Capability of Spectral Flow Cytometry

Contenido

  1. Overview p. 2
    Introduces spectral flow cytometry and FITC combination concept
  2. pp. 3-5
    Presents experimental results combining FITC with various fluorochromes
  3. p. 6
    Summarizes findings on spectral flow cytometry capability
  4. Ordering p. 7
    Lists products used in this application
  5. p. 2 table, 18 rows
  6. p. 3 table, 19 rows

Table 1. Antibody information.

MarkerFluorochromeCloneVendorPart number
CD45PEHI30BioLegend304007
CD45BV510HI30BioLegend304036
CD45FITCHI30Agilent8921014
CD3PerCP-Cy5.5SK7Agilent8931015
CD3FITCSK7Agilent8921016
CD4BV421RPA-T4BioLegend300532
CD4Qdot705S3.5Thermo FisherQ10060
CD4BB515RPA-T4BD564419
CD4Alexa Fluor 647RPA-T4BioLegend300520
CD4PE-Cy5.5SK3Thermo Fisher35-0047-41
CD8Pacific BlueSK1BioLegend344718
CD8FITCSK1Agilent8920038
CD8APCSK1Agilent8920265
CD19APCHIB19Agilent8730007
CD19PerCP-eFluor 710J3-129Thermo Fisher46-0197-42
CCR6 (CD196)PEG034E3BioLegend353410
CXCR3 (CD183)BV711G025H7BioLegend353732
CXCR3 (CD183)PE-Cy7G025H7BioLegend353720

Table 5. Antibody information.

GroupMarkerFluorochromeCloneVendorPart number
Violet-excited fluorochromesCD4BV510OKT4BioLegend317444
Violet-excited fluorochromesCD4317444RPA-T4BioLegend300532
Violet-excited fluorochromesCD4BV421RPA-T4BD746541
Violet-excited fluorochromesCD4BV480RPA-T4Thermo Fisher69-0049-42
Violet-excited fluorochromesCD4eFluor 506RPA-T4Thermo Fisher79-0049-41
Violet-excited fluorochromesCD11cPacific Orange3.9Thermo Fisher48-0116-42
Violet-excited fluorochromesCD4PESK3Agilent8920028
Violet-excited fluorochromesHLA-DRCF568TAL 1B5BiotiumBNC682187-500
Violet-excited fluorochromesCD4PE-Dazzle 594RPA-T4BioLegend300548
Violet-excited fluorochromesCD4PE-Alexa Fluor 610S3.5Thermo FisherMHCD0422
Violet-excited fluorochromesCD4PE-Fire 640SK3BioLegend344663
Violet-excited fluorochromesCD4PE-Cy5RPA-T4BioLegend300510
Violet-excited fluorochromesCD117PE-Cy5.5YB5.B8NovusNBP1-43358PECY55
Violet-excited fluorochromesCD25PE-Alexa Fluor 700CD25-3G10Thermo FisherMHCD2524
Blue-excited fluorochromesCD4Alexa Fluor 700RPA-T4BioLegend300526
Blue-excited fluorochromesCD4APCSK3Agilent8931019
Blue-excited fluorochromesCD4Alexa Fluor 647RPA-T4BioLegend300520
Blue-excited fluorochromesCD4Spark NIR 685SK3BioLegend344658
Blue-excited fluorochromesCD4APC-R700RPA-T4BD564975

Engineering Facts

12 facts · every claim linked to a page
Selection

The instrument's five-laser, 70-detector configuration defines which fluorochromes and panels can be supported, so it must be matched to the intended assay panel.

Laser and detector count determines the maximum panel complexity and which fluorochromes can be excited and resolved.

Source
“equipped with five lasers (349, 405, 488, 561, and 637 nm) and 70 fluorescence detectors.”
Selection

Fluorochromes with nearly identical spectra cannot be separated, so panel design must avoid such pairs regardless of instrument capability.

Prevents selection of fluorochrome pairs that the unmixing algorithm physically cannot resolve.

Source
“When FITC was used with Alexa Fluor 488 or KB520, the Similarity Index was 1.00, indicating almost identical spectra, which prevented signal resolution due to such high similarity.”
Performance

Spectral unmixing does not remove spillover spreading error, so closely overlapping fluorochromes will degrade resolution even when successfully unmixed.

Sets realistic expectations for panel performance and drives selection of minimally overlapping fluorochromes.

Source
“spectral spillover can propagate errors when measuring multiple fluorochromes. Spectral unmixing cannot eliminate these errors.”
Selection

The closer the emission spectra, the larger the spillover spreading error, so panel design should minimize spectral overlap.

Directly guides fluorochrome pairing decisions to preserve resolution in multicolor panels.

Source
“The closer the spectra are, the greater the spillover spreading error between fluorochromes.”
Performance

The system can resolve fluorochromes with a Similarity Index as high as 0.99, defining the practical resolution limit for panel design.

Establishes the upper bound of spectral similarity the instrument can handle when selecting fluorochromes.

Source
“NovoCyte Opteon was able to distinguish fluorochromes with a Similarity Index as high as 0.99 (FITC and BB515).”
Operation

Single-stained compensation beads must be acquired and unmixed to generate the spectral signature required for each fluorochrome.

Defines the reference-sample workflow needed before running multicolor unmixing experiments.

Source
“Unmixed single-stained compensation bead samples were used to generate spectrum signatures for each fluorochrome.”
Operation

Blood samples require lysis, centrifugation at 500 × g, washing, and resuspension in PBS with 1% PFA for consistent acquisition.

Specifies the centrifugation and wash protocol that affects sample quality and measurement reproducibility.

Source
“Centrifuge at 500 × g for 5 minutes at room temperature, discard the supernatant and wash the sample twice with PBS.”
Operation

Antibody staining must be incubated at room temperature for 15 minutes protected from light to preserve fluorochrome integrity.

Light protection and timing are required to avoid fluorochrome photobleaching and staining variability.

Source
“Vortex immediately and incubate at room temperature for 15 minutes, protected from light.”
Operation

PBMC antibody staining requires 30 minutes at room temperature in the dark, followed by two washes in staining buffer.

Defines the staining protocol needed for reproducible multicolor panel results.

Source
“Incubate for 30 minutes at room temperature, protected from light.”
Cumplimiento

The instrument is restricted to research use and must not be used for diagnostic procedures.

Determines the permitted application scope and regulatory classification at installation site.

Source
“For Research Use Only. Not for use in diagnostic procedures.”
Selection

Spectral flow cytometry enables simultaneous use of highly overlapping fluorochromes, expanding panel design flexibility.

Justifies choosing spectral cytometry over conventional systems when panel density is a constraint.

Source
“spectral flow cytometry allows for simultaneous use of fluorochromes with highly overlapping spectra, enhancing flexibility in fluorochrome selection.”
Performance

As the Similarity Index decreases, fluorescent spreading error becomes smaller and resolution increases.

Provides the quantitative relationship used to rank candidate fluorochrome pairs during panel design.

Source
“As the Similarity Index decreases, the fluorescent spreading error becomes smaller, and the resolution increases.”