Waters PATROL UPLC Process Analysis System
Waters ·Made in USA · In Active Production

Waters PATROL UPLC Process Analysis System

Waters PATROL UPLC Process Analysis System — Chromatographs with Flow rate 2 milliliter/minute, Max pressure 0.713 megapascal.

The PATROL UPLC Process Analysis System is a full-loop injection chromatograph with binary or quaternary solvent delivery, up to 32 sample positions, and a 1:100 dilution range. It operates at flow rates up to 2.0 mL/min and 103.4 psi, with column temperature up to 90 °C and injection volumes of 1, 2, or 5 μL. It is intended for automated process and control sample analysis in a system, providing inject-to-inject cycle times under 15 seconds in binary configuration.

Flow rate
2 milliliter/minute
Max pressure
0.713 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 2.0 mL/min
Max pressure 103.4 psi
Column temperature 90.0 °C
Injection volume 5 μL
Total system delay volume mfr <120 μL with binary configuration, with standard
Cycle time mfr <15 s inject-to-inject with binary configuration
Number of solvents mfr Up to four, in any combination of two: A1 or A2,
Flow accuracy mfr ±1.0% at 0.500 mL/min, as per Waters AQT/Systems
Flow precision mfr 0.075 %RSD or ±0.01 min SD, (0.500 to 2.000 mL/m
Priming mfr Wet priming runs at a flow rate of 4 mL/min
Pump compositional accuracy mfr ±0.5% absolute (full scale), from 5% to 95% of f
Pump compositional precision mfr 0.15 %RSD or ±0.04 min SD, whichever is greater,
Maximum sample capacity mfr 32 available sites for process and control sampl
Internal sample capacity mfr 32 positions for standard 8 mL vials
External sampling mfr 1 port standard
Number of sample injections mfr Up to 60 injections per sample vial
Injection mode mfr Full-loop injection only
Injection loop volumes mfr 1 μL, 2 μL, or 5 μL sample loops only
Dilution range mfr Up to 1:100
Injection precision (without dilution) mfr < or =0.5 %RSD for caffeine peak area, full-loop
Injection precision with dilution mfr < or =1.0 %RSD for caffeine peak area, full-loop
Dilution linearity mfr 1 to 50 dilution factor > or =0.999 coefficient
Sample carryover mfr 1 to 50 dilution factor < or =0.01% of the previ
Dimensions, Weight & Mechanical Design
Width 762 mm
Electrical, Power & Drive Specifications
Power consumption 450 VAC
Materials, Media & Process Compatibility
pH range mfr pH 2 to 10

Description

Construction and positioning

Integration & Operating Notes

Electrical
Requires 450 VAC supply with adequate current capacity

The system is rated for 450 VAC; verify that the facility supply can deliver this voltage and the required current without significant voltage drop, as insufficient supply may cause erratic pump or autosampler operation. Ensure proper grounding and circuit protection according to local electrical codes.

Factory Specification Sheet
Process Limits
pH must remain between 2 and 10 to avoid column and system damage

Operating outside the pH range of 2 to 10 can degrade the analytical column and wetted parts, leading to loss of performance and potential leakage. Verify sample and mobile phase pH before introduction, and use appropriate buffers if necessary.

Factory Specification Sheet
Process Limits
Maximum pressure of 103.4 psi must not be exceeded

Exceeding the maximum pressure rating of 103.4 psi can cause seal failure, tubing rupture, or unsafe operation. Monitor system pressure during method development and ensure that column backpressure and flow rate combinations stay within this limit.

Factory Specification Sheet
Control & I/O
External sampling port available for connection to process line

One external sampling port is standard, allowing connection to a process stream or external sample source. Ensure that the sample is compatible with the system's wetted materials and that the pressure and flow are within the specified limits.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the PATROL UPLC Process Analysis System with binary configuration versus quaternary configuration?

Choose binary for faster cycle times (<15 s inject-to-inject) and lower delay volume (<120 μL). Choose quaternary for up to four solvents and higher delay volume (<400 μL) with cycle time <30 s. Select based on required solvent flexibility and throughput.