Agilent 7200 Accurate-Mass QTOF Operation Manual

Contenido

  1. Overview pp. 9-11
    Abbreviations and overview of the 7200 QTOF system
  2. Overview pp. 12-14
    System components, vacuum gauge, and hardware layout
  3. Warning pp. 15-17
    Important safety warnings, explosion and fire precautions
  4. Quality pp. 18-21
    Safety symbols, EMC compliance, and intended use
  5. Procedure pp. 23-31
    Column selection, conditioning, and installation procedures
  6. Procedure pp. 33-56
    EI operation, pumping down, venting, and temperature control
  7. Procedure pp. 57-74
    CI setup, reagent gas flow, and CI autotune procedures
  8. Procedure pp. 75-88
    RIS installation, removal, and ion source exchange
  9. Procedure pp. 89-113
    Tools, ion source cleaning, filament and vacuum maintenance
  10. p. 65 table, 13 rows
  11. p. 24 table, 9 rows
  12. p. 43 table, 2 rows
  13. p. 67 table, 4 rows
  14. p. 90 table, 10 rows
  15. p. 10 table, 21 rows
  16. p. 11 table, 8 rows
  17. p. 40 table, 7 rows
  18. p. 61 table, 2 rows
  19. p. 62 table, 8 rows

Table 7 Typical CI application settings

ParameterMethane PositiveMethane NegativeIsobutane PositiveIsobutane NegativeAmmonia PositiveAmmonia NegativeEI
Ion polarityPositiveNegativePositiveNegativePositiveNegativeN/A
Emission150 μA50 μA150 μA50 μA150 μA50 μA35 μA
Electron energy150 eV150 eV150 eV150 eV150 eV150 eV70 eV
Filament1111111 or 2
Repeller3 V3 V3 V3 V3 V3 V30 V
Ion focus130 V130 V130 V130 V130 V130 V90 V
Entrance lens offset20 V20 V20 V20 V20 V20 V25 V
Shutoff valveOpenOpenOpenOpenOpenOpenClosed
Suggested flow20%40%20%40%20%40%N/A
Source temp250 °C150 °C250 °C150 °C250 °C150 °C230 °C
Quad temp150 °C150 °C150 °C150 °C150 °C150 °C150 °C
Interface temp280 °C280 °C280 °C280 °C280 °C280 °C280 °C
AutotuneYesYesNoYesNoYesYes

N/A Not available

Table 2

Feature7200 Series
High-vacuum pump 1Split flow turbo
High-vacuum pump 2Split flow turbo
High-vacuum pump 3Turbo
Optimal carrier gas flow, mL/min1.0 to 1.5
Reagent gas flow, mL/min1 to 2
Collision cell gas flow1.5
Maximum recommended gas flow, mL/min*2.0
Maximum gas flow, mL/min†2.4
Maximum column id0.32 mm (30 m length)

* Total gas flow into the MS = column flow + collision cell gas flow + reagent gas flow (if applicable) + Agilent Quick Swap flow (if applicable) † Expect degradation of spectral performance and sensitivity.

Table 4 Recommended temperature settings

ZoneEI operation
MS Source230 °C
MS Quad150 °C

Table 8 Default Tune Control Limits, used by CI autotune only

MethaneMethaneAmmoniaAmmonia
Ion polarityPositiveNegativePositiveNegative
Abundance target1x1061x106N/A1x106
Peakwidth target0.70.7N/A0.7
Maximum repeller44N/A4

Reagent gas spans Methane (Positive, Negative) and Ammonia (Positive, Negative); the first column has no header in the printed table.

Table 9 Maintenance schedule

TaskEvery weekEvery 6 monthsEvery yearAs needed
Tune the MSX
Check the foreline pump oil levelX
Check the calibration vial(s)X
Replace the foreline pump oil*X
Clean the ion sourceX
Check the carrier gas trap(s) on the GC and MSX
Replace the worn out partsX
Replace gasket on RIS probeXX
Check the IRM vialX
Replace GC gas suppliesX

*Or as needed.

