Agilent 708-DS Dissolution Apparatus Operator's Manual

Contenido

  1. Overview pp. 6-8
    Product description and key features of the 708-DS
  2. Spec pp. 9-15
    Power, bench space, and technical specifications
  3. Procedure pp. 16-84
    Main screen, methods, menus, and operating procedures
  4. Procedure pp. 85-108
    Cleaning, maintenance schedules, probes, and printer care
  5. Services pp. 109-110
    Warranty, service contacts, accessories, and consumables
  6. Warning pp. 111-118
    Safety symbols, warnings, EEE, and radio interference
  7. pp. 119-120
    Agilent Technologies internet resources
  8. p. 14 table, 15 rows
  9. p. 73 table, 17 rows
  10. p. 15 table, 2 rows
  11. p. 59 table, 6 rows
  12. p. 50 table, 3 rows
  13. p. 66 table, 3 rows
  14. p. 83 table, 1 rows
  15. p. 99 table, 12 rows
  16. p. 100 table, 17 rows

Table 2: Performance specifications of the 708-DS Dissolution Apparatus (G7910A / G7911A)

TypeSpecificationComments
ConditionPerformance may vary depending on environmental conditions (temperature, humidity, altitude, etc.)
EvaporationLess than 1% evaporative loss under specific conditions
Water bath - Temperature rangeAmbient + 5-55 °C (41-131 °F)
Water bath - Probe accuracy± 0.1 °C (± 0.18 °F)
Spindle - Speed range10-250 RPM
Spindle - Speed Accuracy±1% over 25 RPM±2% 10-25 RPM
Spindle - Speed SelectionVia touch screen
SamplingManifold option available for automated sampling
Spindle shaft materialStainless steel
Drive unit liftMotorized drive
Optional featuresManual drive unit lift, Dosage Delivery Module (DDM), Autosampling, AutoTemp, resident sampling cannulas, handheld temperature probe
PrinterOptional internal thermal printer
Local User InterfaceLCD display with integral touch screen
Instrument Control via PCCan be controlled via Dissolution Workstation vers. 1.7 onwards, Cary WinUV Dissolution vers. 5.3 onwards and Dissolution Workflow Manager 1.0 SR1 software; current vers. FW of HW = 3.01
Maintenance and safety-related featuresSpindle hours are tracked in FW

Parameter Limits / Ranges

ParameterRangeResolutionConfigured
Spindle Speed10.0 - 250.0 RPM0.1 RPMMain Screen - Display Parameters on page 27 Run - Manual Operation on page 33 Method Properties Screen 1 on page 47 Menu Screen - Diagnostics Screen 1 on page 65
Temperature (Bath or Vessel)5 - 55 °C (41 - 133 °F)0.1 °C (0.18 °F)Main Screen - Display Parameters on page 27 Method Properties Screen 1 on page 47 Menu Screen - Diagnostics Screen 1 on page 65
Serial Number15 charactersFactory
Profile Print Interval000:01:00 to 024:00:001 secondRun - Manual Operation on page 33 Method Properties Screen 2 on page 48
Test Duration000:00:10 to 999:59:591 secondRun - Manual Operation on page 33 Method Properties Screen 3 on page 49
Final Spin Duration000:00:10 to 001:00:001 secondMethod Properties Screen 2 on page 48
Tester ID12 charactersMenu Screen - Instrument Settings Screen 1 on page 57
Comm ID01 to 9901Menu Screen - Instrument Settings Screen 1 on page 57
Media Volume100 to 2000 mL1 mLRun - Manual Operation on page 33 Method Properties Screen 1 on page 47
DDM Increment000:00:00 to 001:39:591 secondRun - Manual Operation on page 33 Method Properties Screen 2 on page 48
Temp Error Tolerance0 to ±0.5 °C1 secondMenu Screen - Instrument Settings Screen 2 on page 58
Speed Error Tolerance0 to 1.0 RPM1 secondMenu Screen - Instrument Settings Screen 2 on page 58
Manifold Down Time000:00:00 to 001:00:001 secondMethod Properties Screen 3 on page 49
Manifold Lead Time000:00:00 to 000:01:001 secondMethod Properties Screen 3 on page 49
Sample Point Alarm000:00:05 to 000:02:301 secondMethod Properties Screen 3 on page 49
Method Pause Time00:00 to 1:00:001 secondMenu Screen - Alarms on page 56
Timer Alarm00:00 to 24:00:001 secondMenu Screen - Alarms on page 56

