Induztr
Resumen técnico Visor de documentos Modelos relacionados

Determination of Monosaccharides in

Contenido

  1. Overview p. 2
    Background and purpose of monosaccharide analysis method
  2. Procedure p. 3
    Experimental setup, materials, and analytical procedure
  3. Spec p. 4
    Linearity assessment of the analytical method
  4. p. 5
    Summary of results and method conclusions
  5. p. 6
    Bibliographic references cited in the document
  6. p. 2 table, 9 rows
  7. p. 3 table, 16 rows
  8. p. 5 table, 8 rows

Table 1. Migration time (tM) and MS/MS acquisition parameters used for the identification and quantification of monosaccharides in grain.

MonosaccharidetM (min)Q1a (m/z)Q3b (m/z)CEc (V)FEd (V)
Inositol10.6179.187.117101
Fucose11.2163.1103.0470
Galactose11.5179.189.0570
Arabinose11.7149.089.1970
Glucose11.7179.189.0570
Rhamnose11.9163.1103.0470
Xylose12.0149.089.1970
Mannose + Fructose12.1179.189.0570
Ribose12.2149.089.1970

a Precursor ion (Q1), b Fragment ion (Q3), c Collision energy, d Fragmentor energy

Experimental

ParameterValue
CE Conditions
InstrumentAgilent 7100 CE system
Background electrolyte0.5 M TEA, pH 12.3
Applied voltage25 kV
CapillaryAgilent Fused-silica capillary, 20 μm id with 90 cm total length (p/n 160-2660-5, cut down to 90 cm)
Injection30 seconds at 100 mbar
Temperature25 ºC
MS Conditions
InstrumentAgilent 6430 Triple Quadrupole LC/MS
Ion modeESI, negative ionization
Sheath liquid10 mM TEA/methanol (50:50 v:v)
Flow rate6.0 µL/min
Capillary voltage4,500 V
Drying gas flow (N2)6 L/min
Drying gas temperature200 °C
Nebulizer pressure10 psi

Table 2. Linear range, regression equation, determination coefficient, standard deviation of residuals, limits of detection (LOD) and quantitation (LOQ) of the proposed method for monosaccharides in grain.

MonosaccharideY = a + bxR2Sy/xLOD (µM)LOQ (µM)
Fucosey = 0.044 + 1.336 x0.9980.0432.48.1
Galactosey = 0.097 + 0.285 x0.9990.0101.44.6
Arabinosey = -0.103 + 2.355 x0.9980.0791.96.2
Glucosey = 0.246 + 0.441 x0.9970.0151.96.2
Rhamnosey = -0.106 + 2.018 x0.9980.0593.411.3
Xylosey = -0.065 + 3.395 x0.9980.1122.89.3
Mannose + Fructosey = -0.064 ± 2.279 x0.9980.0761.55.0
Ribosey = -0.187 + 2.825 x0.9980.0991.23.8

a Intercept, b Slope, R2 = Determination coefficient, Sy/x = Standard deviation of residuals

Engineering Facts

14 facts · every claim linked to a page
Selection

The method requires a fused-silica capillary of 20 µm id and 90 cm total length, which must be selected to match the CE/MS interface.

Capillary dimensions directly affect separation efficiency, migration times, and compatibility with the MS source; wrong dimensions will degrade performance.

Source
“Agilent Fused-silica capillary, 20 μm id with 90 cm total length (p/n 160-2660-5, cut down to 90 cm)”
Installation

New capillaries must be preconditioned by flushing with NaOH, water, and BGE in a defined sequence before first use.

Proper preconditioning is essential to activate the capillary surface and ensure reproducible electroosmotic flow and migration times.

Source
“New fused silica capillaries were preconditioned by flushing 0.1 M NaOH solution (five minutes), deionized water (10 minutes), and BGE (10 minutes).”
Operation

Between runs, an additional 60-second BGE flush is required to maintain run-to-run reproducibility.

