Induztr
Engineering Digest Document Viewer Related Models

ASM 306S — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ASM 306S helium/hydrogen sniffer leak detector, EN operating instructions (78 pages, translation of original). Applies to part number RSAS00AxMM9A, where 'x' varies according to the selected interface option (I/O, Serial RS-232 D-Sub 9, Ethernet, USB, PROFIBUS DP-V1, PROFINET). Related documents: Maintenance instructions P/N 127443M, Communication interface P/N 130417, Sniffer probe condensed manual P/N 127828. Sniffer probe (P/N PRBxxxxxx) and calibrated leak are separate accessories.

Section-by-Section Technical Breakdown

pp. 15-17
Product description

Describes identification, scope of delivery, connection interfaces, touch control panel, sniffer probe and calibrated leak. Sniffer probe and calibrated leak are essential but ordered separately. Calibrated leaks exist for 100% 4He and 100% H2, value range 3·10⁻⁵–6·10⁻⁵ mbar·l/s.

  • Delivery includes detector, documentation set (USB stick), power cable(s), 6 partitions, sniffer probe sheath, D-Sub 15/37-pin connector kit depending on option
  • Control panel: touch screen, Home button, START/STOP, CAL, ZERO, probe cable connector, 4 magnets
  • Sniffer probe: rod + filter, LED status, LED bargraph, ZERO button, cable connector; LED coding table (Tbl. 1)
  • Calibrated leaks offered: 100% 4He and 100% H2 (3·10⁻⁵–6·10⁻⁵ mbar·l/s = 3·10⁻⁶–6·10⁻⁶ Pa·m³/s), each with calibration certificate; no 3He leak offered
  • Storage box load limit (see commissioning): max 5 kg
pp. 18-19
Installation and electrical connection

Detector must be installed horizontally on its legs with ≥10 cm free space all around, nothing stored under it, ventilation openings unobstructed. Electrical connection is Class I, must be earthed, using only the supplied/original mains cable; shielded cables required in EMC-critical environments.

  • Free space ≥ 10 cm all around; do not obstruct ventilation openings; clean them regularly; store nothing under detector
  • Class I equipment: must be earthed; mains switch to O before connecting
  • Use shielded cables and connections for interfaces in interference-prone environments
pp. 20
Commissioning

Covers precautions before each use, switching on, and powering off. Environment must be free of tracer gas; storage box content limited to 5 kg. After shutdown >3 months, switch on 24 h before use. After a power failure the detector restarts automatically.

  • Max 5 kg total weight of parts/accessories in the storage box
  • Startup sequence: connect mains, connect sniffer probe, breaker to I, set language/unit/date/time on first start, wait for 'Measure' mode
  • After >3 months shutdown: power on 24 hours before use
  • Power-off: breaker to O, unplug, wait 5 minutes before working on/moving the detector
  • On mains failure detector shuts down and restarts automatically when power is restored
pp. 21-33
Operation: sniffing test, calibration, zero, touch screen

Sniffer method: test piece pressurized with tracer gas, probe samples escaping gas; global or localization test. Hydrogen tracer must be 95% N2 / 5% H2 mix. Calibration (external leak or concentration) is prompted 20 minutes after switch-on and recommended daily. Zero function is permanently enabled; touch screen offers main and graph screens with recording, saving (.csv), graph history and zoom.

  • Only tracer gases 4He, 3He, H2; H2 max 5% concentration (95% N2 / 5% H2 forming gas)
  • Ambient [He] function: normal 4He background is 5 ppm; not available for H2/3He or in Standby
  • Calibrate: at least once a day, after power-on prompt (20 min after startup), or every work session (~8 h) for intense operation
  • External calibration: use calibrated leak ~10⁻⁵ mbar·l/s range containing the set tracer gas
  • Test stop in ECO mode via probe: test stops automatically after 10 minutes; without ECO mode, test cannot be started/stopped from the probe
  • Graph parameters: decades -11 to +6 (high) / -12 to +5 (low), max 10 decades span; display time 12 s–1 h; sampling time 100 ms–30 s; auto scale 2 or 4 decades
  • Graph history: 12 s display time → 21 min history; 1 h display time → 105 h (~4 days) buffer history
  • Recordings saved as .csv (default separator: tab) to USB stick or internal memory; recording cleared on power-off
  • Screenshot: press ZERO + [Home] simultaneously; files named ScreenYYYYMMDD_HHMMSS
  • Status feedback: green below reject point, red above; warnings shown via i Next pictogram; critical error E244 requires service
pp. 34-55
Settings menu

Six menus: Measure (tracer gas, set points, correction factor, calibrated leak reference/target, leak settings), Probe (flow unit, probe clogged, ECO mode), Configuration (unit/date/time/language, sound, screen, password), Maintenance, File Manager, Advanced (I/O, Service). Leak rate remains permanently displayed on settings menus. Sniffing method options are Localization (reject point only) and Pass-Fail.

  • Set points: Localization mode uses reject point only; Pass-Fail mode defines go/no-go testing
  • Set points include reject point (red) and warning point (orange) displayed on main/graph screens
  • Password-protected menus indicated by red closed / green open padlock; temporary access re-locks on return to main screen
  • Maintenance menu: history, information, last operations, counters before next maintenance, turbo pump & cell maintenance, parameter import/export
  • Default settings documented in tables Tbl. 9–17 in the appendix (pages 69–75)
pp. 56-76
Troubleshooting, maintenance, accessories and technical data

Troubleshooting guide (fault types: warning, error, critical error E244), maintenance/replacement chapter, Pfeiffer service solutions, accessories table, and technical data with dimensions and unit conversion tables. UL/CSA compliance and EC Declaration of Conformity appended.

