Induztr
Engineering Digest Document Viewer Related Models

Agilent Bond Elut Lipid Extraction

Technical Reference Summary

Covered Models & Document Scope

Agilent Bond Elut Lipid Extraction 96-well plate, 1 plate, part number 5610-2042; method guide for protein precipitation, sample loading, washing, and lipid elution

Technical Data

Parameter Value Condition
Plate format 96-well plate p.1 ↗
Sorbent chemistry EMR—Lipid p.1 ↗
Recommended crashing solvent ACN with 5% MeOH p.1 ↗
Crashing solvent to sample ratio 9:1 p.1 ↗
Washing solvent ACN with 10 to 20% water p.1 ↗
Elution solvent MeOH with DCM, chloroform, or 1-chlorobutane p.1 ↗
Minimum MeOH content for elution 50 % p.1 ↗
Recommended elution solvent DCM/MeOH (v/v 1:2) or chloroform/MeOH (v/v 1:1) p.1 ↗
Plasma sample volume 100 μL p.2 ↗
Collection plate volume 2 mL p.2 ↗
Crashing solvent volume 900 μL p.2 ↗
Sonicating time 10 minutes p.2 ↗ on ice
Wash solvent volume 2 × 1 mL p.2 ↗ ACN/water (v/v 9:1)
Elution volume 1 mL p.2 ↗ DCM/MeOH (v/v 1:2)

Operating Limits

  • For Research Use Only; not for use in diagnostic procedures: p.2 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
Agilent Bond Elut Lipid Extraction, 96-well plate p.2 ↗ 5610-2042 1 plate