Introduces the end-to-end dissolution workflow concept
pp. 3-4
Presents measurement results and discussion of the workflow
p. 5
Summarizes findings and key takeaways
p. 2table, 7 rows
LC method
Parameter
Value
Solvents
A) Water
B) Acetonitrile
Flow Rate
0.5 mL/min
Gradient
Time (min) %B
0 15
5 65
5.1 95
Stop time: 6 min
Post time: 2 min
Injection Volume
5 µL
Needle Wash
3 s in Solvent A
Column Temperature
50 °C
Detection
254/4 nm, Ref. 360/16 nm, data rate 20 Hz
Engineering Facts
10 facts · every claim linked to a page
Selection
Choose an HPLC separation over UV spectroscopy when excipients or multiple APIs compromise compound separation.
Determines whether HPLC is required instead of simpler UV detection for the method.
Source
“Compound separation in UV spectroscopic analysis can be compromised by excipients or the presence of several APIs in the formulation.”
Selection
Use the HDR detection solution when multiple APIs span a broad concentration range so all are detected with best sensitivity in one run.
Guides detector selection for wide dynamic range samples.
Source
“If the concentration of multiple APIs varies over a broad range, the Agilent HDR solution will detect them with the best possible sensitivity.”
Installation
The multisampler must be equipped with a sample ID reader (G4756A or option #110) to enable barcode sample tracking.
Defines the required hardware option for the barcode workflow.
Source
“equipped with a sample ID reader (G4756A or option #110 of all 1260/1290 Infinity III multisamplers).”
Selection
Two DAD units with different path-length cells (60 mm and 3.7 mm) are needed to cover high- and low-abundant compounds.
Specifies the detector configuration required for HDR detection.
Source
“2x Agilent 1290 Infinity III diode array detector (G7117B) with 60 mm Max-Light cartridge cell (G4212-60007) and 3.7 mm Max-Light cartridge cell (G4212-60032)”
Operation
Dissolution sampling must follow the defined time points of 10, 20, 40, 60, and 75 minutes.
Sets the sampling schedule the autosampler sequence must follow.
Source
“Samples were drawn from each vessel at 10, 20, 40, 60, and 75 minutes.”
Operation
Samples of 1 mL are withdrawn, automatically filtrated, and filled into barcoded vials in the 850-DS autosampler.
Defines the sample handling volume and automation path.
Source
“samples of 1 mL were withdrawn at the defined time points and automatically filtrated and filled into the barcoded vials arranged accordingly in the Agilent 850‑DS autosampler.”
Operation
The sample ID reader scans vial barcodes and compares them to the sequence before measurement is submitted.
Explains the verification step that prevents sample mix-ups.
Source
“the sample ID reader scanned the barcodes of the vials and compared them to the obtained sequence.”
Performance
HDR detection combines high- and low-abundant API signals in one run without exceeding the detector dynamic range.
Confirms the performance benefit justifying HDR hardware.
Source
“provides a combined chromatogram of both the high- and low-abundant API compounds in one run4 without exceeding the dynamic range of the detector.”
Selection
HDR detection is a valuable alternative to UV spectroscopy for complex API mixtures.
Supports the decision to replace UV with HDR HPLC.
Source
“making it a valuable alternative to UV spectroscopy analysis in the analysis of complex API mixtures.”
Cumplimiento
Software must be OpenLab CDS revision 2.8 or later with Sample Scheduler for OpenLab.
Defines minimum software version for the LIMS-integrated workflow.
Source
“Agilent OpenLab CDS revision 2.8, or later version with Agilent Sample Scheduler for OpenLab”