  • Critical error 'E244': contact a service center; LED status/bargraph indicate warning, error and critical states
  • Accessories table (Tbl. 4) lists orderable sniffer probe and calibrated leak P/Ns
  • Technical data (Tbl. 5) and environmental conditions (Tbl. 6) on p. 67; pressure and gas throughput unit conversion tables on p. 68
  • Appendix contains full settings-menu tree diagram and default settings tables (Tbl. 9–17)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Hydrogen tracer gas concentration must never exceed 5%; recommended tracer gas is forming gas (95% N2 / 5% H2). p.11 ↗ Specify forming-gas charging of test parts; higher H2 concentrations void warranty and pose explosion risk. ASM 306S (all variants)
Supported tracer gases are 4He, 3He and H2; Ambient [He] function reads 4He background only (normal 5 ppm). p.21 ↗ Use Ambient [He] to verify the test area is not helium-polluted before testing; not usable for H2/3He or in Standby. ASM 306S (all variants)
Manufacturer calibrated leaks cover 3·10⁻⁵–6·10⁻⁵ mbar·l/s (3·10⁻⁶–6·10⁻⁶ Pa·m³/s), available for 100% 4He and 100% H2 only; none offered for 3He. p.17 ↗ For 3He operation, an external third-party calibrated leak must be sourced; each supplied leak comes with a calibration certificate. Calibrated leak accessory
Recommended external calibration leak range is ~10⁻⁵ mbar·l/s (10⁻⁶ Pa·m³/s) with the set tracer gas; calibration is prompted 20 min after power-on and should be done at least daily (every ~8 h shift for intense use). p.23 ↗ Plan calibration into the daily test routine; leak rate values are only trustworthy after calibration. ASM 306S (all variants)
Installation requires ≥10 cm free space all around, ventilation openings unobstructed, detector horizontal on its legs, nothing stored under it. p.18 ↗ Insufficient ventilation causes internal overheating and component deterioration; reserve clearance in cabinet/integration design. ASM 306S (all variants)
Storage box payload must not exceed 5 kg total (parts, accessories); transport cart max load is 26 kg and is not recommended for routine use. p.13 ↗ Do not overload the storage compartment; moving the detector by probe/panel/cables is prohibited. ASM 306S (all variants)
After shutdown exceeding 3 months, the detector must be switched on 24 hours before use; long storage (>3 months) risks elastomer/lubricant damage. p.14 ↗ Schedule pre-use startup time for rarely used detectors; inspect with service if stored long-term. ASM 306S (all variants)
Power-off procedure: breaker to O, unplug mains, wait 5 minutes before working on, opening or moving the detector; components between mains input and circuit breaker remain live after switch-off. p.10 ↗ Include the 5-minute wait and unplugging requirement in lockout/service procedures. ASM 306S (all variants)
Product is Class I equipment and must be earthed; only original manufacturer mains cable may be used, and product must never run with an ungrounded cable. p.7 ↗ Verify protective earth at installation site; order replacement cables only via P/N in the maintenance instructions. ASM 306S (all variants)
ECO mode with probe-initiated testing: test stops automatically 10 minutes after the probe is put down; without ECO mode the test cannot be started or stopped from the sniffer probe. p.23 ↗ Enable ECO mode for probe-driven workflows; otherwise operate tests only from the control panel START/STOP. ASM 306S (all variants)
Graph parameters: high decade -11 to +6, low decade -12 to +5 (max 10 decades span), display time 12 s–1 h, sampling time 100 ms–30 s, auto scale 2 or 4 decades. p.29 ↗ Configure recording resolution and time window per test requirements; sampling down to 100 ms for fast transients. ASM 306S (all variants)
Graph history buffer: 21 min at 12 s display time setting; 105 h (~4 days) at 1 h display time; recordings are volatile (cleared on power-off) and must be manually saved as .csv (tab separator) to internal memory or USB stick. p.29 ↗ Save recordings before powering down; auto-loss of in-progress recording on power failure. ASM 306S (all variants)
Sniffer probe is mandatory for use, not included, and only the manufacturer's probe (P/N PRBxxxxxx) may be used; probe flow reading must be non-zero for valid measurement. p.15 ↗ Budget for probe purchase separately and verify probe flow on the main screen before testing. ASM 306S + sniffer probe accessory
Set point logic: reject point (red) and warning point (orange); screen/bargraph turn green below reject, red above; Localization mode supports reject point but not warning point. p.27 ↗ In localization testing, only reject-based judgment is available; Pass-Fail mode uses both set points. ASM 306S (all variants)
Critical error E244 disables operation and requires return to a Pfeiffer service center; warnings/errors are signaled via i Next pictogram and probe LED/bargraph codes. p.22 ↗ Integrate fault monitoring (i Next indicator, RS-232 remote control) into production line supervision. ASM 306S (all variants)
Power ordering code RSAS00AxMM9A; interface option (I/O, RS-232 D-Sub 9, Ethernet, USB, PROFIBUS DP-V1, PROFINET) selected via the 'x' digit; D-Sub 15 or 37-pin connector kit included depending on option. p.6 ↗ Specify the correct interface digit when ordering; remote I/O mapping documented in Communication interface manual P/N 130417. ASM 306S ordering matrix
Screenshots via simultaneous ZERO + [Home] press are saved internally as ScreenYYYYMMDD_HHMMSS; if the touch screen fails, the detector can still be controlled via RS-232. p.25 ↗ Provide RS-232 fallback control path for broken-screen operation and keep documentation for audit trails. ASM 306S (all variants)
Use is restricted to industrial environments; misuse includes pumping liquids, dust, corrosive/explosive/flammable/toxic fluids, condensable vapors and testing soiled or wet parts. p.11 ↗ Only test clean, dry parts; protect the sniffer path from condensables and particulates to avoid cell contamination. ASM 306S (all variants)