Shimadzu TOC L Series
Shimadzu ·Made in Japan · In Active Production

Shimadzu TOC L Series

Shimadzu TOC L Series — TOC analyzers.

The TOC L Series is a combustion-type total organic carbon analyzer with a vertical tube furnace and NDIR detection, available in standalone and PC-controlled configurations. It covers a working range from 4 µg/L to 35,000 mg/L IC and 5 µg/L to 10,000 mg/L TN, with model-dependent features such as built-in LCD, USB data export, and autosampler compatibility. It is used for laboratory water quality monitoring, including priority sampling with autosamplers and COD/BOD conversion.

Full specifications

Core Performance & Functional Specifications
Measurement range 4 μg/L - 25,000 mg/L TC:
Capacity mfr 120 – 320 g (AP W Series)
Response time for trace measurements (as low as 1 mg) mfr approximately 2 s (as low as 1 mg)
Automatic door opening/closing time (AP W-AD model) mfr approximately 1 s (AP W-AD model)
CSN/CPN mfr 50
COD/BOD conversion mfr Function available (TOC-V CSH/CSN only)
Built-in LCD mfr Black and white
Number and type of data saved (standalone model) mfr 200 samples
Output of data saved (standalone model) mfr Not available
Printer mfr Built-in printer (thermal)
PC connection mfr 232 C
Autosampler cover mfr Fitted cover
Kit for small sample volumes mfr 8 mL
Halogen scrubber for high-salt samples (B-type halogen scrubber) mfr No option available
Dimensions, Weight & Mechanical Design
Weight TOC main unit: W440 x D525 x H460 mm,Approx. 40 kg TN unit added: W600 x D525 x H460 mm, Approx. 48 kg
External dimensions, weight mfr TOC main unit: W440 x D525 x H460 mmï1⁄4ŒApprox.
Electrical, Power & Drive Specifications
Power consumption . 20,000 Wh Approx
Power consumption (Weekly integrated value) Note 2 mfr . 20,000 Wh Approx
Interfaces, Connections & Control
Minimum display mfr 0.1 mg (AP W Series)

Description

Construction and series positioning

The TOC L Series is a combustion TOC analyzer platform. This model supports both standalone and PC operation, with a TFT color LCD on some variants and USB connectivity for data export. The series offers a 36% reduction in weekly power consumption compared to earlier models (approx. 12,700 Wh vs 20,000 Wh). ### Behavior under load and media

Integration & Operating Notes

Process Limits
High-salt samples require dedicated combustion tube and scrubber

Samples with high salt content can rapidly degrade the standard combustion tube and cause halogen interference. The high-salt sample combustion tube kit and B-type halogen scrubber must be installed for such matrices; otherwise, measurement drift and tube failure may occur.

Factory Specification Sheet, extra specs: High-salt sample combustion tube kit, Halogen scrubber for high-salt samples
Control & I/O
Carrier gas problem detection prevents invalid results

The instrument can detect carrier gas supply problems and should be configured to flag or halt measurements when gas pressure or flow is out of range. This prevents reporting data from incomplete combustion. The detection function is available on this model and should be enabled in the method settings.

Factory Specification Sheet, extra specs: Carrier gas problem detection
Maintenance
Regular replacement of combustion tube and scrubber is mandatory

The combustion tube and halogen scrubber are consumables that degrade with use, especially with high-salt samples. Replace them according to the maintenance schedule in the instruction manual. Failure to do so leads to baseline noise, low recovery, and potential damage to the NDIR detector. Keep spare kits on hand for uninterrupted operation.

Instruction Manual, Maintenance Section (typical for combustion TOC analyzers

Selection & Integration Q&A

When should I choose this TOC L Series model over a sibling variant like the CSH or CPH?

Choose this model if you need a wide measurement range (IC up to 35,000 mg/L) and options like high-salt kits and USB data export. Select a CSH/CSN variant if you require a built-in LCD and printer; choose CPH/CPN for PC-controlled operation with validation support.

What are the requirements for connecting this analyzer to a PC or network?

The instrument supports RS232C or USB connection to a PC. For standalone models, data can be exported to a USB memory stick. Ensure the PC has the appropriate driver and software (e.g., for CPH/CPN validation). A USB printer can be connected for direct printing.