Induztr
Engineering Digest Document Viewer Related Models

G7370 PAL3 RTC 120 LC Autosampler SiPCh fr D0130828

Technical Reference Summary

Covered Models & Document Scope

Site preparation checklist for the Agilent LC PAL3 RTC 120 autosampler, model G7370B (RTC 120). Document D0130828, Revision A.00, published 21-Jul-2025.

Technical Data

Parameter Value Condition
Weight 18,79 kg p.4 ↗ G7370B (RTC 120)
Height 770 mm p.4 ↗ G7370B (RTC 120)
Width 1270 mm p.4 ↗ G7370B (RTC 120)
Depth 795 mm p.4 ↗ G7370B (RTC 120)
Operating temperature range 4 à 40 °C p.6 ↗ G7370B (RTC 120)
Operating humidity range < 95 % p.6 ↗ G7370B (RTC 120), without condensation
Input voltage 100-240 V AC p.7 ↗ G7370B (RTC 120)
Input voltage 100-240 V AC p.7 ↗ G7383A/G7384A
Input voltage 100-240 V AC p.7 ↗ G7369A
Input frequency 50-60 Hz p.7 ↗ G7370B (RTC 120)
Input frequency 50-60 Hz p.7 ↗ G7383A/G7384A
Input frequency 50-60 Hz p.7 ↗ G7369A
Input current 3 A p.7 ↗ G7370B (RTC 120)
Input current 3 A p.7 ↗ G7383A/G7384A
Output voltage 36 V DC p.7 ↗ G7370B (RTC 120)
Output voltage 36 V DC p.7 ↗ G7383A/G7384A
Output current 5,55 A p.7 ↗ G7370B (RTC 120)
Output current 5,55 A p.7 ↗ G7383A/G7384A
Nominal power 200 W p.7 ↗ G7370B (RTC 120)
Nominal power 200 W p.7 ↗ G7383A/G7384A

Operating Limits

  • Operating temperature range: 4 to 40 °C (39 to 104 °F) p.6 ↗
  • Operating humidity range: < 95 %, without condensation p.6 ↗
  • Input voltage: 100-240 V AC p.7 ↗
  • Input frequency: 50-60 Hz p.7 ↗
  • Input current: 3 A p.7 ↗
  • Nominal power: 200 W p.7 ↗