Operation Manual

Contenido

  1. Warning p. 2
    Legal and safety notices for the instrument
  2. Overview pp. 3-4
    Table of contents and ATR analysis overview
  3. Overview pp. 5-8
    Instrument introduction and general description
  4. Procedure pp. 9-22
    Setup, power connection, desiccant, software startup, and validation
  5. Procedure pp. 23-34
    Sample analysis, reference caps, cleaning, and ATR technique
  6. Procedure pp. 35-38
    Battery charging, desiccant recharge, and maintenance procedures
  7. Ordering pp. 39-40
    Spare parts list and ordering information
  8. Spec pp. 41-44
    Technical specifications and safety information
  9. pp. 45-52
    Diamond ATR crystal, MAC address access, and library file transfer
  10. p. 21 table, 7 rows
  11. p. 41 table, 15 rows
  12. p. 9 table, 8 rows
  13. p. 39 table, 23 rows
  14. p. 24 table, 3 rows
  15. p. 29 table, 2 rows

Table 1 Diagnostic values

ValueOptimal (green)Marginal (yellow)Critical (red)Comment
Energy (centerburst)29,000 - 15,000>29,000 or <15,000>31,000 or <10,000Indicates the overall alignment and proper gain adjustment for the system.
Battery lifeAC or >30 min5 - 30 min<5 min
Source1.9 A>2.2 A or <1.6 A>2.5 A or <1.0 AIndicates a problem with the source control voltage or that the source has burnt out. Both voltage and amperage are displayed; however, amperage provides a sufficient diagnostic.
Laser signal15,000 - 3,000>15,000 or <3,000>17,000 or <2,000Can verify gross alignment errors even if the reference cap is not in place.
Detector temperature DTGS35 - 44<35 or >44<30 or >48Indicates a problem with the cooling circuitry or an ambient temperature over the specified range.
MCT-51 to -49<-51 or >-49<-53 or >-47The MCT detector is thermo-electrically cooled to approximately -50° C.
CPU temperature10 - 75>75>80Indicates the ambient temperature over the specified range.

Technical

CharacteristicValue
Interferometer geometryHigh-throughput Michelson interferometer with fixed and moving flat mirrors
Standard beamsplitterZinc selenide
Maximum spectral resolution4 cm-1 (both DTGS and MCT detectors)
LaserLow-powered solid state
SourceWire-wound element
DTGS Spectral range4500 to 650 cm-1
MCT Spectral range5000 to 1100 cm-1
Sample interfaceDiamond ATR, Germanium ATR, External Reflectance - 45°, Grazing Angle and Diffuse Reflectance
Detector1.3 mm diameter DTGS; 1.0 mm diameter MCT
Width10.16 cm (7.5 in)
Depth19.05 cm (7.5 in)
Height34.5 cm (13.6 in)
Weight1.88 kg (4.15 lbs) without batteries; 2.22 kg (4.9 lbs) with batteries
Power supply100 to 240 V AC, 3 A, 50 to 60 Hz
IEC/EN 60529IP54

Getting Started

РазделСтраница
Unpacking9
Components11
Connecting power12
Changing the desiccant16
Turning on the instrument17
Starting the software18
Communication to external PC19
Performance verification19

Table 4 Spare parts

Part numberDescription
G8180-67560Coarse Gold Reference Cap (Matte finish)
G8180-67562SS Mirror Reference Cap
G8180-67563Coarse Silver Diff Reference Cap
G8180-67564Gold Mirror Reference Cap
G8180-673004300 HH FTIR Polystyrene Ref Sample
G8180-671474300 HH FTIR Replacement Battery
G8180-670004300 HH FTIR Battery Charger
G8180-671354300 HH FTIR 15V DC Power Supply
G8180-676104300 HH FTIR MCT Retro Reflector
G8181-676114300 HH FTIR DTGS Retro Reflector
X3202-60106O-Ring Removal Tool
G8184-670004300 O-Ring Replacement Kit (PTFE encapsulated silicone O-ring)
G8184-67001O-Ring Replacement Kit (017 EPDM O-ring for external reflectance sample interface)
G8184-67002O-Ring Replacement Kit (013 EPDM O-ring for diffuse reflectance sample interface)
G8180-675144300 HH FTIR Desiccant Cartridge (includes silica gel desiccant)
430-0164USB cable 2 meter A to Micro B
G8180-680014300 HH FTIR D-ATR
G8180-680024300 HH FTIR Ge-ATR
G8180-680034300 HH FTIR Ext Ref
G8180-680044300 HH FTIR Diff Ref
G8180-680054300 HH FTIR GA Ref
G8180-200294300 Handheld FTIR wrist strap
250-01134300 Handheld FTIR stylus

For sampling accessories, upgrades, service contracts, repair and refurbishment services, contact Agilent or go to the Agilent website at www.agilent.com.

