7250 Accurate-Mass Q-TOF GC/MS Operating Manual

Contenido

  1. Overview pp. 11-14
    System overview, abbreviations, and quadrupole analyzer description
  2. Overview pp. 15-22
    Vacuum gauge, safety warnings, symbols, and intended use
  3. Procedure pp. 23-34
    Column installation, conditioning, and interface tip seal procedures
  4. Procedure pp. 35-58
    EI setup, pumping down, temperature and collision cell settings, autotune
  5. Procedure pp. 59-76
    CI setup, reagent gas flow control, CI autotune, and mode switching
  6. Procedure pp. 77-108
    Tools, safety, vacuum and analyzer maintenance, LE-EI ion source procedures
  7. Procedure pp. 109-126
    Axial CI ion source removal, cleaning, assembly, and filament replacement
  8. p. 67 table, 13 rows
  9. p. 24 table, 9 rows
  10. p. 45 table, 2 rows
  11. p. 78 table, 14 rows
  12. p. 90 table, 18 rows
  13. p. 95 table, 18 rows
  14. p. 116 table, 17 rows
  15. p. 12 table, 26 rows
  16. p. 42 table, 5 rows
  17. p. 63 table, 2 rows

Table 7 Typical CI application settings

ParameterMethane PositiveMethane NegativeIsobutane PositiveIsobutane NegativeAmmonia PositiveAmmonia Negative
Ion polarityPositiveNegativePositiveNegativePositiveNegative
Emission150 A50 A150 A50 A150 A50 A
Electron energy150 eV150 eV150 eV150 eV150 eV150 eV
Filament111111
Repeller3 V3 V3 V3 V3 V3 V
Ion focus130 V130 V130 V130 V130 V130 V
Entrance lens offset20 V20 V20 V20 V20 V20 V
Shutoff valveOpenOpenOpenOpenOpenOpen
Suggested flow20%40%20%40%20%40%
Source temp250 °C150 °C250 °C150 °C250 °C150 °C
Quad temp150 °C150 °C150 °C150 °C150 °C150 °C
Interface temp280 °C280 °C280 °C280 °C280 °C280 °C
AutotuneYesYesNoYesNoYes

Emission values printed as '150 A' and '50 A' (microampere symbol lost in text extraction).

Table 2 Gas flows

FeatureGas flows
High-vacuum pump 1Split-flow turbo
High-vacuum pump 2Turbo
High-vacuum pump 3Turbo
Optimal carrier gas flow, mL/min1.0 to 1.5
Collision cell Nitrogen gas flow, mL/min1
Collision cell Helium gas flow, mL/min4
Maximum recommended gas flow, mL/min*2.0
Maximum gas flow, mL/min†2.4
Maximum column id0.32 mm (30 m length)

* Total gas flow into the MS = column flow + collision cell gas flow. † Expect degradation of spectral performance and sensitivity.

Table 4 Recommended temperature settings

ZoneEI operation
MS Source230 °C default. High limit 350 °C. Application dependent.
MS Quad150 °C

Table 9 Maintenance schedule

TaskEvery weekEvery 6 monthsEvery yearAs needed
Check the foreline pump oil levelX
Check the calibration vial(s)X
Replace the foreline pump oil*X
Replace the tip seals on the optional IDPX
Replace the filters on the optional IDPX
Clean the ion sourceX
Check the carrier gas trap(s) on the GC and MSX
Replace the worn out partsX
Replace GC gas suppliesX
Replace CI reagent gas suppliesX
Check the foreline pumpX
Tune the MSX
Leak check the systemX
Lubricate side plate or vent valve O-rings†X

* Or as needed, every 3 months if ammonia is being used as CI reagent gas. † Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be lubricated. Lubricating other seals can interfere with their correct function.

Table 10 Parts list for LE-EI ion source (Figure 21)

Item numberItem descriptionPart number
1Source finger gripG7002-20008
2Filament blockG7002-20019
3Extractor lens (5)*, with 3 mm openingG7004-20061
4Ceramic insulator for extractorG7002-20064
5Entrance lens assembly (1)*G7250-20075
6Ion focus lens (2)*G7004-20068
7Lens insulator/holderG7002-20074
8M2 × 0.4 screw × 12 mm long gold plated screwG7002-20083
9Source bodyG7002-20084
10Post extractor lens 2 (3)*G7004-20090
11Post extractor lens 1 (4)*G7004-20004
12M2 × 6 mm gold plated screwG7002-20109
13Locking ring lens insulatorG7002-20126
14Filament assembly, HE, SingleG3850-60021
15Ring heater/sensor assemblyG7002-60043
16Source mount 1.5 mmG7002-60053
17Repeller assemblyG7002-60057
Complete assemblyIon Source Assy, LE-EI 7250AG7250-67170

*The number in parenthesis is the number engraved on the lens.

