements and accurate results after power-on.
Source
“The spectrometer must have a warmup period of 5 minutes before reliable analysis may begin.”
Operation
Leaving the spectrometer on helps keep moisture out.
Prevents moisture ingress and extends component life when not actively measuring.
Source
“To help keep moisture out of the Cary 635 FTIR, it is recommended to leave the spectrometer on, even when not in use.”
Performance
Performance validation requires at least 120 minutes of warmup before testing.