Background on antibody internalization and assay rationale
Procedurep. 3
Materials, cell lines, labeling and imaging protocol
Procedurep. 4
Methods for quantifying internalization kinetics
pp. 5-7
Findings on direct versus indirect labeling and HER2 internalization
Overviewp. 8
Summary of assay performance and key takeaways
p. 9
Cited literature and sources
Orderingp. 10
List of reagents and products referenced in the application
p. 3table, 5 rows
Table 1. Cell lines and their growth media
Cell Lines
Base Media
Media Supplements
MCF7
DMEM (Gibco, p/n 12430-054)
10% FBS + 1% pen/strep
HCC1954
RPMI 1640 (Gibco, p/n 11875-093)
10% FBS + 1% pen/strep
SKOV3
Moccy's 5A (Gibco, p/n 16600-082)
10% FBS + 1% pen/strep
SKBR3
RPMI 1640 (Gibco, p/n 11875-093)
10% FBS + 1% pen/strep
BT474
Hybri-Care (ATCC, p/n 46-X)
10% FBS + 1% pen/strep
Engineering Facts
12 facts · every claim linked to a page
Operation
The assay requires a 37 °C incubator with 5% CO2, so the lab must provide controlled temperature and CO2 conditions.
Determines the environmental infrastructure needed before selecting and installing the eSight workflow.
Source
“Cell maintenance and assays were conducted in a 37 °C incubator with 5% CO2.”
Selection
CypHer5E is a pH-sensitive dye that is minimally fluorescent at basic pH and maximally fluorescent at acidic pH, which is the basis for detecting internalization.
Confirms the detection chemistry must match acidic endosomal/lysosomal compartments when choosing the reagent.
Source
“is a pH-sensitive cyanine dye derivative that is minimally fluorescent at basic pH and shows maximal fluorescence at an acidic pH.”
Operation
Direct labeling is performed at pH 8.3 in 7.5% sodium bicarbonate with a 20-fold molar excess of dye for one hour at room temperature in the dark.
Defines the conjugation protocol and buffer conditions the operator must reproduce.
Source
“Antibodies were diluted to 1 mg/mL in 7.5% sodium bicarbonate, pH 8.3. A 20-fold molar excess of CypHer5E dye was added to the antibody solutions and incubated for one hour at room temperature in the dark.”
Operation
Unconjugated dye is removed by dialysis against PBS overnight at 4 °C or by a PD minitrap G-25 desalting column.
Specifies the purification step and cold-chain handling needed for reliable labeled antibody.
Source
“The unconjugated dye and low molecular weight contaminants were removed by dialyzing against PBS overnight at 4 °C or fractionating over a PD minitrap G-25 desalting column”
Maintenance
Labeled antibodies can be stored at 4 °C in the dark for 24 hours, and long-term storage requires –15 to –30 °C.
Sets storage requirements that affect freezer availability and reagent logistics.
Source
“Labeled antibodies can be stored at 4°C in the dark for 24 hours. For long-term storage, it is recommended to be stored at –15 to –30 °C.”
Performance
The eSight enables continuous real-time measurement of antibody internalization for more than 72 hours, unlike endpoint detection.
Justifies choosing a kinetic live-cell imaging platform over endpoint assays.
Source
“the xCELLigence RTCA eSight enables continuous real-time measurement of antibody internalization for more than 72 hours.”
Performance
The normalized red area is calculated as red total area divided by brightfield total area to represent internalization rate and reduce seeding/proliferation variation.
Defines the metric engineers must configure in the analysis software for robust comparison.
Source
“normalized red area = red total area / brightfield total area”
Operation
An enhanced red fluorescent filter set with extended wavelength is used, and exposure times may need user adjustment to optimize red intensity.
Informs optical configuration and calibration before running the assay.
Source
“we utilized an enhanced Agilent xCELLigence RTCA eSight filter set to further improve performance of the red fluorescent channel, including extended wavelength.”
Selection
Secondary Fab-CypHer5E labeling is cost-efficient for labeling multiple antibodies at once and avoids D/P ratio variation.
Guides reagent strategy when screening large antibody libraries.
Source
“Secondary Fab-CypHer5E offers a cost-efficient way to label multiple antibodies simultaneously, eliminating the potential disturbance from variations in the D/P ratio.”
Performance
Direct and indirect labeling gave close EC50 values (89.37 and 124.00 ng/mL), showing both methods accurately evaluate internalization kinetics.
Supports method selection between direct and indirect labeling based on equivalent performance.
Source
“The EC50 values for trastuzumab-CypHer5E and trastuzumab-Fab-CypHer5E were very close (89.37 and 124.00 ng/mL, respectively)”
Operation
Images were captured for 48 hours at one-hour intervals after adding labeled antibody 24 hours post-seeding.
Defines the imaging schedule and timing the operator must program.
Source
“Brightfield and red fluorescence images were captured for 48 hours at one-hour intervals.”
Cumplimiento
The product is for research use only and not for diagnostic procedures.
Constrains the intended use and regulatory classification at installation.
Source
“For Research Use Only. Not for use in diagnostic procedures.”