Table 1 Abbreviations

AbbreviationDefinition
ACAlternating current
ALSAutomatic liquid sampler
BFBBromofluorobenzene (calibrant)
CCCollision cell
CIChemical ionization
DCDirect current
DFTPPDecafluorotriphenylphosphine (calibrant)
EIElectron impact
EPCElectronic pneumatic control
eVElectron volt
GCGas chromatograph
idInside diameter
LANLocal Area Network
m/zMass to charge ratio
MFCMass flow controller
MSMass spectrometer
NCINegative chemical ionization
OFNOctafluoronaphthalene (sample)
PCIPositive chemical ionization
PFDTDPerfluoro-5,8-dimethyl-3,6,9-trioxydodecane (calibrant)
PFET2,4,6-tris (Pentafluoroethyl)-1,3,5-triazine

Table 1 Abbreviations (continued)

AbbreviationDefinition
PFTBAPerfluorotributylamine (calibrant)
Q-TOFQuadrupole time-of-flight
QuadQuadrupole mass filter
RFRadio frequency
RFPARadio frequency power amplifier
TOFTime-of-flight
TorrUnit of pressure, 1 mm Hg
TurboTurbomolecular vacuum pump

Table 3 Influence of carrier and collision cell gas flows on vacuum

Column Flow (mL/min)CC Gas On N2 = 1.5 mL/min Rough VacCC Gas On N2 = 1.5 mL/min Quad VacCC Gas On N2 = 1.5 mL/min TOF VacCC Gas Off Rough VacCC Gas Off Quad VacCC Gas Off TOF Vac
0.51.16 * 10-13.41 * 10-54.20* 10-78.29 * 10-23.17 * 10-72.15 * 10-7
0.71.18 * 10-13.41 * 10-54.20 * 10-78.67 * 10-23.45 * 10-72.15 * 10-7
11.22 * 10-13.43 * 10-54.22 * 10-79.17 * 10-23.79 * 10-72.16 * 10-7
1.21.24 * 10-13.43 * 10-54.22 * 10-79.49 * 10-24.06 * 10-72.16 * 10-7
21.32 * 10-13.44 * 10-54.22 * 10-71.07 * 10-15.17 * 10-72.17 * 10-7
31.42 * 10-13.44 * 10-54.25 * 10-71.19 * 10-16.51 * 10-72.20 * 10-7
41.50 * 10-13.46 * 10-54.27 * 10-71.29 * 10-17.84 * 10-72.21 * 10-7

These pressures are approximate and will vary from instrument to instrument by as much as 30%.

Table 5 Temperatures for CI operation

Ion sourceAnalyzerGC/MS interface
PCI300 °C150 °C280 °C
NCI150 °C150 °C280 °C

Table 6 Typical analyzer vacuum with reagent gas flow

MFC (%)Collision cell gas flow on N2 = 1.5 mL/min — Rough Pump (mTorr)Collision cell gas flow on N2 = 1.5 mL/min — Quadrupole (Torr)Collision cell gas flow on N2 = 1.5 mL/min — Flight Tube (Torr)Collision cell gas flow off N2 = 0 mL/min — Rough Pump (mTorr)Collision cell gas flow off N2 = 0 mL/min — Quadrupole (Torr)Collision cell gas flow off N2 = 0 mL/min — Flight Tube (Torr)
01.36e+023.62e-053.35e-079.13e+015.98e-071.64e-07
101.36+023.62e-053.37e-071.14e+011.27e-061.65e-07
151.43+023.66e-053.37e-071.23e+011.62e-061.67e-07
201.5+023.71e-053.39e-071.31e+011.96e-061.67e-07
251.57+023.73e-053.41e-071.39e+012.32e-061.70e-07
301.63+023.77e-053.41e-071.46e+012.64e-061.71e-07
351.69+023.81e-053.41e-071.52e+013.00e-061.71e-07
401.74+023.83e-053.43e-071.58e+013.34e-061.72e-07

MFC is calibrated for methane and the vacuum gauge is calibrated for nitrogen, so these measurements are not accurate, but are intended as a guide to typical observed readings. Measurements will vary by as much as 30% from one MS to the next.

