Agilent Technologies Compact microplate stacking, serial dilution, plate stamping
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies Compact microplate stacking, serial dilution, plate stamping

Agilent Technologies Compact microplate stacking, serial dilution, plate stamping — Automated liquid handling.

The Agilent Bravo BenchCel Workstation is an integrated automated liquid handling platform combining a Bravo pipetting head (300 nL to 200 µL) with a BenchCel 2R microplate handler and 660 mm front loading racks, driven by VWorks Automation Control Software. It is configured for serial dilution, plate stamping and microplate stacking/sealing workflows on 96-well bottom plates. Average pipetting precision is 2.7 %.

Full specifications

Core Performance & Functional Specifications
Average Precision 2.7 %
Pipetting volume range mfr 300 μL
Microplate handler front loading rack size mfr 660 mm
Mix volume per mix cycle mfr 150 μL
Mix amplitude mfr 6 mm
Number of mix cycles mfr three three
Processing time mfr 568 sec/plate
Accuracy ratio mfr 1
Microplate handler model mfr 2 R
Microplate sealer model mfr PlateLoc Thermal Microplate Sealer Sealer
Microplate type mfr 96 bottom
Control software mfr VWorks Automation Control Software Software
Dimensions, Weight & Mechanical Design
Front loading rack height mfr 660 mm
Standards, Safety & General Data
Number of standard deck positions mfr nine nine
Tartrazine standard concentration mfr 0.36 mM

Description

Construction and positioning

Behavior under load and in different modes

Installation, connection and control

Integration & Operating Notes

Control & I/O
VWorks must sequence pipetting and plate-handler moves

The Bravo pipetting head and the BenchCel 2R handler are separate devices coordinated by VWorks Automation Control Software. The plate handler must not present or remove a plate while the pipetting head is over the deck; the VWorks protocol must define handshake steps and safe park positions before any plate transfer.

Application Note 5991 8090EN; Application Note 5991 8091EN
Process Limits
Pipetting volume must stay within 300 nL to 200 µL

The Bravo pipetting head is documented for 300 nL to 200 µL. Protocols calling for volumes outside this band require a different head or a dilution step; the 2.7 % average precision figure applies to the validated serial dilution workflow, not to arbitrary volumes.

Application Note 5991 8090EN
Mounting
Reserve 660 mm front clearance for loading racks

The BenchCel 2R handler uses 660 mm front loading racks. The bench or enclosure must provide unobstructed front access of at least this depth so plates can be loaded and unloaded without disturbing the deck; bench depth and door swing must be checked before siting.

Application Note 5991 8091EN
Maintenance
Sealer and barcode labeler are separate modules

The PlateLoc Thermal Microplate Sealer and the Microplate Barcode Labeler are documented as adjacent modules integrated with the 2R handler, not as part of the base workstation. Sealing requires a peelable aluminum thermal seal roll, which is a consumable ordered separately; plan roll replacement and sealer maintenance as separate service items.

Application Note 5991 8091EN

Selection & Integration Q&A

When should I choose this Bravo BenchCel Workstation instead of the sibling compact liquid handler or the sibling microplate handler?

Choose this workstation when a protocol needs both pipetting (300 nL to 200 µL) and automated plate stacking in one VWorks run, e.g. serial dilution at 568 sec/plate. Pick the compact liquid handler if only pipetting is needed, or the microplate handler if only stacking/sealing is required.

What adjacent modules and consumables must be specified alongside this workstation for a sealing workflow?

The 2R handler workflow integrates the PlateLoc Thermal Microplate Sealer or the Microplate Barcode Labeler under VWorks control. The sealer uses a peelable aluminum thermal seal roll, a consumable not in base delivery; confirm sealer and labeler are quoted as separate line items.

Which items are not included in base delivery and must be ordered or scheduled separately?

The PlateLoc Thermal Microplate Sealer, the Microplate Barcode Labeler and the peelable aluminum thermal seal roll are separate from the base workstation. VWorks Automation Control Software is the control layer; confirm software licenses and any integration cabling are quoted separately.

Dimensional Drawings & Technical Documentation

Technical documentation