Waters SFC Prep 150 AP System
Waters ·Made in USA · In Active Production

Waters SFC Prep 150 AP System

Waters SFC Prep 150 AP System — Chromatographs with Flow rate 150 milliliter/minute, Max pressure 30 megapascal.

The Waters SFC Prep 150 AP System is a preparative supercritical fluid chromatography (SFC) platform configured for high-throughput purification, with a 150 mL/min flow ceiling and 300 bar maximum pressure. It distinguishes itself within the SFC Prep series by integrating automated sample and fraction handling for up to 15 microplates and 480 test tubes, enabling unattended method development and scale-up. The system is intended for laboratory-scale purification and analysis, bridging analytical SFC and production-scale chromatography.

Flow rate
150 milliliter/minute
Max pressure
30 megapascal

Full specifications

Core Performance & Functional Specifications
Flow rate 150 mL/min
Max pressure 300 bar
Column temperature 70 °C
Injection volume 1000 µL
Recovery mfr No Ketoprofen detected, therefore carry over is
Pressure rating mfr 10,000 psi (680 bar)
Maximum operating pressure mfr 8700 psi (592 bar)
Flow range mfr 20 to 200 g/min
Wetted materials mfr 316 stainless-steel, 440C, 17-4 PH stainless-ste
Number of solvents mfr Four solvents selected by the Solvent Select Val
Solvent conditioning mfr Sparging one channel per solvent
Operating flow rate range mfr 0.50 to 150.00 mL/min, in 0.01 mL/min increments
Maximum operating pressure mfr 6000 psi up to 100 mL/min, roll-off to 5000 psi
Plunger seal wash mfr Integral, active, programmable
Delay volume mfr <6.5 mL
Compressibility compensation mfr Automatic and continuous
Priming mfr Self prime from solvent container on floor with
Compositional accuracy mfr ±3% absolute (full scale) from 5% to 95%; and fr
Compositional precision mfr Retention time variation <0.15 min SD (uracil/ca
Sample capacity mfr Injection: Up to 15 microplates (up to 384 well)
Syringe delivery accuracy mfr >99.0% at full stroke with 1000 μL syringe
Dimensions, Weight & Mechanical Design
Height 1020 mm
Width 959 mm
Operating Conditions, Environment & Protection
Cooling mfr Circulating coolant
Ambient operating temperature mfr 15 to 40 °C
Ambient operating humidity mfr 20 to 80% relative humidity at 40 °C

Description

Construction and Positioning

The SFC Prep 150 AP System is a benchtop preparative SFC instrument (1020 mm H × 959 mm W) that combines a high-pressure pumping module, an autosampler with large-capacity sample and fraction decks, and a column oven. Unlike analytical SFC variants, this modification prioritizes sample throughput and fraction collection capacity: it accepts up to 15 microplates (384-well), 72 vials, or 544 test tubes for injection, and collects into up to 480 test tubes, 12 deep-well plates, or 128 28 mm vials. The system uses four solvents selected by a solvent select valve (SSV) with sparging on one channel per solvent, and employs circulating coolant for column temperature control up to 70 °C. ### Behavior Under Load and in Different Media

The pump delivers 0.50–150.00 mL/min in 0.01 mL/min increments, with a maximum operating pressure of 6000 psi (414 bar) up to 100 mL/min, rolling off to 5000 psi (345 bar) at 150 mL/min. The pressure rating of 10,000 psi (680 bar) and maximum operating pressure of 8700 psi (592 bar) apply to the high-pressure plumbing, not the system’s rated flow envelope. Compositional accuracy is ±3% absolute (full scale) from 5% to 95% and across 1–150 mL/min for methanol/water step gradients at 600 psi; compositional precision shows retention time variation <0.15 min SD (uracil/caffeine) across 1, 5, 20, 50, and 150 mL/min. The system self-primes from solvent containers on the floor with dry lines, and automatic, continuous compressibility compensation maintains flow stability. Wetted materials include 316 stainless steel, 440C, 17-4 PH stainless steel, sapphire, GFPM, PTFE, and aluminum, compatible with common SFC modifiers and CO2. ### Installation, Connection, and Operation

The system requires an ambient operating temperature of 15–40 °C and 20–80% relative humidity at 40 °C. It features an integral, active, programmable plunger seal wash to extend seal life, and a delay volume of <6.5 mL. The autosampler syringe (1000 µL) delivers >99.0% accuracy at full stroke. Recovery testing with ketoprofen showed no detectable carryover (<0.01%). The system is controlled via Waters Empower software (not detailed in provided facts), and installation requires connection to solvent lines, coolant supply, and waste.

Integration & Operating Notes

Process Limits
Do not exceed 5000 psi at 150 mL/min

The pump’s maximum operating pressure rolls off from 6000 psi at ≤100 mL/min to 5000 psi at 150 mL/min. Exceeding this curve risks seal failure, overpressure faults, and invalid results. Always verify method pressure at the intended flow rate before injection.

Factory Specification Sheet
Maintenance
Enable plunger seal wash to extend pump seal life

The integral, active, programmable plunger seal wash must be primed and programmed with an appropriate wash solvent (e.g., water/isopropanol) to prevent buffer salt accumulation and seal abrasion. Neglecting this leads to premature seal failure and pressure instability.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the SFC Prep 150 AP System over other SFC Prep series models?

Choose this model when you need automated high-throughput purification with up to 15 microplates or 544 test tubes for injection and 480 test tubes for collection, and flow up to 150 mL/min. For lower sample capacity or analytical-scale work, consider smaller variants in the series.

What are the requirements for connecting this system to my existing CO2 supply and fraction collection?

The system self-primes from floor-level solvent containers and uses four solvents via SSV. It requires a CO2 supply with appropriate pressure and a coolant source for the column oven. Fraction collection is integrated for test tubes, deep-well plates, or vials; ensure compatible racks and waste lines.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation