Safety warnings, EEE disposal and regulatory notices
Overviewpp. 19-30
ScreenTape architecture and TapeStation product description
Overviewpp. 31-36
Differences between TapeStation models and variants
Procedurepp. 37-54
Unpacking, hardware installation and setup steps
Procedurepp. 55-66
TapeStation Controller software settings, diagnostics and security
Procedurepp. 67-96
Running samples, good measurement and pipetting practices
Procedurepp. 97-114
Mixing, centrifugation and sample preparation recommendations
Specpp. 115-120
RINe algorithm functional range and interpretation
Procedurepp. 121-138
System verification, diagnosis and troubleshooting checks
Procedurepp. 139-195
Preventive maintenance, electrode cartridge exchange and cleaning
Procedurepp. 196-252
Error scenarios, sizing issues and communication problems
pp. 253-256
Supplementary reference information and index
p. 30table, 14 rows
p. 132table, 2 rows
p. 228table, 4 rows
p. 68table, 8 rows
p. 122table, 8 rows
p. 136table, 2 rows
p. 212table, 6 rows
p. 213table, 7 rows
p. 215table, 4 rows
Table 2 Physical Specifications
Type
Specification
Weight
25.5 kg (56.3 lbs) (G2991A)
24.8 kg (54.7 lbs) (G2991B)
Dimensions (w×d×h)
435 x 510 x 445 mm (17.1 x 20.0 x 17.5 inches)
Line voltage
100 – 240 VAC ± 10 %
Line frequency
50 or 60 Hz ± 5 %
Power consumption
50 VA / 50 W
Interface
USB (Instrument: USB-B; PC: USB-A)
Ambient operating temperature
10 – 40 °C (50 – 104 °F)
Operating humidity
15 – 80 %, non-condensing
Storage temperature
-40 to 70 °C (-40 to 158 °F)
Operating altitude
Up to 3000 m (9842 ft)
Safety standards
IEC, EN, CSA, UL, Overvoltage category II, Pollution degree 2
For indoor use only.
ISM Classification
ISM Group 1 Class A
According to CISPR 11
Sound pressure
<70 dB (A) According to ISO 7779:1988/EN 27779/1991
Assay specific temperature range
12 – 37 °C (54 – 99 °F) for D1000 ScreenTape
17 – 37 °C (63 – 99 °F) for High Sensitivity D1000 ScreenTape
15 – 30 °C (59 – 86 °F) for Genomic DNA ScreenTape
14 – 30 °C (57 – 86 °F) for RNA and High Sensitivity RNA ScreenTape
15 – 30 °C (59 – 86 °F) for D5000 ScreenTape
15 – 30 °C (59 – 86 °F) for High Sensitivity D5000 ScreenTape
16 – 30 °C (61 – 86 °F) for Cell-free DNA ScreenTape
NOTE This is an ISM Group 1 Class A product intended for use in industrial environment. In a domestic environment, this product may cause radio interference, in which case the user may be required to take adequate measures. NOTE The assay specific temperature range is measured at the ScreenTape nest. Choose the location of the instrument and the ambient conditions to fit to the working range of the assay else the start of runs will be disallowed.
Table 6 Computer Environment
System Test Suite
Troubleshooting Test Suite
Computer Specifications
Checks the computer specifications and configuration.
Yes
Yes
Software Environment
Checks the compatibility of TapeStation Controller Software with the computer.
Yes
Yes
Table 18 RNA - Possible causes of incorrect quantification in order of probability.
Root cause
Explanation
Solution
Insufficient Mixing
Buffer and sample was not sufficiently vortexed. The dye could not incorporate completely into the sample.
See “Mixing Recommendations” on page 100 and the Agilent Information Center for more information.
Sample concentration outside recommended range for application
Linear concentration estimation cannot be achieved for samples exceeding the limits in the specification table.
Either dilute or concentrate the sample until it is within the recommended range for the application as stated in the respective assay Quick Guide; then prepare a new run. Use the High Sensitivity assay if applicable (“Quantitative range” on page 114, Table 3 on page 68).
Incorrect pipetting technique or pipette calibration
Incorrect pipetting technique or uncalibrated pipettes can lead to poor pipetting performance which in turn alter sample/buffer ratio in an unpredictable manner.
See “Good Pipetting Practice” on page 91 and follow guidelines of the respective pipette vendor.
Sample not collected at bottom of vial
Insufficient volume pipetted by the robotics and applied to the lane may affect the migration behavior and alter the band shape.
Make sure to collect all liquid at the bottom of the vial by centrifugation. See “Centrifugation Recommendations” on page 103. Ensure that after removing strip caps and inserting the tube strips or 96-well sample plate into the TapeStation instrument liquid is still located at the bottom of the well, and is not spilled.
Table 3 ScreenTape Assay portfolio
ScreenTape Assay
Quick Guide Number
Agilent D1000 ScreenTape Assay
G2991-90031
Agilent High Sensitivity D1000 ScreenTape Assay
G2991-90131
Agilent RNA ScreenTape Assay
G2991-90021
Agilent High Sensitivity RNA ScreenTape Assay
G2991-90121
Agilent Genomic DNA ScreenTape Assay
G2991-90041
Agilent D5000 ScreenTape Assay
G2991-90051
Agilent High Sensitivity D5000 ScreenTape Assay
G2991-90151
Agilent Cell-free DNA ScreenTape Assay
G2991-90060
Table 5 Test suite groups
Test suite groups
Description
Computer Environment
Evaluation of computer specification and installation status.
Setup
Prepares the instrument, requests insertion of material, checks status LED functionality.
Information on current Maintenance counters, past Electrode Cartridges, past needle changes and current installed software is reported.
Vision
Camera and LED illumination are checked using the TapeStation Test Tape.
Electrophoresis
The voltage supply, current readout, and electrodes are checked using the TapeStation Test Tape.
Robotics
Robotics parts are checked by dedicated movements. Tip and ScreenTape device handling are checked using a full tip rack and one regular ScreenTape device.
Pipetting
This check procedure in the Hardware Troubleshooting visually checks pickup of liquid from 96-well sample plates and tube strips.
Cleanup
Resets the instrument to normal operational mode and requests removal of material.
Table 10 Robotics
System Test Suite
Troubleshooting Test Suite
Needle To Pump
Yes
Yes
Tape Handling
Yes
Yes
Tape Handling, Teach Pump-ML and Teach Pump-ML-Z represent tests which are only relevant for 4200 TapeStation instruments G2991A and G2991B.
Table 13 DNA - Possible causes of incorrect quantification results in order of probability
Root cause
Explanation
Solution
Insufficient Mixing
Buffer and sample was not sufficiently vortexed. The dye could not incorporate completely into the sample.
See “Mixing Recommendations” on page 100 and the Agilent Information Center for more information.
Sample concentration outside recommended range for application.
Linear concentration estimation cannot be achieved for samples exceeding the limits in the specification table.
Either dilute or concentrate your sample until it is within the recommended range for application as stated in the respective assay Quick Guide, then prepare a new run. Use the High Sensitivity assay if applicable (Table 3 on page 68).
Incorrect peak integration
The algorithm cannot establish a correct concentration estimation if marker or sample peaks are not completely encompassed as a signal from either marker or sample is left out of the calculation.
Ensure that all sample peaks and markers are integrated correctly in the TapeStation Analysis software, by clicking and dragging so that the whole peak is encompassed. See the Agilent Information Center for more information.
Incorrect analysis type
The Gel and Electropherogram view are designed to analyze discrete peaks within samples. The Region view is optimized for analyzing samples with broad peaks. Region view uses a different method to calculate concentration and sizing.
Use Gel view for samples with narrow peaks and the Region view for samples with broad peaks or smears such as libraries. See “Peak Maxima Versus Average Molecular Weight Sizing and Molarity” on page 111 for more information.
Sample and the marker relevant for quantification have merged.
Additional sample overlaying with the marker will alter the known concentration of the marker that is used for concentration calculation.
Ensure that your sample is within the recommended sizing range for the applications, see “Sizing Range” on page 113.
Contaminating beads can alter the run behavior of the sample and can introduce artifacts, see “TapeStation Analysis in the Agilent SureSelect and HaloPlex Workflow” on page 226.
If using the Sure Select Protocol, ensure that all AMPure beads are removed by increasing the time that the sample is on the magnetic plate.
Table 13 DNA - Possible causes of incorrect quantification results in order of probability
Root cause
Explanation
Solution
Sample not collected at bottom of vial
Insufficient volume pipetted by the robotics and applied to the lane may affect migration behavior and alter the band shape.
Make sure to collect all liquid at the bottom of the vial by centrifugation. See “Centrifugation Recommendations” on page 103. Ensure that after removing strip caps and inserting the tube strips or 96-well sample plate into the TapeStation instrument, liquid is still located at the bottom of the well, and is not spilled.
Incorrect protocol used
Differences exist between Standard and High Sensitivity assay protocols.
Ensure that the correct sample protocol for the application is used, see “Use the Correct Protocol” on page 99.
Incorrect reagents used
Reagents from different assays may differ in dye and marker concentration. Furthermore markers used are different and upper markers may not be present.
Ensure that the correct reagent buffer for the application is used, see “Use the Correct Protocol” on page 99.
Sample prepared too long before analysis
Evaporation or settling may alter the sample to buffer ratio.
Ensure that your run is started immediately after sample preparation.
The sample/buffer mix may evaporate over time if a 96-well sample plate was used which was not sealed with the recommended seal foil.
Always cover with the recommended foil seal (PN 5067-5154) and ensure tight sealing throughout the 96-well sample plate.
Incorrect pipetting technique or pipette calibration.
Incorrect pipetting technique or uncalibrated pipettes can lead to poor pipetting performance which in turn alter sample/buffer ratio in an unpredictable manner.
See “Good Pipetting Practice” on page 91 for more information or follow guidelines of the respective pipette vendor.
Incorrect Marker peaks picked up
Depending on the sample pattern, wrong peaks can get assigned as marker if running very close to the actual marker.
Use the Aligned button to detect the correct markers. Compare to aligned/unaligned data of good, for example the ladder, and questionable samples, then assign correct markers by right-clicking. See the Agilent Information Center for more information. Ensure that all sample peaks are within the recommended sizing range for the application.
Table 14 DNA - Possible causes of incorrect sizing results in order of probability
Root cause
Explanation
Solution
Incorrect Marker peaks picked up
Depending on the sample that is used, wrong peaks can get assigned as marker if the sample size is very close to the actual marker.
Use the Aligned button in the TapeStation Analysis software to detect the correct markers. Compare to aligned/unaligned data of good and questionable samples; then assign correct markers by right-clicking. See the Agilent Information Center for more information. Ensure that all sample peaks are within the recommended sizing range for the application.
Incorrect analysis type
The Gel and Electropherogram view are designed to analyze discrete peaks within samples. The Region view is optimized for analyzing samples with broad peaks. Region view uses a different method to calculate concentration and sizing.
Use Gel view for samples with narrow peaks and the Region view for samples with broad peaks or smears such as libraries. See “Peak Maxima Versus Average Molecular Weight Sizing and Molarity” on page 111 for more information.
Issues with the ladder lane
Any issues with the run ladder lane will be annotated within the TapeStation Analysis software sample table under Observations.
Manually insert or delete affected peaks. In assays where an electronic ladder is available (D1000, High Sensitivity D1000, and Cell-free DNA Assay), insert the electronic ladder to regain sizing information, see the Agilent Information Center for more information. In assays where an electronic ladder is not available (D5000, High Sensitivity D5000, and Genomic DNA Assay), re-run affected samples and ensure that Good Measurement Practices are followed, see “Good Measurement Practices” on page 87. For Genomic DNA Ladder considerations, see “Genomic DNA Assay” on page 210.
Insufficient migration due to presence of salt
Sample matrices exceeding the maximum buffer strength specifications may alter currents in an unpredictable manner leading to short running lanes, see “Maximum Buffer Concentration in Sample” on page 116.
Dilute your samples to ensure low levels of buffer salt. True migration profiles can be seen by unaligning your gel image by pressing the Aligned button in the TapeStation Analysis software. Please refer to the salt tolerance guidelines for your assay in the appropriate Quick Guide (Table 3 on page 68).
Engineering Facts
13 facts · every claim linked to a page
Selection
The G2991A model requires a trained service engineer for Electrode Cartridge exchange, while G2991B and G2992A allow user exchange, which affects maintenance planning and model choice.
Determines whether on-site user maintenance is possible or a service contract is needed.
Source
“Only G2991B and G2992A posses an Electrode Cartridge with lever mounting suitable for the exchange by users during maintenance. For model G2991A the Electrode Cartridge can be exchanged only by a trained service engineer during a scheduled preventive maintenance service.”
Installation
The instrument requires 75 mm clear space around it and 250 mm additional height above the listed dimensions to open the lid, so bench layout must account for this.
Prevents installation in spaces that block airflow, connections, or lid operation.
Source
“Allow approximately 75 mm (3 inches) of clear space around the instrument for air circulation and electrical connections.”
Installation
Two plug sockets must be available near the bench for the instrument and a laptop, which must be planned in the site requirements.
Ensures adequate power outlets are available before installation.
Source
“Two plug sockets must be available near the allocated bench space for the instrument itself and a laptop.”
Installation
If shipment conditions are unknown or outside operating conditions, wait at least 5 hours before unpacking to allow temperature and humidity equilibration and avoid condensation damage.
Prevents condensation-induced electronics damage during commissioning.
Source
“If actual shipment conditions are unknown or were outside operating conditions, wait for at least 5 h before unpacking to allow the instrument to adjust to the ambient temperature and humidity of the operating environment.”
Installation
The instrument is heavy and requires a second person for lifting during unpacking to prevent injury.
Ensures safe handling and prevents back strain or dropping during installation.
Source
“Because the instrument is heavy, lifting it will require the assistance of a second person to prevent injury.”
Operation
The ladder must always be loaded in location A1 on the tube strip holder for analysis.
Incorrect ladder placement will cause run failure or invalid results.
Source
“Always load the ladder in location A1 on the tube strip holder.”
Operation
Running samples from the 96-well plate and tube strips in one analysis is not supported; all samples must be combined on one format.
Prevents run setup errors that would require re-running samples.
Source
“Running samples from the 96-well sample plate and from sample tube strips in one analysis is not supported. Combine all samples you want to run on one instrument either to the 96-well sample plate or to the tube strips.”
Performance
Insufficient mixing degrades results; the recommended protocol is vortex mixing at 2000 rpm for 1 minute followed by brief centrifugation.
Ensures reproducible electropherogram and gel results by following validated mixing conditions.
Source
“The green trace shows the recommended protocol of vortex mixing using the IKA vortexer and adaptor at 2000 rpm for 1 minute followed by brief centrifugation.”
Performance
Electropherogram view reports the highest point of a peak as size, while region view reports the center of the region's mass, so sizing interpretation depends on the selected view.
Prevents misinterpretation of sizing and molarity results depending on analysis mode.
Source
“Electropherogram view is designed for use with discrete peaks, and the default size reported is that of the highest point of the peak.”
Maintenance
After an Electrode Cartridge change, a System Validation test is recommended for good laboratory practice even though it is not mandatory to continue working.
Guides maintenance workflow to confirm instrument performance after part replacement.
Source
“This additional system validation is not mandatory to continue working with the instrument but is recommended for good laboratory practice reasons.”
Safety
The instrument is not completely disconnected from mains when the power switch is off, so the power cable must be unplugged for full isolation.
Ensures safe servicing by fully removing power before maintenance.
Source
“The apparatus is not completely disconnected from the mains supply when the power switch is in the Off position”
Safety
Bearers of pacemakers or implanted defibrillators must stay at least 55 mm from the magnets, and magnets must be kept 25 mm from sensitive devices.
Prevents serious injury to personnel with implanted medical devices during installation and operation.
Source
“Bearers of heart pacemakers or implanted defibrillators must stay off at least 55 mm from the magnets.”
Cumplimiento
The product is ISM Group 1 Class A and may cause radio interference in domestic environments, requiring adequate measures by the user.
Informs site selection and EMC compliance planning for the laboratory environment.
Source
“This is an ISM Group 1 Class A product intended for use in industrial environment. In a domestic environment, this product may cause radio interference, in which case the user may be required to take adequate measures.”