Table 11 Parts list for LE-EI ion source (Figure 24)

Item numberItem descriptionPart number
1Source finger gripG7002-20008
2Filament blockG7002-20019
3Extractor lens (5)*, with 3 mm openingG7004-20061
4Ceramic insulator for extractorG7002-20064
5Entrance lens assembly (1)*G7250-20075
6Ion focus lens (2)*G7004-20068
7Lens insulator/holderG7002-20074
8M2 × 0.4 screw × 12 mm long gold plated screwG7002-20083
9Source bodyG7002-20084
10Post extractor lens 2 (3)*G7004-20090
11Post extractor lens 1 (4)*G7004-20004
12M2 × 6 mm gold plated screwG7002-20109
13Locking ring lens insulatorG7002-20126
14Filament assembly, HE, singleG3850-60021
15Ring heater/sensor assemblyG7002-60043
16Source mount 1.5 mmG7002-60053
17Repeller assemblyG7002-60057
Complete assemblyIon Source Assy, LE-EI 7250AG7250-67170

*The number in parenthesis is the number engraved on the lens.

Table 12 Parts list for Axial CI ion source

Item numberItem descriptionPart number
1Source finger gripG7002-20008
2Filament blockG7002-20019
3Filament, HES CIG7250-60075
45973 Gold Fil Screw (M2 × 0.4 10MM-LG)G1999-20021
5in-WSHR-Shoulder 0.25 0.37 0.06 Alumina3050-2670
6HES CI source mount assemblyG7250-60009
7HES CI repeller assyG7250-60102
8GCMS Ring heater/sensor assemblyG7002-60058
9HES CI BodyG7250-20097
10Drawout plateG1999-20446
11HES CI Drawout cylinderG7250-20098
12HES Ramp CI Ion focusG7250-20096
13Entrance lens assy, STD, Half-RampedG7250-20075
14HES CI Lens insulatorG7250-20095
15Locking ring for lens insulatorG7002-20126
16Screw, M2 × 6 Pan Head Torx, gold platedG7002-20109
Complete assemblyAxial CI Ion Source Assy (Support)G7250-67404

Table 1 Abbreviations

AbbreviationDefinition
ACAlternating current
ALSAutomatic liquid sampler
CCCollision cell
CIChemical ionization
DCDirect current
EIElectron impact
EPCElectronic pneumatic control
eVElectron volt
GCGas chromatograph
GC/MSGas chromatograph mass spectrometer
idInside diameter
IDPIsolated dry pump
LANLocal Area Network
LE-EILow energy EI
m/zMass-to-charge ratio
MFCMass flow controller
MSMass spectrometer
OFNOctafluoronaphthalene (sample)
PFTBAPerfluorotributylamine (calibrant)
Q-TOFQuadrupole time-of-flight
QuadQuadrupole mass filter
RFIRadio frequency interference
RFPARadio frequency power amplifier
TOFTime-of-flight
TorrUnit of pressure, 1 mm Hg
TurboTurbomolecular vacuum pump

Table 3 Influence of carrier and collision cell gas flows on vacuum

Column flow (mL/min)CC Gas On N2 = 1 mL/min He = 4 mL/min Rough Vac (mTorr)CC Gas On N2 = 1 mL/min He = 4 mL/min Quad Vac (Torr)CC Gas On N2 = 1 mL/min He = 4 mL/min TOF Vac (Torr)CC Gas Off Rough Vac (mTorr)CC Gas Off Quad Vac (Torr)CC Gas Off TOF Vac (Torr)
0.71.25 E+023.50 E-052.61 E-072.70 E+015.78 E-071.58 E-07
11.30 E+023.49 E-052.61 E-073.61 E+016.21 E-071.58 E-07
1.21.34 E+023.49 E-052.61 E-074.13 E+016.55 E-071.58 E-07
21.48 E+023.51 E-052.61 E-076.18 E+018.09 E-071.59 E-07
31.63 E+023.52 E-052.61 E-078.45 E+019.99 E-071.60 E-07

Pressures are approximate and will vary from instrument to instrument by as much as 30%.

