Agilent JetClean Self-Cleaning Ion Source for GC/MS e-Book
Technical Reference Summary
Covered Models & Document Scope
Fuente de iones autolimpiante Agilent JetClean para GC/MS; disponible como opción en los nuevos sistemas GC/MS de cuadrupolo simple y triple de Agilent y como actualización para sistemas existentes: 5977A/B, 7000B/C/D, 7010A/B, 5977B con Intuvo, 7000D GC/TQ con Intuvo.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Cleaning gas | hydrogen p.2 ↗ | controlled flow |
| Manual source cleaning interval without JetClean | once every two weeks p.2 ↗ | user experience, non-Agilent GC/MS |
| Manual source cleaning interval with JetClean | once in nine months p.2 ↗ | user experience, Agilent GC/MSD with JetClean |
| Area %RSD, Benzo[a]anthracene | 1.8 %RSD p.4 ↗ | Day 1, n=12 |
| Area %RSD, Chrysene | 1.2 %RSD p.4 ↗ | Day 1, n=12 |
| Area %RSD, Benzo[b]fluoranthene | 1.9 %RSD p.4 ↗ | Day 1, n=12 |
| Area %RSD, Benzo[a]pyrene | 1.7 %RSD p.4 ↗ | Day 1, n=12 |
| Calibration linearity R², Benzo[a]pyrene | 0.9976 p.4 ↗ | Day 1 |
| MRM ratio, 252.1 -> 250.1 / 252.1 -> 252.1 | 56.2 (87.2 %) p.4 ↗ | Benzo[a]pyrene |
| MRM ratio, 250.1 -> 248.1 / 252.1 -> 252.1 | 8.7 (50.1 %) p.4 ↗ | Benzo[a]pyrene |
Operating Limits
- Compatible GC/MS systems (retrofit): 5977A/B, 7000B/C/D, 7010A/B, 5977B with Intuvo, 7000D GC/TQ with Intuvo p.2 ↗
- Cleaning gas: hydrogen p.2 ↗