Eppendorf New Brunswick S41i - CO 2 Incubator Shaker
Eppendorf New Brunswick S41i - CO 2 Incubator Shaker — Incubators with Capacity 170 L.
The New Brunswick S41i is a 170 L benchtop CO2 incubator shaker combining a triple-eccentric counterbalanced Innova drive (25–400 rpm, 2.5 cm orbit) with 4–50 °C and 0.2–20 % CO2 control in a seamless chamber. It is stackable or floor-mounted, intended for suspension-cell and microbial cultivation requiring simultaneous temperature, CO2 and agitation.
Full specifications
| Capacity | 170 L |
| Audible And Visual Alarms | yes |
| Available Program Modes | Constant speed and temperature |
| Benchtop (b), Under Bench (u), Floor (f) Or Stackable (s) | U, F, S |
| Bms Relay | standard |
| Co 2 Control Increment | 0.1 % |
| Co 2 Range | 0.2...20 % |
| Co 2 Stability At 5 % Co 2 | 0.2 % |
| Co 2 Uniformity | 0.1 % |
| Door Opening | opening to front |
| Drive Mechanism | triple-eccentric counterbalanced drive |
| High Temperature Disinfection (htd) | 120 °C |
| Memory | non volatile with automatic power failure restart |
| Motor Type | Solid state, DC brushless motor |
| On-board Data Logging | standard |
| Orbit | 2.5 cm |
| Platform Size (cm/in) | 36 cm |
| Remote Device Monitoring And Notifications (visionize ® ) | VisioNize ® box enabled |
| Sealed Inner Glass Doors | standard |
| Segmented Door Options | – |
| Temperature Accuracy | 0.2 °C |
| Temperature Control Increment | 0.1 °C |
| Temperature range | 4...50 °C |
| Temperature Stability At 37 °c | 0.1 °C |
| Temperature uniformity | 0.25 °C |
| No. of compartments mfr | 1 (2. optional) |
| Timer mfr | Continuous |
| Dimensions (W × D × H) mfr | 87,5 × 73 × 85 cm / 34,4 × 28,7 × 33,5 in |
| Speed range mfr | 25 – 400 rpm (stacked: 25 - 250/400 rpm for top/ |
| Perforated shelves mfr | standard |
| Shaking mfr | standard |
| Seamless chamber mfr | standard |
| In-line gas regulator mfr | standard |
| Product Weight | 154 kg |
| External depth (with lid/door open) mfr | 144.8 cm (57 in) |
| Power supply mfr | 230 V ± 10 %, 50/60 Hz |
| Data Export Interface | Ethernet |
| Power Plug | power plug type F |
| Access port mfr | 1 |
| Cetlus And Ce Marked | yes |
| Number Of Shelves (standard/max) | 1 |
Description
Construction and positioning
- Single-compartment, 170 L chamber with a seamless interior, one standard perforated shelf and a 36 cm shaking platform; sealed inner glass doors open to the front.
- Drive is a triple-eccentric counterbalanced mechanism on a solid-state DC brushless motor, giving a 2.5 cm orbit across 25–400 rpm.
- The unit is configured for under-bench (U), floor (F) or stackable (S) installation; when stacked, the top and bottom units are limited to 25–250/400 rpm respectively. External envelope is 87.5 × 73 × 85 cm and 154 kg, with 144.8 cm depth required when the door is open.
- A second compartment is available as an option; the standard delivery has one.
Behavior under load and in different modes
- Temperature is held over 4–50 °C with 0.1 °C set increment, 0.2 °C accuracy, 0.1 °C stability at 37 °C and 0.25 °C uniformity.
- CO2 is controlled from 0.2 to 20 % in 0.1 % increments, with 0.2 % stability at 5 % CO2 and 0.1 % uniformity.
- Program modes are constant speed and constant temperature; the timer is continuous. No digitized performance curves are supplied, so no derating or load-dependent speed/temperature behavior can be stated beyond the stacked-unit speed limits.
Control, data and operation
- On-board data logging is standard, memory is non-volatile with automatic power-failure restart, and an Ethernet interface plus a standard BMS relay and VisioNize box enable remote monitoring and notifications.
- Audible and visual alarms are fitted; an in-line gas regulator and one access port are standard.
- High-temperature disinfection (HTD) is available at 120 °C.
- Electrical supply is 230 V ± 10 %, 50/60 Hz with a type F power plug; the unit is cETLus and CE marked.
Integration & Operating Notes
The external depth with the door open is 144.8 cm, so bench or floor placement must leave that clearance in front of the unit for loading and for the sealed inner glass doors; the closed footprint is 87.5 × 73 cm and the unit weighs 154 kg, which also governs floor loading and lifting equipment.
When two units are stacked, the speed range is restricted to 25–250/400 rpm for top/bottom unit respectively; a protocol requiring more than 250 rpm on the upper unit cannot be run in a stacked configuration and must be assigned to the lower position or a non-stacked installation.
The unit is specified only for 230 V ± 10 % at 50/60 Hz and is fitted with a type F (Schuko) power plug; a 115 V supply or a non-F receptacle requires a site change or an appropriate adapter arrangement, and the ±10 % tolerance must be respected at the point of connection.
Because memory is non-volatile with automatic power-failure restart, the unit resumes its set speed, temperature and CO2 conditions after a mains interruption; engineers integrating the BMS relay or VisioNize box should treat a power-loss event as a restart, not a shutdown, and alarm logic must distinguish the two.
The CO2 range is 0.2–20 % with 0.1 % increments; protocols needing a true 0 % CO2 atmosphere (for example certain anaerobic or non-CO2 incubations) fall outside the specified control range and require a different chamber or gas strategy.
High-temperature disinfection is available at 120 °C; scheduling an HTD cycle requires the chamber to be emptied of samples and heat-sensitive accessories, and the cooldown back into the 4–50 °C working range must be accounted for in the run schedule.
Selection & Integration Q&A
Pick the S41i for a 170 L single-compartment benchtop workload at 25–400 rpm and 4–50 °C, especially where under-bench, floor or stacked placement is needed. The CS220 is the larger alternative referenced in the brochure when 170 L and the 36 cm platform are insufficient.
A 230 V ± 10 %, 50/60 Hz supply with a type F receptacle, plus an Ethernet drop for data export and VisioNize box monitoring. A standard BMS relay is wired for remote alarm signalling, and one access port is available for probe or gas-line routing.
Yes: it is rated for under-bench, floor or stacked use, but the open door needs 144.8 cm clear depth and the 154 kg weight must be supported. In a stacked pair the top unit is limited to 250 rpm while the bottom can reach 400 rpm.
The standard unit ships with one perforated shelf, one access port and an in-line gas regulator. A second compartment is optional, and additional shelves or probe feedthroughs are not part of the base configuration; confirm these against the order.