Table 5 Temperatures for CI operation

Ion sourceAnalyzerGC/MS interface
PCI300 °C150 °C280 °C
NCI150 °C150 °C280 °C

Engineering Facts

14 facts · every claim linked to a page
Safety

Hydrogen gas must not be used with this instrument under any circumstances.

Carrier gas selection must exclude hydrogen; using it violates the manufacturer's safety requirements.

Source
“The use of hydrogen gas is specifically prohibited with this product.”
Installation

The analyzer side plate top thumbscrew must be finger-tight only, and the collision cell top plate shipping brackets must remain fastened during normal operation.

Improper torque causes vacuum leaks, and removing the shipping brackets compromises explosion protection.

Source
“You MUST ensure the top thumbscrew on the analyzer side plate is fastened finger-tight. Do not over-tighten the thumbscrews; this can cause air leaks.”
Safety

Combustible or absorbent materials must not be left in the oil pan under the foreline pump.

Housekeeping around the foreline pump is a fire-safety requirement during installation and operation.

Source
“Combustible materials (or flammable/nonflammable wicking material) placed under, over, or around the foreline (roughing) pump constitutes a fire hazard.”
Installation

The GC/MS interface temperature should be slightly above the maximum GC oven temperature but must never exceed the maximum column temperature.

This sets the allowable interface setpoint relative to oven and column limits to avoid column damage.

Source
“the interface temperature should be slightly higher than the maximum GC oven temperature, but never higher than the maximum column temperature.”
Operation

The quadrupole setpoint is software-limited to 200 °C and the ion source to 350 °C.

Defines the maximum allowable heated-zone setpoints when configuring the method.

Source
“The software will not allow you to exceed 200 °C for the quadrupole, or 350 °C for the ion source.”
Operation

Carrier gas flow must be turned on before any GC heated zones are energized.

Prevents column damage during pumpdown and startup sequencing.

Source
“Do not turn on any GC heated zones until carrier gas flow is on. Heating a column with no carrier gas flow will damage the column.”
Operation

After the 'Okay to run' message, wait 2 hours for thermal equilibrium before acquiring data.

Data acquired before thermal equilibrium may not be reproducible, so this wait is required for valid results.

Source
“After the message Okay to run appears, wait 2 hours for the MS to reach thermal equilibrium.”
Operation

Turbo pump speed should reach 80% within 10–15 minutes and eventually 95%; otherwise the electronics shut off the foreline pump and a power cycle is required.

Provides the acceptance criterion for a successful pumpdown and the recovery action if it fails.

Source
“Within 10 to 15 minutes, the turbo pump speed should be up to 80%.”
Operation

The MS must never be vented by admitting air through either end of the foreline hose.

Defines the only acceptable venting method to avoid contaminating the analyzer.

Source
“Never vent the MS by allowing air in through either end of the foreline hose.”
Maintenance

When running ammonia five or more hours per day, the foreline pump must be ballasted with air at least one hour per day, and the MS purged with methane after ammonia use.

Ammonia CI imposes a mandatory daily maintenance routine to protect pump seals.

Source
“When running ammonia for five or more hours a day, the foreline pump must be ballasted (flushed with air) for at least one hour a day to minimize damage to pump seals.”
Operation

The ammonia supply pressure must be kept below 5 psig and the tank kept upright below the flow module level.

Higher pressure can condense ammonia to liquid and damage the flow module.

Source
“The pressure of the ammonia supply must be less than 5 psig.”
Maintenance

Do not use compressed air to blow dust out of the optional IDP foreline pump when replacing its tip seal.

Compressed air would spread potentially hazardous chemical-laden dust into the laboratory.

Source
“When replacing the pump's tip seal, do NOT use compressed air to blow out this dust.”
Selection

The instrument mainframe weighs 159 kg, with the foreline pump adding 25 kg (DS202 rotary vane) or 45 kg (IDP).

Bench/site load-bearing and rigging requirements depend on these masses.

Source
“The weight of the instrument mainframe is 159 kg.”
Selection

The flight tube extends 136 cm above the top of the instrument, which must be accounted for in lab ceiling/clearance planning.

Vertical clearance above the bench is a site-preparation constraint.

Source
“The flight tube extends 136 cm over the top of the instrument.”