Amino Acid Analysis How to guide
Technical Reference Summary
Covered Models & Document Scope
Solución integral Agilent AdvanceBio Amino Acid Analysis (AAA): instrumentación InfinityLab LC Series (1290/1260 Infinity II), columnas analíticas y de guarda AdvanceBio AAA LC, reactivos de derivatización OPA/FMOC, tampón borato, reactivo DTDPA, estándares de aminoácidos y consumibles (viales, tapones, septos).
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| AdvanceBio AAA LC column dimensions | 3.0 x 100 mm p.20 ↗ | 2.7 µm particle |
| AdvanceBio AAA LC column particle size | 2.7 µm p.20 ↗ | |
| AdvanceBio AAA guard column dimensions | 4.6 x 5 mm p.20 ↗ | 2.7 µm, 3/pk |
| Borate buffer concentration | 0.4 M p.20 ↗ | in water, pH 10.2, 100 mL |
| Borate buffer pH | 10.2 p.20 ↗ | 0.4 M in water |
| FMOC reagent concentration | 2.5 mg/mL p.20 ↗ | in ACN, 10 x 1 mL ampoules |
| DTDPA reagent quantity | 5 g p.20 ↗ | |
| AA standard concentration | 1 nmol/µL p.20 ↗ | 10 x 1 mL |
| AA standard quantity | 250 pmol p.20 ↗ | 10/pk |
| Vial volume | 2 mL p.20 ↗ | screw top, amber, certified, 100/pk |
| Mobile phase A composition | 10 mM Na2HPO4, 10 mM Na2B4O7 p.6 ↗ | pH 8.2 |
| Derivatization reaction temperature | Room temperature p.9 ↗ | RNH2 |
| Fluorescence detection excitation (primary) | 340 nm p.11 ↗ | 0.00 min, emission 450 nm |
| Fluorescence detection excitation (secondary) | 260 nm p.11 ↗ | 5.53 min, emission 325 nm |
| PMT gain | 10 p.11 ↗ | as needed |
| Gradient %B at 0.35 min | 2 % p.11 ↗ | |
| Gradient %B at 13.4 min | 57 % p.11 ↗ |
Operating Limits
- Mobile phase A pH: 8.2 p.6 ↗
- Borate buffer pH: 10.2 p.20 ↗
- Mobile phase replacement interval: every two days p.19 ↗
- Derivatization reagent / buffer / standards / wash water replacement: daily p.19 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| AdvanceBio AAA guard columns, 4.6 x 5 mm, 2.7 µm, 3/pk p.20 ↗ | 820750-931 | |||
| AdvanceBio AAA LC column, 3.0 x 100 mm, 2.7 µm p.20 ↗ | 695975-322 | |||
| AdvanceBio AAA guard columns, 3.0 x 5 mm, 2.7 µm, 3/pk p.20 ↗ | 823750-946 | |||
| Borate Buffer, 0.4 M in water, pH 10.2, 100 mL p.20 ↗ | 5061-3339 | |||
| FMOC Reagent, 2.5 mg/mL in ACN, 10 x 1 mL ampoules p.20 ↗ | 5061-3337 | |||
| Dithiodipropionic Acid (DTDPA) reagent, 5 g p.20 ↗ | 5062-2479 | |||
| Vial, screw top, amber with write-on spot, 2 mL, certified, 100/pk p.20 ↗ | 5182-0716 | |||
| Cap, screw, green, PTFE/white silicone septum, 100/pk p.20 ↗ | 5182-0721 | |||
| Vial, screw top, clear, flat bottom, for LC, 6 mL, certified, 100/pk p.20 ↗ | 9301-1377 | |||
| AA standard, 1 nmol/µL, 10 x 1 mL p.20 ↗ | 5061-3330 |