NOTE: A value of 0 disables the tolerance check. (для строк Temp Error Tolerance и Speed Error Tolerance)

Specifications of the 708-DS Dissolution Apparatus (G7910A / G7911A)

TypeSpecificationComments
GLP featuresErrors can be tracked via print-out or in the software (DWS, DWM, Cary WinUV Dissolution); user management via local UI or software
HousingAll materials are recyclable

Menu Screen - System Menu

OptionFunction
Temperature TolerancesAllowable temperature variation from the set point. When alarms are enabled in Instrument Settings, deviations greater than this will cause an alarm but not stop the method.
Speed ToleranceAllowable spindle speed variation from the set point. When alarms are enabled in Instrument Settings, deviations greater than this cause an alarm but do not stop the method.
VesselsThis parameter sets the current installed vessel size (2000, 1000, 250, 200, or 100 mL). This value is used to determine various settings throughout the operation of the instrument but most importantly sampling manifold depth which can be programmed from the Calibration screen.
Number ActiveSets the number of vessels used in a test. The remainder of the vessels are ignored until this parameter is changed.
Enable AlarmsIf checked, all alarms are activated. If not checked, all alarms are suppressed on the user interface and on a printout if a printer is installed.
Enable DDMEnables the Dosage Delivery Modules. This setting overrides the method setting. If DDM is not enabled, the DDM Increment box is grayed out in Method Editor.

Menu Screen - System Menu

OptionFunction
Manifold Down DurationThis is the amount of time that the manifold remains at its sampling position. The sum of the Lead Time and Duration must be greater than or equal to 30 seconds for accurate temperature readings.
Enable Sample Point AlarmEnabling this option activates the sample notification dialog as well as activating the audible and visual alarms. This alarm draws the operator to the machine at the timepoint for manual sampling scenarios.
Alarm Lead TimeThe Timepoint Approaching screen (see Dialog Screens (Green / Blue) on page 75) is available to notify the user of upcoming timepoints. The time entered is how long before the timepoint the Timepoint Approaching screen displays. This alarm alerts the operator to an upcoming sample timepoint. Audible beeps will increase in frequency as the timepoint approaches.

From the Method Properties screen: • Press Cancel to exit without saving the values and return to the Select Method screen. • Press Back to save all values to permanent memory and return to the previous Method Properties screen. • Press Save to save all values to permanent memory and return to the Select Method screen. • Press Next to save all values to permanent memory and proceed to the next Method Properties screen.

Menu Screen - System Menu

OptionFunction
Dosage Delivery Module ControlThe DDMs are solenoid activated doors that hold a tablet above each vessel. Press the corresponding button to open a DDM. Press All to open all DDMs simultaneously.
DrivePress Drive to raise the drive unit. Keep pressing it to raise the drive unit to the maximum height. Release the button to stop the motion. Press Drive to lower the drive unit.
ManifoldPress Manifold to raise the manifold to the maximum height. Press Manifold to lower the manifold to the appropriate position based on the apparatus type and volume.

From the Diagnostics screen: • Press Next to proceed to the second Diagnostics screen. • Press Finish to return to the Menu screen.

