Nabertherm NA 2000/45 LS — Forced Convection.
The NA 2000/45 LS is a forced-convection chamber furnace with a 1100 × 1500 × 1200 mm working chamber and 450 °C maximum temperature, built to EN 1539 safety technology for solvent-laden loads. Its 250-unit exhaust air flow and graduated solvent limits distinguish it from non-LS NA variants, making it suitable for drying and heat treatment where flammable volatiles are released.
Full specifications
| Heating power | 72 watt |
| Heating power | 72 watt |
| Exhaust air flow rate mfr | 250 |
| Inner dimensions (W × D × H) mfr | 1100 × 1500 × 1200 mm |
| Outer dimensions (W × D × H) mfr | 2115 × 2650 × 2575 mm |
| Max temperature | 450 °C |
| Max temperature | 450 °C |
Description
Construction and series positioning
- The NA 2000/45 LS is the safety-technology (LS) version of the NA 2000 forced-convection chamber furnace, with a 2000-litre-class chamber measuring 1100 × 1500 × 1200 mm (W × D × H) and an outer envelope of 2115 × 2650 × 2575 mm.
- The LS designation follows EN 1539 safety technology, which governs operation with solvent-containing charge materials; non-LS NA models in the same size class do not carry this solvent-rated exhaust architecture.
- Heating power is listed as 72 W and maximum temperature as 450 °C; the low power figure should be verified against the order-specific electrical data before sizing the supply.
Behavior with solvent loads
- The exhaust air flow rate is 250 (unit as printed). Solvent admission is not a single figure but a temperature-dependent limit: the maximum solvent mass in grams falls as temperature rises, from 172 g at the lowest listed temperature down to 18 g at the highest.
- This means the charge composition and the programmed temperature ramp must be matched to the solvent limit at each setpoint; a load acceptable at low temperature may exceed the limit at the upper end of the 450 °C range.
Control and documentation
- The controller documentation covers the Nabertherm 500 series and MyNabertherm app, including process data storage and AMS2750F temperature-uniformity instrumentation types; recorder and controller options are selected separately from the furnace.
- No digitized curve points or certificates were supplied for this model, so heat-up and uniformity performance must be taken from the order documentation rather than this card.
Integration & Operating Notes
The admissible solvent mass decreases with temperature, from 172 g down to 18 g across the listed range. Exceeding the limit for the actual chamber temperature risks exceeding the safe solvent concentration in the exhaust stream, so the batch recipe must be checked against the limit at every ramp segment, not only at the final setpoint.
The solvent limits are tied to the stated exhaust air flow rate of 250. Ducting, extraction fans or building ventilation connected to the furnace outlet must sustain this flow; a restricted or shared exhaust path invalidates the solvent rating and is a safety issue, not merely a performance loss.
The specification lists 72 W heating power, which is not consistent with a 2000-litre 450 °C chamber and should be treated as a data-entry artifact. The actual connected load, voltage and phase must be taken from the order-specific nameplate before preparing the power supply and protection devices.
The 500-series controller and MyNabertherm app functions, including process data storage and AMS2750F instrumentation types, are documented as controller options rather than furnace-integrated defaults. Specify the required controller, recorder and uniformity class at ordering; retrofitting instrumentation later changes the validation basis.
Selection & Integration Q&A
Choose the LS version when the charge releases flammable solvents during heating: it carries EN 1539 safety technology and a documented temperature-dependent solvent limit (172 g down to 18 g) with a 250 exhaust flow. For solvent-free loads, a non-LS NA model avoids the exhaust and solvent-limit constraints.
The exhaust path must sustain the stated flow rate of 250 continuously during solvent operation, since the solvent limits are tied to it. Ducting, extraction fan capacity and any building ventilation interface should be sized and verified before the first solvent-containing batch is run.
The 500-series controller functions, MyNabertherm app connectivity, process data storage, AMS2750F uniformity instrumentation and recorder options are documented as selectable controller and instrumentation options. They must be specified at ordering; the furnace alone does not include them.