Shimadzu UV 2600i2700i Plus
Shimadzu UV 2600i2700i Plus — Spectrophotometers.
The Shimadzu UV-2600i/2700i Plus is a double-beam UV-Vis spectrophotometer with a sealed optical bench and integrated PC control. It covers a 185–900 nm range with high photometric accuracy, distinguishing it from single-beam or extended-range variants. It is used for quantitative absorbance and transmittance measurements in pharmaceutical, chemical, and environmental laboratories.
Full specifications
Description
Construction and positioning
- Double-beam optics with a sealed, nitrogen-purgeable sample compartment reduce drift and enable measurements below 200 nm. - The 2600i/2700i Plus configuration includes a built-in PC and UVProbe software, unlike base models that require an external computer. ### Optical performance
- Wavelength range: 185–900 nm, covering far-UV to near-IR. - Photometric range: −4 to 4 Abs, with accuracy of ±0.002 Abs at 1 Abs and ±0.004 Abs at 2 Abs. - Stray light: ≤0.02% at 220 nm (NaI) and ≤0.02% at 340 nm (NaNO2). - Baseline stability: ±0.0003 Abs/h (after 2 h warm-up). ### Installation and operation
- Requires a stable AC power supply (100–240 V, 50/60 Hz) and a grounded outlet; avoid lines with heavy noise. - The sample compartment can be purged with nitrogen for measurements below 200 nm; purge pressure should be 0.02–0.05 MPa. - Operation via UVProbe software on the integrated PC; data export via USB or Ethernet. - The instrument should be installed in a dust-free, vibration-free environment with ambient temperature 15–35 °C and humidity 20–80% RH (non-condensing).
Integration & Operating Notes
The optical bench is sensitive to vibration and tilt; install on a rigid, level surface away from centrifuges, pumps, or other vibration sources. Exceeding ±1° tilt may misalign optics and degrade baseline stability.
Connect to a properly grounded outlet with voltage within 100–240 V AC ±10%. Avoid sharing the circuit with high-power equipment; voltage fluctuations beyond ±10% can cause lamp instability and measurement errors.
For measurements below 200 nm, purge the sample compartment with nitrogen (99.99% purity) at 0.02–0.05 MPa. Insufficient purge causes absorption by oxygen, leading to inaccurate results.
The deuterium and tungsten lamps have a typical life of 2,000 hours. Replace when the lamp intensity monitor shows a significant drop or when baseline noise increases. Always power off and unplug before replacement.
The integrated PC communicates with the spectrophotometer via USB. If the connection is lost during a measurement, the run aborts and data may be lost. Use a ferrite core on the USB cable and avoid hot-plugging.
Selection & Integration Q&A
Choose this double-beam model for high-stability measurements and low stray light (≤0.02% at 220 nm). For routine work above 220 nm, a single-beam model may suffice; for wavelengths beyond 900 nm, select an extended-range variant.
It requires a grounded 100–240 V AC, 50/60 Hz supply and a PC with USB port. The integrated PC runs UVProbe; ensure the outlet is not shared with noisy equipment and use a ferrite core on the USB cable.
Operate at 15–35 °C and 20–80% RH (non-condensing), away from vibration and dust. Continuous operation is supported, but allow 2 hours warm-up for baseline stability; avoid rapid temperature fluctuations.
Replace lamps after ~2,000 hours or when intensity drops. Base delivery excludes nitrogen regulator, tubing, and replacement lamps; order these separately. Regular cleaning of the sample compartment is recommended.