Agilent Technologies 5100 ICP OES
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 5100 ICP OES

Agilent Technologies 5100 ICP OES — Atomic spectroscopy.

The Agilent 5100 ICP OES is a dual-view (RV/DV) inductively coupled plasma optical emission spectrometer configured with a fully demountable Easy-fit torch. Its standard quartz outer tube-set with 1.8 mm id injector serves most aqueous and acidic samples, while the ceramic outer tube-set with 2.4 mm id inert alumina injector extends the working range to HF digests and high-TDS matrices up to 35% TDS. It is intended as the excitation and sample-introduction front end of an atomic emission analytical system.

Full specifications

Core Performance & Functional Specifications
Injector Internal Diameter (id) 1.8 mm
Maximum Total Dissolved Solids (tds) 35 %
Operating Conditions, Environment & Protection
Cooling time before torch removal mfr at least five minutes minutes

Description

Construction and series positioning

Behavior across sample matrices

Installation, control, and operation

Integration & Operating Notes

Process Limits
Do not exceed 35% TDS with the 2.4 mm alumina injector

The 2.4 mm id inert alumina injector G8020-47004 is rated for HF digests and high-TDS samples up to 35% total dissolved solids. Exceeding this matrix load risks injector blockage and devitrification-related failure; dilute or switch to a more suitable sample-introduction configuration before running higher-TDS digests.

User Manual 5994-8863EN, page 7
Maintenance
Cool torch at least five minutes after plasma switch-off

The torch and torch compartment remain hot after the plasma is extinguished. Allow at least five minutes of cooling before attempting removal, and maintain a straight, parallel movement when inserting or removing the torch and its quartz or ceramic components to avoid breakage and hand injury.

User Manual 5994-8863EN, page 4
Maintenance
Hand-tighten the injector locking nut only

The locking nut must be tightened by hand; using tools or overtightening can damage both the injector and the torch body. During reassembly, seat the top seal and align the outer tube-set groove with the 'T' mark before pressing the tube-set fully into the body and inserting the injector until it clicks.

User Manual 5994-8863EN, pages 3 and 5
Process Limits
Match outer tube-set and injector to the sample matrix

Standard quartz with 1.8 mm id injector suits most aqueous and acidic samples; organic high-purity quartz with 0.8 mm id (volatile) or 1.4 mm id (semi-volatile) injectors resist solarization in organic matrices; ceramic outer tube-set with 2.4 mm id inert alumina is reserved for high-TDS, HF digest, and fusion work. Selecting the wrong combination shortens torch life or compromises analytical results.

User Manual 5994-8863EN, Table 1, page 2
Maintenance
Ceramic tube-set kit ships without an injector

The fully demountable torch kit with ceramic outer tube-set does not include an injector and must be ordered separately. The quartz intermediate tube of the ceramic outer tube-set has a fixed polymer adapter that should not be removed; use the supplied tube-set tool when separating the ceramic outer tube from the quartz intermediate tube.

User Manual 5994-8863EN, pages 2 and 5

Selection & Integration Q&A

When should I choose the fully demountable torch over the one-piece or semi-demountable Easy-fit torch for the 5100 ICP OES?

Choose the fully demountable torch when the sample matrix varies and you need to swap injectors (quartz or inert alumina, 0.8–2.4 mm id) and outer tube-sets (quartz, organic high-purity quartz, or ceramic) independently. The one-piece and semi-demountable torches are ready to use out of the box and suit fixed, routine aqueous/acidic workloads where reconfiguration is not required.

What must be ordered alongside the ceramic outer tube-set kit, and which injector is compatible with HF digests and high-TDS samples?

The ceramic outer tube-set kit does not include an injector, so a compatible quartz or inert alumina injector must be ordered separately. For HF digests and high-TDS matrices, the 2.4 mm id inert alumina injector G8020-47004 is specified, with total dissolved solids limited to 35%.

What installation and handling limits apply when fitting or removing the torch on the 5100 ICP OES?

The torch is shipped without the injector installed; insert it until it clicks into the retaining clamp and hand-tighten the locking nut only. After plasma switch-off, allow at least five minutes for the torch and compartment to cool, and keep a straight parallel movement when inserting or removing the torch and its quartz or ceramic parts to avoid breakage.

Which consumables and spare parts are not included in the base delivery and should be stocked for maintenance?

Replacement injectors (quartz or inert alumina, 0.8/1.4/1.8/2.4 mm id), outer tube-sets (standard quartz, organic high-purity quartz, ceramic for RV or DV), top seal pack of 3 (G8014-60023), and injector locking nut (G8020-60810) are separate parts. Cleaning and maintenance procedures are covered in publication 5994-8864EN.

Dimensional Drawings & Technical Documentation

Technical documentation