Using a Deans Switch to Divert Derivatizing Agents from the EI Source for Improved Robustness
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación sobre el uso de un conmutador Deans para desviar agentes de derivatización (MSTFA) de la fuente de EI en flujos de trabajo de GC/MS/MS para control de dopaje. Abarca dos configuraciones de instrumentos: Agilent 7000 Series GC/TQ con gas portador hidrógeno y Agilent 7010 Series GC/TQ con gas portador helio; dos tipos de entrada (entrada split/splitless SSL y entrada de vaporizador de temperatura programable).
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Purge flow to split vent | 50 mL/min p.6 ↗ | SSL inlet, at 0.75 min |
| Purge flow to split vent | 50 mL/min p.6 ↗ | PTV inlet, at 0.75 min |
| Injection volume | 1.5 µL p.6 ↗ | SSL inlet |
| Switching time | 4.9 min p.6 ↗ | SSL inlet, from vent to MS |
| Column temperature program | 2.5 °C/min to 199 °C (0 min hold) °C p.6 ↗ | SSL inlet |
| Carrier gas and flow rate | Hydrogen, 1 mL/min constant flow p.6 ↗ | SSL inlet |
| Transfer line temperature | 300 °C p.6 ↗ | SSL inlet |
| Electron energy | 70 eV p.6 ↗ | SSL inlet |
| Electron energy | 70 eV p.6 ↗ | PTV inlet |
| Quench gas (helium) | 2.25 mL/min p.6 ↗ | PTV inlet |
| Collision gas (nitrogen) | 1.5 mL/min p.6 ↗ | SSL inlet |
| Collision gas (nitrogen) | 1.5 mL/min p.6 ↗ | PTV inlet |
| Ion source temperature | 280 °C p.6 ↗ | SSL inlet |
| Ion source temperature | 280 °C p.6 ↗ | PTV inlet |
| Tune | Etune: atunes.eiex.tune.xml p.6 ↗ | SSL inlet |
| Cleaning interval | 350–400 to ≥ 1,500 injections p.1 ↗ | Agilent 7000 Series GC/TQ, hydrogen carrier gas |
Operating Limits
- Source cleaning interval improvement (7000 Series GC/TQ, hydrogen): from approximately 350–400 to ≥ 1,500 injections p.1 ↗
- Source cleaning interval improvement (7010 Series GC/TQ, helium): from approximately every 300 to 350 injections to approximately 900 injections p.1 ↗
- Routine batch size: 80 samples p.1 ↗
- Source cleaning frequency with routine batches: shifted from weekly to every 2 to 3 weeks p.1 ↗