0.005 ng/μL – 2 ng/μL input DNA (adjustable by dilution of sample)
1 Results using DNA ladder or DNA fragment standards initially prepared with 1x TE Buffer
2 The 22 cm effective, 47 cm total length capillary array is only available for 12-capillary Fragment Analyzer system.
Gel preparation — Prepare gel/dye mixture for 5200 Fragment Analyzer system
# of samples to be analyzed1
Volume of Intercalating Dye
Volume of gel
12
1.0 µL
10 mL1
24
1.5 µL
15 mL
36
2.0 µL
20 mL
48
2.5 µL
25 mL
96
4.5 µL
45 mL
1 Typically one sample well per separation is dedicated to the ladder.
Engineering Facts
11 facts · every claim linked to a page
Selection
The kit's DNA sizing range is 100 bp to 6,000 bp, so samples outside this range cannot be sized with this kit.
Determines whether the kit is compatible with the expected fragment sizes of the CRISPR experiment.
Source
“DNA sizing range 100 bp – 6,000 bp”
Selection
Separation resolution depends on the capillary array: 10 bp at 500 bp for the 22-47 array and 8 bp at 500 bp for the 33-55/55-80 arrays.
Array choice directly affects the ability to resolve closely sized fragments near 500 bp.
Source
“Separation resolution 10 bp @ 500 bp (22-47 Array)2 8 bp @ 500 bp (33-55/55-80 Array)”
Selection
The 22 cm effective, 47 cm total length capillary array is only available for the 12-capillary Fragment Analyzer system.
Prevents selecting an incompatible array for non-12-capillary instruments.
Source
“The 22 cm effective, 47 cm total length capillary array is only available for 12-capillary Fragment Analyzer system.”
Performance
Input DNA concentration must be within 0.005 ng/μL to 2 ng/μL, adjustable by diluting the sample.
Samples outside this concentration window require dilution or concentration before analysis.
Source
“DNA fragment concentration range 0.005 ng/μL – 2 ng/μL input DNA (adjustable by dilution of sample)”
Installation
The 1x 930 dsDNA Inlet Buffer tray must be replaced daily.
Daily replacement is a mandatory maintenance step that affects run-to-run reproducibility.
Source
“Replace 1x 930 dsDNA Inlet Buffer daily.”
Maintenance
Capillary Storage Solution in row H must be replaced every 2-4 weeks.
Defines the capillary storage maintenance interval to prevent capillary degradation.
Source
“Replace Capillary Storage Solution every 2-4 weeks.”
Operation
The MDK DNA Ladder is ready to use without dilution and 24 µL must be placed in well 12 of each row of samples analyzed.
Ladder placement and volume are required for correct sizing and must not be diluted.
Source
“Place 24 µL of MDK DNA Ladder (“ready to use”; no dilution) into well 12 of each row of samples analyzed.”
Operation
The marker plate row A must be filled with 30 µL per well plus a mineral oil overlay.
Correct marker volume and oil overlay are needed for proper marker detection.
Source
“Fill row A (30 µL/well + mineral oil overlay).”
Operation
The CRP-910-(22, 33 or 55) – CRISP Discovery method must be selected from the dropdown menu in the software.
Using the wrong method leads to incorrect separation parameters and invalid results.
Source
“Select Add to Queue, select the CRP-910-(22, 33 or 55) – CRISP Discovery method from the dropdown menu.”
Safety
Reagent handling may pose health risks, so material safety datasheets must be consulted and protective equipment worn.
Mandates safety review and PPE before handling kit chemicals.
Source
“The handling of reagents and chemicals might hold health risks.”
Cumplimiento
The kit is for research use only and not for diagnostic procedures.
Restricts the kit to research applications and excludes clinical diagnostic use.
Source