Bath Heater Error

ActionHeater Probe Open or No Comm.Pump Failed or Heater OffBath Temperature 2 ºC Over SetpointBath OvertemperatureBath Probe OpenBath Probe ShortHeater Probe Short
• Ensure the bath heater cable is connected. • Check cable.• Contact Agilent Customer Care (Obtaining Warranty and Other Services on page 109).• Check for a faulty probe. • Contact Agilent Customer Care (Obtaining Warranty and Other Services on page 109).• Turn off the bath heater. • Contact Agilent Customer Care (Obtaining Warranty and Other Services on page 109).• Ensure the bath probe is plugged in. • Replace the bath probe.• Replace the bath probe. • Contact Agilent Customer Care (Obtaining Warranty and Other Services on page 109).• Contact Agilent Customer Care (Obtaining Warranty and Other Services on page 109).

Table 3: Resistance Value of Temperature Probe

Temp. [°C (°F)]Resistance [Ω]Temp. [°C (°F)]Resistance [Ω]Temp. [°C (°F)]Resistance [Ω]
-5.0 (23.0)953024.0 (75.2)235453.0 (127.4)724.50
-4.0 (24.8)904625.0 (77.0)225254.0 (129.2)697.90
-3.0 (26.6)858626.0 (78.8)215655.0 (131.0)672.50
-2.0 (28.4)815127.0 (80.6)206456.0 (132.8)648.10
-1.0 (30.2)774128.0 (82.4)197757.0 (134.6)624.80
0.0 (32.0)735529.0 (84.2)189458.0 (136.4)602.40
1.0 (33.8)698930.0 (86.0)181559.0 (138.2)580.90
2.0 (35.6)664431.0 (87.8)173960.0 (140.0)560.30
3.0 (37.4)631932.0 (89.6)166761.0 (141.8)540.50
4.0 (39.2)601133.0 (91.4)159962.0 (143.6)521.50
5.0 (41.0)571934.0 (93.2)153363.0 (145.4)503.30
6.0 (42.8)544435.0 (95.0)147164.0 (147.2)485.80

Temperatures between the values listed may be interpolated.

Temp. [°C (°F)] / Resistance [Ω]

Temp. [°C (°F)]Resistance [Ω]Temp. [°C (°F)]Resistance [Ω]Temp. [°C (°F)]Resistance [Ω]
7.0 (44.6)518336.0 (96.8)141265.0 (149.0)469.00
8.0 (46.4)493737.0 (98.6)135566.0 (150.8)452.90
9.0 (48.2)470338.0 (100.4)130167.0 (152.6)437.40
10.0 (50.0)448239.0 (102.2)124968.0 (154.4)422.50
11.0 (51.8)427340.0 (104.0)120069.0 (156.2)408.20
12.0 (53.6)407441.0 (105.8)115270.0 (158.0)394.50
13.0 (55.4)388642.0 (107.6)110771.0 (159.8)381.20
14.0 (57.2)370843.0 (109.4)106472.0 (161.6)368.50
15.0 (59.0)353944.0 (111.2)102373.0 (163.4)356.20
16.0 (60.8)337845.0 (113.0)983.8074.0 (165.2)344.50
17.0 (62.6)322646.0 (114.8)946.2075.0 (167.0)333.10
18.0 (64.4)308147.0 (116.6)910.2076.0 (168.8)322.30
19.0 (66.2)294448.0 (118.4)875.8077.0 (170.6)311.80
20.0 (68.0)281449.0 (120.2)842.8078.0 (172.4)301.70
21.0 (69.8)269050.0 (122.0)811.3079.0 (174.2)292.00
22.0 (71.6)257251.0 (123.8)781.1080.0 (176.0)282.70
23.0 (73.4)246052.0 (125.6)752.20

Engineering Facts

14 facts · every claim linked to a page
Selection

The instrument supports Apparatus 1, 2, 5, and 6 and vessel sizes from 100 mL to 2 L, so it can cover a broad range of dissolution methods with one platform.

Determines whether the apparatus can be used for the required USP apparatus type and vessel volume before purchase.

Source
“The instrument can be configured for use with baskets (Apparatus 1), paddles (Apparatus 2), paddle over disk assemblies (Apparatus 5), and rotating cylinders (Apparatus 6), and can accommodate vessel sizes from 100 mL to 2 L.”
Installation

The instrument needs 10 cm side clearance, about 20 cm rear clearance, and at least 30 cm above when fully raised, so bench layout must be planned accordingly.