Table 2 Optimal uses of external reflectance sampling interfaces

Sampling InterfaceOptimal Uses
External reflectance• Suitable for analysis of smooth, opaque samples that reflect infrared light • Suitable for analysis of thin films and coatings greater than 1 micron in thickness on reflective metal surfaces
Grazing angle• Suitable for analysis of ultra-thin films less than 1 micron in thickness • Suitable for analysis of very low-level contamination (nanometer thickness level)
Diffuse reflectance• Suitable for analysis of samples that reflect little light

Table 3 Internal reflectance sampling interfaces

Sampling interfaceOptimal Uses
Diamond ATR• Ideal for analysis of solids, liquids, pastes and gel • Only the top 2-3 microns are analyzed
Germanium ATR• Ideal for analysis of highly absorbing solids and liquids, e.g. carbon filled elastomers and rubbers • Only the top 0.5 to 2 micrometers is measured

See Appendix A for more details on the Diamond ATR crystals.

Engineering Facts

14 facts · every claim linked to a page
Safety

The instrument is not intrinsically safe, so it must only be used in atmospheres verified free of flammable materials.

Determines whether the analyzer can be deployed in hazardous or potentially explosive field environments.

Source
“4300 Handheld FTIR systems are NOT intrinsically safe. Use the system only in atmospheres that have been tested for flammable materials.”
Selection

The standard DTGS detector covers 4500 to 650 cm-1 and the instrument tolerates 0 to 50 C and up to 95% humidity.

Defines the spectral coverage and environmental envelope to match against the intended application.

Source
“The instrument comes with a DTGS detector capable of covering the spectral range of 4500 to 650 cm-1.”
Installation

After delivery the shipping case must be left at room temperature for several hours before opening to avoid condensation.

Prevents condensation damage to optics during initial setup and installation.

Source
“place the shipping case in a room-temperature environment and allow several hours for the contents of the container to reach the room ambient temperature.”
Installation

The instrument must be kept away from hot surfaces and sources of electromagnetic interference.

Guides placement of the instrument to avoid thermal and EMI-induced measurement errors.

Source
“Keep the instrument away from hot surfaces and any sources of electromagnetic interference.”
Operation

Batteries are hot-swappable, allowing continuous use without shutdown, and four batteries give about 8 hours of runtime.

Enables planning of field runtime and battery logistics without interrupting measurements.

Source
“You can open the front of the battery compartment and swap out a battery while the system is in use.”
Operation

Batteries cannot be charged inside the instrument and require the dedicated external charger.

Requires provisioning of the external charger for field battery rotation.

Source
“Batteries can only be charged in the appropriate battery charger. They do not charge in the 4300 Handheld FTIR system.”
Installation

Only the supplied 15 V DC power supply may be used, and a UL-approved surge protection strip is required.

Prevents damage from incompatible supplies and power surges during AC operation.

Source
“Use only the included 4300 Handheld FTIR power supply. Use of other power supplies may damage the 4300 Handheld FTIR.”
Installation

The power cord must be connected to the spectrometer first, then to the outlet and computer.

Defines correct power-up wiring sequence to avoid startup damage.

Source
“Connect the power cord to the spectrometer system first and then to the power outlet and computer to prevent any power startup damage or problems.”
Maintenance

The desiccant cartridge is recharged in a lab oven at about 120 C until the silica gel returns to dark blue.

Specifies the desiccant regeneration procedure needed to protect the optics from moisture.

Source
“The recommended temperature is approximately 120°C. Drying time may vary dependent on the amount of moisture.”
Operation

The instrument needs a 5-minute warm-up after power-on for reliable analysis, and 30 minutes before performance validation tests.

Sets the minimum stabilization time before collecting trustworthy data.

Source
“Allow the spectrometer to warm up for a period of 5 minutes for reliable analysis.”
Installation

The integrated CPU screen must not be rotated beyond vertical, or the flat flex cable will be damaged.

Prevents a common handling error that damages the display cable.

Source
“The integrated CPU screen rotated beyond the vertical position will result in damage to the flat flex cable connected to the screen.”
Installation

The USB cable must be inserted at a slight upward angle and the instrument placed on its side when connected.

Avoids damage to the USB port during PC communication and software updates.

Source
“The USB cable must be inserted at a slight upwards angle into the instrument USB connector”
Performance

Diagnostic values should be checked daily or weekly, while full performance validation runs quarterly or bi-annually.

Establishes the routine verification schedule to keep measurements within specification.

Source
“Performance validation should be run on a quarterly or bi-annual basis.”
Maintenance

The exterior may be wiped with a damp cloth and mild detergent but must never be fully immersed in water.

Defines safe cleaning limits to avoid water ingress into the handle and trigger.

Source
“Do not completely immerse the 4300 Handheld FTIR in water.”