Without this intermediate flush, carryover and BGE depletion can cause drift in migration times and peak areas.

Source
“An additional flushing step with BGE for 60 seconds was included between the runs.”
Operation

Samples are injected hydrodynamically for 30 seconds at 100 mbar, which must be set precisely on the CE instrument.

Injection parameters control the amount of sample loaded; deviations affect sensitivity and peak shape.

Source
“Samples were introduced hydrodynamically in 30 seconds at 100 mbar and analyzed with an applied voltage of 25 kV.”
Operation

The BGE must be 0.5 M triethylamine at pH 12.3 to ionize the monosaccharides for negative-mode ESI-MS/MS.

The high-pH BGE is required for carbohydrate ionization; incorrect pH or concentration will prevent separation and detection.

Source
“All separations were performed at 25 °C using 0.5 mM TEA, pH 12.3 as the BGE.”
Operation

The sheath liquid for ESI must be 10 mM TEA in methanol/water (50:50 v/v) delivered at 6.0 µL/min.

Sheath liquid composition and flow rate are critical for stable electrospray and sensitive MS detection in negative ion mode.

Source
“Sheath liquid 10 mM TEA/methanol (50:50 v:v)”
Operation

The mass spectrometer must be operated in negative ionization mode with MRM for one specific transition per monosaccharide.

MRM in negative mode provides the selectivity and sensitivity needed to quantify underivatized monosaccharides at low µM levels.

Source
“The mass spectrometer was operated in negative ionization mode, using multiple reaction monitoring (MRM) mode for one specific transition.”
Operation

Sample hydrolysis uses 1.0 M H2SO4 at 90 °C for 150 minutes with manual stirring every 30 minutes.

Hydrolysis conditions determine the completeness of polysaccharide breakdown and the accuracy of monosaccharide quantification.

Source
“The tube was heated at 90 °C for 150 minutes and manually stirred every 30 minutes.”
Operation

Neutralization with Ba(OH)2 is necessary because simple neutralization does not reduce sample conductivity enough for CE.

High conductivity from residual H2SO4 causes poor electrophoretic separation; Ba(OH)2 precipitation of BaSO4 is required to make the sample compatible with CE.

Source
“A simple neutralization does not diminish the conductivity to an acceptable level, so the neutralization with Ba(OH)2 was carried out”
Operation

After neutralization, samples must be centrifuged at 6,000 rpm for one minute and filtered through a 0.2 µm PVDF/PP membrane.

Centrifugation and filtration remove BaSO4 precipitate and particulates that would otherwise clog the capillary and damage the MS.

Source
“the samples were centrifuged for one minute at 6,000 rpm and filtered through a 0.2 μm PVDF and PP membrane”
Performance

The method cannot fully resolve mannose and fructose, so they are quantified together as a single peak.

Engineers must know that mannose and fructose are reported as a combined result, which limits specificity for these two analytes.

Source
“However, the resolution was not sufficient to separate mannose and fructose.”
Performance

The method achieves LODs between 1.2 and 3.4 µM for all monosaccharides, suitable for trace-level grain analysis.

LOD defines the minimum detectable concentration and determines whether the method meets the required sensitivity for a given grain sample.

Source
“The proposed method enabled us to detect monosaccharides at levels between 1.2 to 3.4 µM.”
Performance

The total analysis time is under 14 minutes per sample, enabling high-throughput screening.

Short analysis time and low reagent consumption reduce operating cost and improve laboratory throughput.

Source
“spending less than 14 minutes per sample and uses a small amount of sample with low reagent consumption.”
Cumplimiento

The method was validated on soybean, corn, and coffee grain samples, demonstrating applicability to real food matrices.

Matrix validation confirms the method is fit for purpose in these food commodities and supports regulatory or quality-control use.

Source
“The applicability of this method for the analysis of real samples was demonstrated with soybean, corn, and coffee grain samples.”