Product introduction and key features of the chiller
Specpp. 1-2
Technical data and performance parameters per DIN 12876
p. 2table, 4 rows
p. 1table, 38 rows
Technical data according to DIN 12876
Parameter
Value
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S459713
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
digital
Internal temperature sensor
Pt100
Resolution of display
0,1 K
Interface digital
USB (Device), RS232
Interface
Temperature stability at -10°C
0,2 K
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at 15°C
1 kW
Cooling power at 0°C
0,8 kW
Cooling power at -10°C
0,5 kW
Cooling power at -20°C
0,15 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,33 kg
CO2 equivalent
0,5 t
Circulation pump
B
at 0,5 bar
21 l/min
at 1,0 bar
17 l/min
at 1,5 bar
11 l/min
at 2,0 bar
6 l/min
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Pump connection
G3/4 male
min. filling capacity
3,8 l
expansion tank
1,7 l
Overall dimensions WxDxH **
350x496x622 mm
Net weight
53 kg
sound pressure level +/- 4 dB(A)
57 dB(A)
Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
14,5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window before selection.
Defines whether the chiller can cover the required process temperatures at all.
Source
“Operating temperature range
-20...100 °C”
Performance
Cooling capacity falls from 1 kW at 15 °C to only 0.15 kW at -20 °C, so low-temperature duty needs a much larger unit or longer pull-down.
Prevents undersizing when the application runs near the cold end of the range.
Source
“at 15°C
1 kW
at 0°C
0,8 kW
at -10°C
0,5 kW
at -20°C
0,15 kW”
Installation
The refrigeration machine is air-cooled, so sufficient room ventilation and respect for space requirements are mandatory at the installation site.
Air-cooled condensers reject heat to the room and need adequate airflow clearance.
Source
“Refrigeration machine
air-cooled, CFC- and
HCFC-free”
Installation
The power supply requirement is 220-240 V single/two-phase 50/60 Hz with a maximum current of 14.5 A and a 16 A fuse, which dictates the site electrical supply.
The electrical connection, cable and protection must be sized for 14.5 A and a 16 A fuse.
Source
“Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
14,5 A
Fuse
16 A”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is explicitly not allowed.
Site power quality must be checked; an out-of-tolerance combination can invalidate operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
The correct plug type must be stated on the order to avoid a non-connectable unit.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The plastic pump is not resistant to acetone and polyglycols, so these media must be avoided or a more resistant plastic must be ordered.
Prevents chemical attack and pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection water should be mixed with glysantine or ethylene glycol rather than used pure.
Defines the correct heat transfer fluid preparation for sub-zero operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Performance
The circulation pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar, so high-pressure-drop circuits reduce available flow.
The pump curve must be checked against the pressure drop of the external circuit.
Source
“at 0,5 bar
21 l/min
at 1,0 bar
17 l/min
at 1,5 bar
11 l/min
at 2,0 bar
6 l/min”
Performance
Maximum delivery is 25 l/min at a maximum delivery pressure of 2.5 bar, setting the hydraulic limit of the unit.
Defines the upper bound of flow and pressure available to the application.