The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase power connection.
Determines electrical infrastructure and whether a plug or hard wiring is needed.
Source
“Power supply requirement 400V 3~ 50Hz”
| Parameter | Value |
|---|---|
| Delivery at 5,0 bar | 42 l/min |
| Pump connection | M38x1,5 male |
| max. permissible kin. viscosity | 100 mm²/s |
| Cooling water connection | G3/4 male |
| min. cooling water differential pressure | 1 bar |
| max. cooling water differential pressure | 4 bar |
| max. cooling water pressure | 6 bar |
| max. System pressure | 6 bar |
| min. filling capacity | 8,9 l |
| Filling capacity expansion tank | 19,9 l |
| Overall dimensions WxDxH ** | 918x963x1771 mm |
| Power supply requirement | 400V 3~ 50Hz |
| Pressure equipment category | II |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | 1.0/24 |
** Please respect space requirements. See operating conditions at www.huber-online.com
| Parameter | Value |
|---|---|
| Operating temperature range | -50...200 °C |
| Temperature stability at -10°C | 0,01 K |
| temperature set point / display | 5,7" colour Touchscreen |
| Resolution of display | 0,01 K |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| digital input | ECS ONE |
| digital output | POKO ONE |
| Interface analog | optional |
| Alarm message | optic, acoustic, relay |
| Safety classification | III / FL |
| Heating power | 24 kW |
| Cooling power with Thermooil at 200°C | 41 kW |
| Cooling power with Thermooil at 100°C | 41 kW |
| Cooling power with Thermooil at 20°C | 41 kW |
| Cooling power with Ethanol at 0°C | 37 kW |
| Cooling power with Ethanol at -20°C | 30 kW |
| Cooling power with Ethanol at -40°C | 15 kW |
| Cooling power with Ethanol at -50°C | 8 kW |
| Refrigeration machine | water-cooled, natural refrigerant |
| Refrigerant (ASHRAE, GHS) | R-744 (A1, H280) |
| Global Warming Potential (GWP) | 1 |
| Circulation pump: MK pump max. delivery | 201 l/min |
| Circulation pump: MK pump max. delivery pressure | 5,3 bar |
| Delivery at 1,0 bar | 179 l/min |
| Delivery at 2,0 bar | 154 l/min |
| Delivery at 3,0 bar | 124 l/min |
| Delivery at 4,0 bar | 87 l/min |
Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' объединены с соответствующими температурными точками.
The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase power connection.
Determines electrical infrastructure and whether a plug or hard wiring is needed.
“Power supply requirement 400V 3~ 50Hz”
The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar, with G3/4 male connections.
Cooling water supply pressure and connection size must be verified before installation.
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
The unit is designed for cooling water intake between 5 °C and 20 °C, and lower temperatures require insulating the cooling water pipes.
Cooling water temperature affects cooling capacity and condensation risk.
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Cooling capacity drops as cooling water temperature rises, and cooling water flow rate may need to increase.
Performance must be derated if the available cooling water is warmer than the reference conditions.
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
Water quality must be compatible with these materials to avoid corrosion or fouling.
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and opposite-sign deviations are not allowed.
Site power quality must be checked against these limits to avoid invalid operating conditions.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation scope and whether a plug must be sourced separately.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
The unit has safety classification III / FL and adjustable overtemperature protection according to DIN 12876.
Safety class and overtemperature protection are relevant for risk assessment and application suitability.
“Safety classification III / FL”
The pump connection is M38x1.5 male, and pipework or hoses must use ball socket per DIN 3863 and sleeve nut per DIN 3870.
Ensures correct mating of hoses and pipework to the circulation pump.
“Pump connection M38x1,5 male”
The maximum permissible kinematic viscosity of the medium is 100 mm²/s.
The heat transfer fluid must be selected within this viscosity limit for proper pump operation.
“max. permissible kin. viscosity 100 mm²/s”
The unit requires a minimum filling capacity of 8.9 l and has a 19.9 l expansion tank.
Determines the required heat transfer fluid volume and expansion allowance.
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
The unit must be installed in an ambient temperature range of 5 °C to 40 °C.
Ambient conditions affect reliable operation and must be respected at the installation site.
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.
Class A equipment may require EMC precautions in residential or sensitive environments.
“Classification (disturbance) to EN55011: Class A, Group 1”
The refrigeration machine is water-cooled and uses natural refrigerant R-744 with GWP 1.
Cooling water supply and refrigerant type affect maintenance and environmental compliance.
“Refrigeration machine water-cooled, natural refrigerant”