Insufficient clearance can block airflow, hinder connections, and prevent safe operation of the raised drive unit.

Source
“It needs an additional 10 cm (4 inches) of space on either side and approximately 20 cm (8 inches) in the rear for air circulation and easy access to liquid and electric connections. Ensure a minimum clearance of 30 cm (12 inches) above the unit (when fully raised).”
Installation

The power supply accepts 115/230 V ±10% at 50/60 Hz with no voltage selector or externally accessible fuses, simplifying site power planning.

Confirms the instrument can be connected to either 115 V or 230 V laboratory supply without reconfiguration.

Source
“It accepts any line voltage in the range described in Specifications of the 708-DS Dissolution Apparatus (G7910A / G7911A) on page 13. Consequently there is no voltage selector in the rear of the instrument.”
Safety

The power connector must remain easily reachable and unpluggable, and sufficient space behind the socket is required for emergency disconnection.

Emergency disconnection must be possible at any time; poor placement creates an electrical hazard.

Source
“Make sure the power connector of the instrument can be easily reached and unplugged.”
Safety

Only the Agilent-supplied power cord may be used and it must not be repurposed for other equipment.

Using an incorrect cord risks electric shock or equipment damage and may violate warranty conditions.

Source
“Use only the power cord supplied by Agilent with this instrument.”
Safety

The instrument must be connected to a power outlet with a protective ground connection; operating without proper grounding is prohibited.

Missing ground can cause electric shock or short circuit, so the site outlet must be verified before installation.

Source
“Do not operate the instrument from a power outlet that does not have a protective ground connection.”
Operation

Retaining ties must be removed before power-on, otherwise the instrument can be significantly damaged.

This is a mandatory first-power-on step that prevents shipping-damage-related failures.

Source
“Failure to remove the retaining ties could result in significant damage to the instrument.”
Operation

Up to 35 automated methods can be stored and selected by number from the Select Method screen.

Method capacity affects how many routine test protocols can be kept on-board without external software.

Source
“You can enter up to 35 automated methods through the Select Method screen.”
Operation

Pausing a method stops the spindles and starts a pause timer; exceeding the configured Pause Duration triggers an alarm.

Operators must know that pause behavior affects elapsed test time and can trigger alarms if media change takes too long.

Source
“The spindles stop.”
Operation

Up to twelve timepoints per test can be programmed, each with its own spindle speed and vessel volume.

Complex dissolution profiles with speed or volume changes require this timepoint capacity to be sufficient.

Source
“Using this Method Properties screen, you can specify up to twelve timepoints per test.”
Maintenance

Paddles and basket shafts must be cleaned by hand only; laboratory dishwashers can damage the fluorocarbon coating.

Dishwasher cleaning can degrade the coating and alter dissolution performance, so cleaning SOPs must reflect hand cleaning.

Source
“We recommend that you do not use a laboratory dishwasher. Clean paddles and basket shafts only by hand.”
Maintenance

Corrosive media such as hydrochloric acid or salt-containing media require immediate rinsing with deionized water and thorough drying after each use.

Delayed rinsing can corrode stainless steel shafts and baskets, affecting test reproducibility and component life.

Source
“When using with corrosive materials such as hydrochloric acid or media containing salts, be sure to rinse them thoroughly with deionized water immediately after each use, and dry thoroughly with a soft towel or cloth.”
Maintenance

Paddles and basket shafts must be stored in the original styrofoam container or another suitable container to prevent nicks, chips, and scratches.

Improper storage damages critical surfaces and can introduce variability into dissolution results.

Source
“Place them back into the original styrofoam shipping container or other appropriate container between uses.”
Operation

The manifold can hold up to two cannulas and one probe per vessel position, and each can be adjusted via a thumbscrew.

This determines sampling and temperature-monitoring configuration per vessel and affects manifold setup.

Source
“The manifold of the apparatus may contain up to two cannulas and one probe per vessel position depending on the options configured.”