Engineering Facts

13 facts · every claim linked to a page
Safety

Hydrogen carrier gas must never be used with this system, so the gas supply and safety plan must be built around an alternative carrier.

Carrier gas selection and lab safety design must exclude hydrogen before the instrument is specified or installed.

Source
“The use of hydrogen gas is specifically prohibited with this product.”
Installation

The column installation procedure is instrument-specific, so technicians must not reuse procedures from other MS systems.

Using a wrong installation procedure can damage the column or the MS during setup.

Source
“The column installation procedure for the 7200 Q-TOF GC/MS is different from that for other MSs.”
Installation

Carrier gas purity must be at least 99.9995%, which drives the choice of gas supply and traps.

Gas purity directly affects column bleed, ion source contamination, and MS sensitivity.

Source
“Always use carrier gas that is at least 99.9995% pure.”
Maintenance

Columns must be conditioned before connection to the MS interface to limit stationary-phase deposits in the ion source.

Skipping conditioning increases ion source cleaning frequency and reduces sensitivity.

Source
“Conditioning a column before it is connected to the GC/MS interface is essential.”
Installation

New ferrules can loosen after thermal cycling, so tightness must be rechecked after a few heating cycles.

Leak checks after the first heating cycles prevent vacuum loss and air leaks.

Source
“Because of thermal expansion, new ferrules may loosen after heating and cooling a few times.”
Operation

The GC/MS interface and oven must not be heated until carrier gas is on and the column is purged of air.

Heating without gas flow damages the column and contaminates the MS.

Source
“Ensure that the carrier gas is turned on and the column has been purged of air before heating the GC/MS interface or the GC oven.”
Operation

The analyzer cover should only be opened for pumpdown or sideplate thumbscrew access, not for general work.

Unnecessary opening risks vacuum loss and contamination of the analyzer.

Source
“You should not open the analyzer cover for any other reason.”
Operation

Reagent gas flow rates differ between PCI and NCI startup, so the flow controller must be set per mode.

Wrong reagent gas flow prevents stable CI tuning and sensitivity.

Source
“PCI mode set reagent gas flow to 20 (1 mL/min)”
Performance

EI performance must be verified before switching to CI, so a baseline EI check belongs in the workflow.

This confirms the system is healthy before introducing reagent gas and changing source conditions.

Source
“Always verify MS performance in EI before switching to CI operation.”
Maintenance

The RIS allows EI/CI source switching in under 30 minutes while keeping the instrument under vacuum.

This short cycle time affects method scheduling and uptime planning.

Source
“Switching back and forth between CI and EI ion sources takes less than 30 minutes with the new removable ion source (RIS).”
Selection

The system requires a methane/isobutane gas purifier for CI, which must be included in the installation scope.

The purifier removes oxygen, water, hydrocarbons, and sulfur compounds that would degrade CI performance.

Source
“For the CI system, a methane/isobutane gas purifier is provided and is required.”
Safety

The foreline pump exhaust must be vented outside or to a fume hood when toxic solvents or chemicals are analyzed.

Lab exhaust routing must be planned before installation to protect operators.

Source
“attach a hose (11-mm id) to the oil mist filter to take the foreline pump exhaust outside or to a fume (exhaust) hood.”
Installation

The instrument weighs 152 kg and the foreline pump adds 21.5 kg, so floor loading and rigging must be planned.

Bench strength and moving equipment must be sized for the total installed weight.

Source
“The weight of the instrument mainframe is 152 kg.”